Back to Endress + Hauser Liquid analysis
Endress+Hauser logo

Datasheet summary

Memosens CPS42E

Engineering summary from PDF text extraction for Memosens CPS42E. Verify every value with the OEM datasheet.

Summary

Product identification

  • Product name: Memosens CPS42E
  • Product type: ORP sensor for process technology
  • Technology: Digital with Memosens 2.0 technology

Application

  • Media: Low conductivity, or high proportion of organic solvents or alcohol.
  • Industries: Chemical industry, organic chemicals, power stations, laboratory measurements.
  • Hazardous Areas: ATEX, IECEx, CSA C/US, NEPSI, Japan Ex and INMETRO approvals for Zone 0, Zone 1, and Zone 2.

Operating limits

  • Pressure: Up to 11 bar (159.5 psi) absolute when counterpressure is applied.
  • Temperature (Process): -15 to 135 °C (5 to 275 °F).
  • Temperature (Ambient): Not specified below 0 °C (32 °F) due to risk of frost damage. Storage temperature: 0 to 50 °C (32 to 122 °F).

Feed water

  • Conductivity: Minimum 5 μS/cm (Reference system SB, with minimized flow; pressure and temperature must remain constant).

Product water

  • ORP Measurement Range: -1500 to 1500 mV.

Electrical

  • Power Supply: Via Memosens data cable (CYK10 or CYK20) connected to a Memosens transmitter.
  • Data Transmission: Digital, non-contact, inductive signal transmission.

Mechanical / hydraulic

  • Reference System: Ag/AgCl reference lead with 3M KCl liquid electrolyte.
  • Junction: Ceramic junction with specified KCl flow rate.
  • Temperature Sensor: Integrated NTC 30K.
  • Material:
    • Sensor shaft: Glass
    • ORP measuring element: Platinum
    • O-ring: FKM
    • Process coupling: PPS fiber-glass reinforced
    • Plug-in head: Memosens plug-in head, pressure resistance 16 bar (232 psi) relative.
  • Dimensions: Various installed lengths available (120 mm, 225 mm, 360 mm, 425 mm).
  • Weight: Ranges from 40 g to 100 g depending on installed length.
  • Process Connection: Pg 13.5.
  • Degree of Protection: IP 68.
  • Installation Angle: Minimum 15° from horizontal is required to ensure electrolyte contact.

Certificates and approvals

  • CE Mark: Complies with EU directives.
  • Ex Approvals: ATEX, IECEx, NEPSI, CSA C/US, Japan Ex, INMETRO.
  • Other: EAC certification.

Disclaimer

Extraction may miss figures, footnotes, or revisions; contractual data must match the OEM PDF revision used on the project.

Official datasheet (PDF)

Lenntech datasheet mirror

Curated from selected public technical reference material for discovery and preliminary comparison. This summary is not a substitute for a current certified manufacturer datasheet. Verify revisions and design limits before use.