Datasheet summary
Oxymax W COS41
Engineering summary from PDF text extraction for Oxymax W COS41. Verify every value with the OEM datasheet.
Summary
Product identification
Oxymax W COS41: Potentiostatic amperometric two-electrode sensor.
Application
The sensor is designed for the continuous measurement of dissolved oxygen in various water management areas, including:
- Sewage treatment plants: For oxygen measurement and regulation in activated sludge basins to ensure efficient biological cleaning.
- Water monitoring: As an indicator of water quality in rivers, lakes, and seas.
- Water treatment: For status monitoring of drinking water, such as oxygen enrichment or corrosion protection.
- Fish farming: For oxygen measurement and regulation to optimize living and growth conditions for fish.
Benefits
- Maximum measurement accuracy with long maintenance intervals and intelligent sensor self-monitoring.
- Membrane-covered design provides high O₂ selectivity, minimum maintenance, and minimal calibration effort.
Measuring principle
Dissolved oxygen molecules diffuse through a membrane and are reduced to hydroxide ions (OH⁻) at the cathode. At the anode, silver is oxidized to silver ions (Ag⁺), forming a silver halide layer. A current proportional to the oxygen content flows between the electrodes. This current is converted by a transmitter and displayed as oxygen concentration (mg/l, ppm), saturation index (% SAT), or partial pressure (hPa).
Measuring system
A complete measuring system typically includes:
- Oxygen sensor (Oxymax W COS41)
- Transmitter (e.g., Liquisys M COM223/253)
- Special measuring cable
- Assembly (e.g., flow assembly COA250, immersion assembly CYA611, or retractable assembly COA451)
Optional components include a universal suspension assembly support, junction box VBM, and automatic spray cleaning system Chemoclean.
Input
- Measured variable: Dissolved oxygen
- Units: mg/l, ppm, % SAT, or hPa
- Measuring range (with Liquisys M COM 223/253-DX/DS):
- 0.05 ... 20.00 mg/l (ppm)
- 0.00 ... 200 % SAT
- 0 ... 400 hPa
Wiring
The sensor can be connected directly to the transmitter using a special measuring cable. For cable extensions beyond the fixed cable length, a junction box VBM and the special measuring cable CYK71 are required.
Special measuring cable CYK71 assignments:
- Terminal S: Outer shield
- Terminal 12: Active inner shield (NTC)
- Terminal 90: Cathode
- Terminal 91: Anode
- Terminal 11: NTC temperature sensor
Performance characteristics
- Response time (T90): 3 minutes (at 20 °C / 68 °F)
- Response time (T99): 9 minutes (at 20 °C / 68 °F)
- Reference operating conditions: 25 °C (77 °F), 1013 hPa (15 psi)
- Signal current in air: approx. 300 nA
- Zero current: Zero current free
- Measured value resolution: 0.01 mg/l (0.01 ppm)
- Maximum measured error: ±1 % of measured value
- Long-term drift: < 1 % per month (with permanent polarization)
- Influence of medium pressure: Pressure compensation not necessary.
- Polarization time: < 60 minutes
- Oxygen intrinsic consumption: Approx. 90 ng/h in air at 25 °C (77 °F)
Installation
- Angle of installation: Permissible up to horizontal (0 ... 180°); overhead installation is not permitted.
Environment
- Ambient temperature range: –5 to 50 °C (20 to 120 °F)
- Storage temperature:
- Filled with electrolyte: –5 to 50 °C (20 to 120 °F)
- Without electrolyte: –20 to 60 °C (0 to 140 °F)
- Ingress protection:
- Fixed cable versions: IP 68 (10 m / 33 ft water column at 25 °C / 77 °F in 30 days)
- TOP68 plug-in head versions: IP 68 (1 m / 3.3 ft water column at 50 °C / 122 °F in 7 days)
Process
- Process temperature: –5 to 50 °C (20 to 120 °F)
- Process pressure: max. 10 bar (145 psi) permissible overpressure. Underpressure operation is not permissible.
Mechanical construction
- Weight:
- With cable length 7 m / 23 ft: 0.7 kg / 1.5 lbs.
- With cable length 15 m / 49 ft: 1.1 kg / 2.4 lbs.
- With TOP68 plug-in connection: 0.3 kg / 0.66 lbs.
- Material:
- Sensor shaft: POM
- Membrane cap: POM
- Cathode: Gold
- Anode/Reference electrode: Silver / silver bromide
- Process connection: G1 and NPT ¾"
- Maximum cable length: 50 m (164 ft)
- Membrane thickness: approx. 50 μm
- Electrolyte: Alkaline electrolyte
Scope of delivery
- Oxygen sensor with transport protection cap
- Accessories set: 2 replacement membrane caps, 10 electrolyte ampoules, 1 sealing kit (3 O-rings), 6 abrasive sheets.
- Operating Instructions (CD)
- Brief Operating Instructions (paper)
Accessories
- Assemblies: Flow assembly COA250, Immersion assembly Dipfit W CYA611.
- Zero solution: For producing oxygen-free solution.
- Measuring cable: COK41 (for TOP68 plug-in head), CYK71 (non-terminated, sold by the meter).
- Junction box: VBM (IP 65, aluminum).
- Transmitter: Liquisys M COM 223/253.
Disclaimer: This extraction may miss figures, footnotes, or revisions. Contractual data must match the OEM PDF revision used on the project.
Official datasheet (PDF)
Curated from selected public technical reference material for discovery and preliminary comparison. This summary is not a substitute for a current certified manufacturer datasheet. Verify revisions and design limits before use.