Datasheet summary
Oxymax COS61
Engineering summary from PDF text extraction for Oxymax COS61. Verify every value with the OEM datasheet.
Summary
Product identification
- Product Name: Oxymax COS61D/COS61
- Product Type: Optical sensor for measuring dissolved oxygen
- Technology: Digital optical sensor based on fluorescence quenching principle.
- Protocols: With or without Memosens protocol.
- Model Variants:
- COS61D: Liquiline sensor, Memosens protocol, optionally with M12 plug.
- COS61: Liquisys sensor, compatible with COM2x3W and COM2x3D (with conversion kit).
Application
- Wastewater treatment plants (activated sludge basin oxygen regulation, outlet monitoring)
- Water monitoring (rivers, lakes, seas for water quality indication)
- Water treatment (status monitoring, e.g., drinking water enrichment, corrosion protection)
- Fish farming (oxygen measurement and regulation for optimal conditions)
Key Features and Benefits
- Optical Technology: Minimum maintenance, maximum availability.
- Digital Signal Processing:
- Calibration data saved in sensor.
- High EMC protection via digital communication.
- Extended maintenance intervals and long-term stability.
- Intelligent self-monitoring.
- Measurement: No flow required, possible in still water.
- Plug & Play: Safe communication based on Memosens protocol (COS61D).
- Compatibility: COS61 is compatible with COS31 and COS41.
Measuring Principle
The sensor uses oxygen-sensitive molecules (markers) integrated into an optically active layer. The measurement process relies on fluorescence quenching:
- Green light pulses are sent to the fluorescence layer.
- Markers fluoresce with red light pulses.
- The duration and intensity of the red light pulses are dependent on the oxygen content. Higher oxygen levels mask markers, resulting in shorter and less intense signals.
Measurement Parameters
- Measured Variables: Dissolved oxygen.
- Units: mg/l, µg/l, ppm, ppb, %SAT, or hPa.
- Compensation: Medium temperature is automatically recorded. Air pressure can be set statically or entered via an additional sensor.
Operating Limits
-
Measuring Ranges (with Liquiline CM44x or Liquisys COM2x3-W):
- 0 to 20 mg/l
- 0 to 400 hPa
- 0 to 200 % SAT (These ranges apply for 20 °C and 1013 hPa)
-
Process Temperature: -5 to +60 °C (20 to 140 °F)
-
Process Pressure: Ambient pressure 1 to 10 bar (14.5 to 145 psi) abs.
-
Ambient Temperature: -20 to 60 °C (-4 to 140 °F) at 95% relative air humidity, non-condensing.
-
Storage Temperature: -20 to 70 °C (-4 to 158 °F) at 95% relative air humidity, not condensing.
Electrical
- COS61D Connection: Fixed cable with terminated cores, or optional M12 plug. Connects directly to transmitter terminals or via M12 sensor socket.
- COS61 Connection: Fixed cable or TOP68 plug connection. Connects via SXP connector to field device or terminal connections for cabinet devices (e.g., Liquisys COM223).
- Maximum Cable Length: max. 100 m (330 ft), including extension.
- Interface:
- COS61D: Memosens protocol.
- COS61: RS 485.
Performance Characteristics
- Response Time (t90, from air to nitrogen at reference conditions): 60 s.
- Maximum Measured Error:
- COS61D: < 12 mg/l: 0.01 mg/l or ±1% of reading; 12 mg/l to 20 mg/l: ±2% of reading.
- COS61: < 12 mg/l: 0.02 mg/l or ±1% of reading; 12 mg/l to 20 mg/l: ±2% of reading.
- Repeatability: ±0.5% of end of measuring range.
- Operating Life of Sensor Cap: > 2 years (under reference conditions, protect from direct sunlight).
Mechanical / Hydraulic
-
Degree of Protection:
- COS61D: IP 68 (10 m water column, 25 °C, 30 days).
- COS61 (Fixed cable): IP 68 (10 m water column, 25 °C, 30 days).
- COS61 (Top68 plug): IP 68 (1 m water column, 50 °C, 7 days).
-
Materials (Parts in contact with medium):
- Sensor cap: PVC / POM
- Fluorescence layer: Silicone
- Orifice plate: PET
- O-rings: EPDM
- Pin holder: 1.4404
- Shaft tube: 1.4571
- Housing connection: POM
- Protection guard: POM
- Housing air purge unit: POM
-
Process Connection:
- COS61D: G1, NPT 3/4"
- COS61: G1
-
Weight:
- With 7 m cable: 0.7 kg (1.5 lbs)
- With 15 m cable: 1.1 kg (2.4 lbs)
- With TOP68 plug: 0.3 kg (0.66 lbs) depending on version.
-
Dimensions: See figures 12, 13, 14, 15, 16 in the original document.
Environment and Compatibility
-
Electromagnetic Compatibility (EMC):
- COS61D: EN 61326: 2005, Namur NE 21:2007.
- COS61: EN 61326: 1997 / A1: 1998.
-
Certificates and Approvals:
- Declaration of Conformity (CE mark).
- EAC (TP TC 004/2011, TP TC 020/2011).
- CSA GP (Class 2 power supply, Overvoltage category I, max. height 2000 m).
- CSAus NI Cl 1, Div 2 (Hazardous areas, specific installation requirements apply).
- Explosion Protection: Class 1, Division 2, Groups A, B, C, D T6; IP67/IP68.
Disclaimer: Extraction may miss figures, footnotes, or revisions; contractual data must match the OEM PDF revision used on the project.