Memosens CPS76E

Engineering summary from PDF text extraction for Memosens CPS76E. Verify every value with the OEM datasheet.

Summary

Product identification

Memosens CPS76E pH/ORP sensor for process technology with Memosens 2.0 technology.

Application

Designed for process technology and monitoring of processes with:

  • Rapidly changing pH values
  • High proportion of electrode poisons such as H₂S

Approvals for hazardous areas include ATEX, IECEx, CSA C/US, NEPSI, Japan Ex, and INMETRO for Zone 0, Zone 1, and Zone 2.

Features and Benefits

  • Simultaneous measurement of pH, ORP, and rH value (in rH mode).
  • Platinum electrode for measuring reference impedance.
  • Acrylamide-free gel bridging electrolyte, sterilizable and resistant to chemical corrosion.
  • Suitable for CIP/SIP cleaning and autoclavable.
  • Long service life due to poison-resistant reference with ion trap.
  • Integrated NTC 30K temperature sensor for effective temperature compensation.
  • Glass breakage and blockage detection by measuring:
    • Resistance of glass membrane
    • Reference impedance
  • Memosens technology provides:
    • Maximum process safety via non-contact, inductive signal transmission.
    • Digital data transmission for data security.
    • Easy handling as sensor data are saved in the sensor.
    • Predictive maintenance through recording sensor load data.

Measuring Principles

  • pH Measurement: Utilizes the electrochemical potential generated by the membrane glass, which depends on the H⁺ ion accumulation. An integrated Ag/AgCl reference system is used. The Nernst equation converts measured voltage to the pH value.
  • ORP Measurement: Measures the potential using a platinum or gold electrode, with an integrated Ag/AgCl reference system.
  • rH Measurement: Calculated using the formula: rH = 2 * (mV/S) + 2 pH. Requires simultaneous measurement of pH and ORP.
    • rH values indicate the oxidation or reducing power of a process solution.
    • 0 to 9: Strong reducing power
    • 9 to 17: Weak reducing power
    • 17 to 25: Undetermined medium
    • 25 to 34: Weak oxidizing power
    • 34 to 42: Strong oxidizing power

Measuring System Components

A complete measuring system includes:

  • pH/ORP sensor CPS76E
  • Memosens data cable CYK10 or CYK20
  • Transmitter (e.g., Liquiline CM44, Liquiline CM42)
  • Assembly (e.g., immersion, flow, retractable, permanent installation)
  • Optional: Automatic cleaning and calibration system.

Communication and Data Processing

  • Digital sensors with Memosens technology must be connected to Memosens-compatible transmitters.
  • Sensor data stored include: Manufacturer data, serial number, order code, calibration data, application data, operational history, etc.
  • This data can be displayed with specific Endress+Hauser transmitters and software.

Input

Measured Variable

  • pH value
  • ORP
  • rH value
  • Temperature

Measuring Range

  • ORP: –1500 to 1500 mV
  • Application B:
    • pH: 0 to 14
    • Temperature: 0 to 140 °C (32 to 284 °F)
  • Application H:
    • pH: 0 to 12
    • Temperature: 0 to 140 °C (32 to 284 °F)

Performance Characteristics

Reference System

  • TB and TU: Ag/AgCl reference lead with ion trap, gel KCl, 3M, AgCl-free bridging electrolyte.
  • TP: Ag/AgCl reference lead with ion trap, gel KCl, 3M, AgCl-free bridging electrolyte, pressurized to 7 bar (102 psi) absolute.

Installation

  • Orientation: Install with an angle of inclination of at least 15° from the horizontal to prevent air bubble formation. Upside-down installation is possible depending on the order code for the reference system.
  • Tightening Torque: 3 Nm (2.21 lbf ft) by hand (for Endress+Hauser assemblies).

Environment

  • Ambient Temperature Range: 0 to 50 °C (32 to 122 °F). Avoid temperatures below freezing.
  • Storage Temperature: 0 to 50 °C (32 to 122 °F).
  • Degree of Protection: IP 68 (10 m (33 ft) water column, 25 °C (77 °F), 45 days, 1 M KCl).
  • Electromagnetic Compatibility (EMC): Meets EN 61326-1, EN 61326-2-3, and NAMUR NE21.

Process

  • Process Temperature Range: 0 to 140 °C (32 to 284 °F) for Applications B and H, and Version TB. For Versions TU and TP, 0 to 140 °C (32 to 284 °F), with 140 °C (284 °F) limited to sterilization. Continuous operation limited to 100 °C (212 °F) due to pressure loss.
  • Process Pressure Range:
    • Application B: 0.8 to 14 bar (11.6 to 203 psi) absolute.
    • Application H: 0.8 to 7 bar (11.6 to 101.5 psi) absolute.
  • Conductivity: Minimum 10 μS/cm (at atmospheric pressure, without flow).

Mechanical Construction

  • Dimensions: Various installed lengths available: 120 mm (4.72 in), 225 mm (8.86 in), 360 mm (14.17 in), 425 mm (16.73 in).
  • Weight: Ranges from 40 g (1.4 oz) to 100 g (3.5 oz) depending on the installed length.
  • Materials:
    • Sensor Shaft: Glass
    • pH Membrane Glass: Type B, Type N
    • Metal Lead: Ag/AgCl
    • Reference: Ceramic junction (zirconium dioxide)
    • ORP Measuring Element: Platinum
    • O-ring: FKM
    • Process Coupling: PPS fiber-glass reinforced
    • Plug-in Head: Memosens plug-in head (pressure resistant to 16 bar (232 psi) relative).
  • Process Connections: Pg 13.5.

Certificates and Approvals

  • CE marking
  • Ex Approval: ATEX, IECEx, NEPSI, CSA C/US, Japan Ex, INMETRO (various classifications including Ex ia IIC T3/T4/T6 Ga). Hazardous area versions are indicated by a red/orange ring on the plug-in head.
  • TÜV certificate for Memosens plug-in head (pressure resistance 16 bar (232 psi) relative).
  • EAC conformity.

Accessories

A range of assemblies, buffer solutions, and measuring cables are available.


Disclaimer: This extraction may miss figures, footnotes, or revisions. Contractual data must match the OEM PDF revision used on the project.