Back to Endress + Hauser Liquid analysis
Endress+Hauser logo

Datasheet summary

Memosens CPS16E

Engineering summary from PDF text extraction for Memosens CPS16E. Verify every value with the OEM datasheet.

Summary

Product identification

Memosens CPS16E pH/ORP sensor for standard applications in process technology and environmental engineering.

Application

Long-term monitoring or limit control in processes with stable conditions, including:

  • Chemical processes
  • Pulp and paper industry
  • Incinerator plants and gas scrubbers
  • Mining
  • Wastewater

The sensor has ATEX, IECEx, CSA C/US, NEPSI, Japan Ex and INMETRO approvals for use in hazardous areas Zone 0, Zone 1 and Zone 2.

Measuring principles

  • pH measurement: Measures acidity or alkalinity based on an electrochemical potential generated by H⁺ ion accumulation on the membrane glass. An Ag/AgCl reference system is used.
  • ORP measurement: Measures the balance between oxidizing and reducing components using a platinum electrode and an Ag/AgCl reference system.
  • rH measurement: Calculated from simultaneous pH and ORP measurements using the formula: rH = 2 · (mV/S) + 2 pH. The rH value indicates the oxidation or reducing power of a solution, with a scale from 0 to 42.

Your benefits

  • Simultaneous measurement of pH, ORP, and rH value (in rH mode).
  • Platinum electrode for measuring reference impedance.
  • Low-maintenance and robust with a large PTFE ring junction.
  • Can be used at pressures up to 17 bar (absolute).
  • Process glass suitable for very alkaline applications.
  • Long service life due to poison-resistant reference with ion trap.
  • Integrated NTC 30K temperature sensor for temperature compensation.
  • Glass breakage and blockage detection by measuring glass membrane resistance and reference impedance.
  • Memosens technology provides:
    • Maximum process safety via non-contact, inductive signal transmission.
    • Data security through digital data transmission.
    • Ease of use with sensor data saved in the sensor.
    • Predictive maintenance by recording sensor load data.

Input

Measured variable

  • pH value
  • ORP
  • rH value
  • Temperature

Measuring range

  • ORP: –1500 to 1500 mV
  • Application A:
    • pH: 1 to 12
    • Temperature: –15 to 80 °C (5 to 176 °F)
  • Application B:
    • pH: 0 to 14
    • Temperature: 0 to 135 °C (32 to 275 °F)

Process limits

Process temperature range

  • Application A: –15 to 80 °C (5 to 176 °F)
  • Application B: 0 to 135 °C (32 to 275 °F)

Process pressure range

  • Applications A and B: 0.8 to 17 bar (11.6 to 246.5 psi) absolute

Conductivity

  • Reference system TA: Minimum 50 μS/cm (minimized flow; pressure and temperature must remain constant)

Mechanical construction

Dimensions

  • Installed lengths available: 120 mm, 225 mm, 360 mm, 425 mm.
  • Sensor shaft diameter: Ø 11.9 mm (0.47 in).

Weight

  • 120 mm installed length: 40 g (1.4 oz)
  • 225 mm installed length: 60 g (2.1 oz)
  • 360 mm installed length: 90 g (3.2 oz)
  • 425 mm installed length: 100 g (3.5 oz)

Materials

  • Sensor shaft: Glass
  • pH membrane glass: Type A, B
  • Metal lead: Ag/AgCl
  • Junction: Ring-shaped PTFE
  • ORP measuring element: Platinum
  • O-ring: FKM
  • Process coupling: PPS fiber-glass reinforced
  • Nameplate: Ceramic metal oxide
  • Temperature sensor: NTC 30K
  • Plug-in head: Memosens plug-in head, pressure resistance 16 bar (232 psi) relative.

Process connections

  • Pg 13.5

Environment

Ambient temperature range

  • Not specified below 0 °C (32 °F) due to risk of frost damage.

Storage temperature

  • 0 to 50 °C (32 to 122 °F)

Degree of protection

  • IP 68 (10 m (33 ft) water column, 25 °C (77 °F), 45 days, 1 M KCl)

Electromagnetic compatibility (EMC)

  • Interference emission and interference immunity as per EN 61326-1:2013, EN 61326-2-3:2013, NAMUR NE21:2017.

Certificates and approvals

  • CE mark
  • Ex approvals: ATEX, IECEx, NEPSI, CSA C/US, Japan Ex, INMETRO (Zone 0, Zone 1, Zone 2)
  • TÜV certificate for Memosens plug-in head (16 bar relative pressure resistance)
  • EAC

Disclaimer

This data sheet was extracted from a datasheet for the Endress+Hauser Memosens CPS16E. The extraction may miss figures, footnotes, or revisions. Contractual data must match the OEM PDF revision used on the project.

Official datasheet (PDF)

Lenntech datasheet mirror

Curated from selected public technical reference material for discovery and preliminary comparison. This summary is not a substitute for a current certified manufacturer datasheet. Verify revisions and design limits before use.