Orbitex W CPS21

Engineering summary from PDF text extraction for Orbitex W CPS21. Verify every value with the OEM datasheet.

Summary

Product identification

  • Product Name: Orbitex W CPS21
  • Product Type: pH sensors with open ring junction for wastewater technology
  • Document: TI00029C/07/EN/12.22-00

Application

  • Municipal wastewater
  • Electroplating wastewater
  • Wastewater from metal processing plants

Your benefits

  • Resistant to blocking media, quick response time and reliable measurement thanks to open ring junction.
  • Can be used under pressures up to 6 bar (87 psi) without counterpressure thanks to solid Polytex electrolyte.
  • Long service life due to optimized diffusion path of 180 mm (7.09 in) and thus a long poison diffusion path of the reference system.
  • TOP68/ESA plug-in head, water proof (IP 68).

Measuring principle

pH measurement

The pH value is a measure of acidity or alkalinity. The sensor's membrane glass generates an electrochemical potential based on the medium's pH. An integrated Ag/AgCl reference system serves as the reference electrode. The measured voltage is converted to a pH value using the Nernst equation.

Measuring system

A complete measuring system typically includes:

  • CPS21 pH sensor
  • Transmitter (e.g., Liquiline CM42)
  • Measuring cable (e.g., CPK9)
  • Immersion or flow assembly (e.g., Dipfit CPA111)

Input

  • Measured variables: pH value, Temperature
  • Measuring range:
    • pH: 2 to 12
    • Temperature: –15 to 60 °C (5 to 140 °F)

Electrical

  • Energy supply: Memosens sensors, sensors with TOP68 plug-in head, sensors with GSA plug-in head. Specific connection details depend on the measuring cable (CYK10, CYK20, CPK9, CPK1).

Installation

  • Orientation: Sensors should not be installed upside-down.
  • Angle of inclination: Must be at least 15° from the horizontal to ensure the pH membrane is covered with inner electrolyte.
  • Assembly mounting: Hand-tighten sensor to 3 Nm (2.21 lbf ft) when installing in Endress+Hauser assemblies. Ensure threads, O-rings, and sealing surfaces are clean and undamaged.

Environment

  • Operating temperature range: –15 to 60 °C (5 to 140 °F)
  • Storage temperature: 0 to 50 °C (32 to 122 °F)
  • Degree of protection:
    • IP 67: GSA plug-in head (with closed connector system)
    • IP 68: TOP68 plug-in head, up to 135 °C (275 °F) (1 m (3.3 ft)) water column, 50 °C (122 °F), 168 h.
  • Electromagnetic compatibility (EMC): Interference emission and interference immunity as per EN 61326: 2012.

Process

  • Process temperature range: –15 to 60 °C (5 to 140 °F)
  • Process pressure range: 0 to 6 bar (0 to 87 psi)
  • Conductivity: Min. 500 µS/cm

Mechanical construction

  • Design, dimensions:
    • Overall length: 120 mm (4.70 in)
    • Diameter: ø 12 mm (0.47 in)
  • Weight: 0.1 kg (0.2 lb)
  • Materials:
    • Electrode shaft: Glass
    • pH membrane glass: Type A
    • Metal lead: Ag/AgCl
    • Diaphragm: Open ring junction
    • Temperature sensor: Integrated
    • Plug-in head:
      • ESA: Threaded plug-in head Pg 13.5, TOP68, 16 bar (232 psi) overpressure protection (threefold)
      • GSA: Threaded plug-in head Pg 13.5
  • Reference system: Ag/AgCl with Polytex gel 3M KCl, AgCl-free.
  • Process connections: Pg 13.5

Certificates and approvals

Available certificates and approvals can be selected via the Product Configurator on www.endress.com.

Ordering information

  • Product page: www.endress.com/cps21
  • Use the Product Configurator to configure and obtain a valid order code. CAD drawings may be available.

Scope of delivery

  • Ordered version of the sensor
  • Operating Instructions
  • Safety instructions (if Ex approval ordered)
  • Supplementary sheet for certificates (if ordered)

Accessories

  • Assemblies: Dipfit CPA111, Dipfit CPA140, Flowfit CPA240, Flowfit CYA27, Ecofit CPA640.
  • Buffer solutions: CPY20.
  • Measuring cable: CPK9, CPK12, CPK1.