Datasheet summary
OUSAF44
Engineering summary from PDF text extraction for OUSAF44. Verify every value with the OEM datasheet.
Summary
Product identification
- Product Name: OUSAF44
- Product Type: Optical sensor for UV absorption measurement
Application
The OUSAF44 sensor measures the spectral absorption of process liquids in the ultraviolet (UV) region. Applications include:
- Protein concentration measurement
- Chromatography monitoring
- Filtration monitoring
- Organic compound concentration measurement
- Aromatic compound detection
Operating limits
Environmental conditions
- Ambient temperature: 0 to 55 °C (32 to 131 °F)
- Storage temperature: -10 to +70 °C (+10 to +160 °F)
- Humidity: 5 to 95 %
- Degree of protection: IP 65 (NEMA 4) for all optical parts
Process conditions
- Process temperature: 0 to 90 °C (32 to 194 °F) continuous; Max. 130 °C (266 °F) for 2 hours
- Process pressure: Max. 100 bar (1450 psi) absolute, depending on material, pipe size, and process connection of the flow assembly.
Feed water
Measured variable
- UV-absorption
Measuring range
- 0 to 2.5 AU
- Max. 50 OD (depending on the optical path length)
Wavelength
- Discrete wavelengths at 254, 280, 295, 302, 313, or 365 nm. More wavelengths are available on request.
Product water
Not specified in the extract.
Electrical
- Power supply: Provided by associated transmitter via cable set.
- Electrical connection: Via pre-terminated or labeled cable set CUK80 (for CM44P) or OUK40 (for CVM40).
- Lamp voltage:
- 10.0 ± 0.1 V (when connected to CM44P)
- 11.9 ± 0.1 V (when connected to CVM40)
- Maximum cable length: 100 m (330 ft)
Mechanical / hydraulic
- Sensor housing material: Stainless steel 316
- Cable connector ends material: Nickel-plated brass
- Lamp type: Low-pressure mercury lamp
- Lamp operating life: Typically 3000 hours, minimum 1000 hours.
- Warm-up period: 30 minutes for the lamp to reach full capacity.
- Detector type: UV silicon detector, hermetically sealed.
- Filter type: Multilayer interference filter.
- Dimensions:
- Lamp module (A): 142 mm (5.60 inches)
- Standard detector (B): 60.2 mm (2.37 inches)
- Easycal detector (B): 69 mm (2.70 inches)
- Weight:
- UV lamp: 0.58 kg (1.28 lbs)
- UV lamp with wire-braided cable and junction box (hazardous area): 3.2 kg (6.66 lbs)
- Easycal detector: 0.65 kg (1.43 lbs)
- Standard detector: 0.36 kg (0.794 lbs)
- Flow assembly: OUA260 (dimensions and details in TI00418C)
- Hygienic design: Compliance with life sciences sector requirements, with FDA- and USP-compliant sealing materials. SIP/CIP resistant.
- Optional features: Air purge ports, adjuster for optical path length.
Hazardous area information
- Approvals: FM and ATEX approved lamps available.
- Detector: Intrinsically safe, suitable for Zone 1 and Class I, Division 1 environments when connected via a safety barrier (MTL7760AC).
- Lamp connection: Requires connection via a certified junction box for hazardous areas.
Calibration
- Integrated reference detector: For lamp compensation.
- Optional Patented Easycal system: Enables cost-effective, time-efficient, and in-line calibration traceable to NIST without liquid standards.
Disclaimer: This summary is extracted from a datasheet and may not include all figures, footnotes, or revisions. Contractual data must be confirmed against the specific revision of the OEM PDF used for the project.