Product Identification
- Product Name: Memosens measuring cable and installation accessories
- Model Series: CYK10 / CYK11
Application
- Measuring cable for all Memosens sensors and sensors with Memosens technology for nearly all parts of process engineering.
- CYK10: Memosens data cable used with sensors utilizing contactless inductive signal transmission. Available in versions for use in hazardous areas.
- CYK11: Memosens extension cable for connecting digital fixed-cable sensors with Memosens protocol. Available versions include: Ferrule with ferrule, M12 female connector with ferrule, and M12 female connector with M12 plug. CYK10 based installations can also be extended with CYK11.
- Both CYK10 and CYK11 can be extended with junction boxes.
Your Benefits
- CYK10 Data Cable:
- Absolute waterproof connection (ingress protection IP 68)
- No corrosion of contacts
- No influence of interfering potentials thanks to galvanic isolation of transmitter and medium
- CYK11 Extension Cable:
- Absolute waterproof connection (ingress protection IP 68)
- Flexible and fast in application
Function and System Design
The Memosens system offers:
- Non-contact, digital signal transmission for optimum galvanic isolation.
- No contact corrosion.
- Completely watertight connections.
- Possibility for laboratory sensor calibration, increasing measured value availability.
- Predictive maintenance through recording of sensor data such as:
- Total hours of operation
- Hours of operation with very high or very low measured values
- Hours of operation with high temperatures
- Number of steam sterilizations
- Sensor condition
Measuring System Components
- Transmitter
- Memosens extension cable CYK11 (optional)
- Junction box (optional)
- Memosens data cable CYK10 or fixed cable
- Sensor
Connection in Explosion-Hazardous Areas
- It is not permitted to use CYK11 as an extension for CYK10-based measuring systems in hazardous areas.
- For hazardous areas, it is recommended to use a continuous and not extended Memosens measuring cable CYK10.
Cable Specification
Performance Characteristics
| Parameter | CYK10 | CYK11 |
|---|---|---|
| Conductor/Construction | 4 cores | 6 cores |
| Outer diameter (mm) | 6.3 | 6.3 |
| Shielding | Yes | Yes |
| Bending radius solid | 5xD | 5xD |
| Bending radius flexible | 10xD | 10xD |
| UV resistant | Yes | Yes |
| Clear of ozon-depleting substances | Yes | Yes |
| Halogen-free | No | Yes |
| RoHS-compliant | Yes | Yes |
| Flame-resistant | Yes | Yes |
| Oil-resistant | - | Yes |
Environment
| Parameter | CYK10 | CYK11 |
|---|---|---|
| Temperature min. | -25 °C (-13 °F) | -30 °C (-22 °F) |
| Temperature max. | 135 °C (277 °F) | 90 °C (194 °F) |
| Degree of protection | IP 68 (Memosens: 1 bar, 25 °C, 45 days; M12: 0.1 bar, 50 °C, 30 days) | IP 68 (M12: 0.1 bar, 50 °C, 30 days) |
Note: IP 68 values are specified for mated condition and specific media/conditions.
Mechanical Construction
- CYK10 and CYK11 Cable Sheath Material: TPE
M12 Plug Assignment
| Pin | CYK10 & CYK11 Signal | Color |
|---|---|---|
| 1 | V (+24 V) | PK |
| 2 | GND | GY |
| 3 | V (Memosens sensor) | BN |
| 4 | GND (Memosens sensor) | WH |
| 5 | RS 485 A | GN |
| 6 | RS 485 B | YE |
Junction Box
- Material: Aluminum, painted
Variants
- Junction box M12 connector/cable: Cable extension for Memosens sensors, Liquiline. Order no. 71145498.
- Junction box cable/cable: Cable extension for Memosens sensors, Liquiline. Order no. 71145499.
Certificates and Approvals
- Declaration of Conformity: Product meets requirements of harmonized European standards and EU guidelines. Marked with the 4 mark.
- Ex approval (CYK10):
- IECEx: Ex ia IIC T3/T4/T6 Ga
- ATEX: II 1G Ex ia IIC T3/T4/T6 Ga
- FM/CSA: IS/NI CLI Div1&2 GP ABCD
- Nepsi: II 1G Ex ia IIC T3/T4/T6 Ga
Ordering Information