Datasheet summary
Smartec CLD134
Engineering summary from PDF text extraction for Smartec CLD134. Verify every value with the OEM datasheet.
Summary
Product identification
Product: Smartec CLD134 Type: Hygienic measuring system for conductivity and concentration. Application Industries: Food, beverage, pharmaceutical, and biotechnology.
Application Areas
- Phase separation (product/water, product/product) in pipe systems.
- Control of Cleaning In Place (CIP) processes in return lines.
- Concentration control for CIP cleaning agent remaking.
- Product monitoring in pipes, bottling plants, and quality assurance.
- Leakage monitoring.
Measuring Principle
Inductive conductivity measurement using an oscillator, primary coil, induced current flow in the medium, secondary coil, and receiver to determine conductivity.
Advantages of Inductive Measurement:
- No electrodes, preventing polarization effects.
- Accurate measurement in heavily polluted media prone to buildup.
- Complete galvanic isolation between the measurement and the medium.
Key Features
- Hygienic Design: Injection-molded PEEK sensor, seamless and crevice-free. Virgin PEEK in contact with the medium ensures biological safety. Designed according to ASME BPE guidelines and meets EC Regulation No. 1935/2004.
- Certified Hygiene: FDA-listed materials for medium contact, 3-A standards certification, EHEDG certification (CLS54 sensor for cleanability), optional USP class VI biological reactivity testing with lot traceability.
- Process Connections: Available with commonly used hygienic connections, with custom options on request.
- Process Temperature & Pressure:
- Continuous temperatures up to 125 °C (257 °F).
- Temporary sterilization temperatures up to 150 °C (302 °F) (max. 60 minutes).
- Pressure resistance up to 12 bar (174 psi) abs at temperatures up to 90 °C (194 °F).
- Suitable for underpressure applications.
- Temperature Measurement: Integrated temperature sensor with response time t90 < 26 s. Embedded without seals for long service life.
- Temperature Compensation: Options include linear (user-configurable coefficient α), IEC 60746-3 (NaCl), and user-programmable coefficient tables (up to 10 elements).
- Concentration Measurement: Switchable from conductivity mode. Offers user-programmable and pre-programmed concentration curves (e.g., for CIP solutions), displaying concentration directly as %.
- Remote Parameter Set Configuration (MRS): Allows for measurement range switching, adjustment of temperature compensation, and switching between concentration curves.
- Device Versions: Available as compact or remote systems. Transmitter can be ordered separately.
Measured Values
- Conductivity
- Concentration
- Temperature
Measuring Ranges
- Conductivity: Recommended range: 100 μS/cm to 2000 mS/cm (uncompensated).
- Concentration:
- NaOH: 0 to 15 %
- HNO₃: 0 to 25 %
- H₂SO₄: 0 to 30 %
- H₃PO₄: 0 to 15 %
- User-defined: Up to 4 tables available with MRS function.
- Temperature: -35 to +250 °C (-31 to +482 °F) (Pt 1000).
Electrical Specifications
- Input:
- Binary inputs: 10 to 50 V, max. 10 mA at 50 V.
- Output:
- Conductivity/Concentration: 0/4 to 20 mA, galvanically isolated.
- Temperature (optional second current output).
- Signal on alarm: 2.4 or 22 mA.
- Load: Max. 500 Ω.
- Separation voltage: Max. 350 VRMS / 500 V DC.
- Minimum output signal spread:
- Conductivity: 200 μS/cm to 200 mS/cm depending on range.
- Temperature: 15 °C or 27 °F.
- Auxiliary voltage output: 15 V ± 0.6 V, max. 10 mA.
- Contact outputs: Max. 2 A switching current, max. 250 V AC / 30 V DC switching voltage.
- Switching power: Max. 500 VA AC / 60 W DC (ohmic load).
- Power Supply:
- Versions: 100/115/230 V AC ±10/-15 %, 48-62 Hz or 24 V AC/DC +20/-15 %.
- Power consumption: Max. 7.5 VA.
- Mains fuse: 250 V/3.15 A (semi-delay).
- Electrical Connection: Terminals for signal outputs, auxiliary voltage, binary inputs, and power supply.
Performance Characteristics
- Measured value resolution: Temperature: 0.1 °C.
- Response time:
- Conductivity: t95 < 1.5 s.
- Temperature: t90 < 26 s.
- Measured error:
- Sensor (Conductivity): ± (0.5 % of reading + 10 μS/cm) after calibration.
- Sensor (Temperature): Pt 1000 Class A (IEC 60751).
- Transmitter (Conductivity): Max. 0.5 % of measured value ± 4 digits (display); Max. 0.75 % of current output range (output).
- Transmitter (Temperature): Max. 0.6 % of measuring range (display); Max. 0.75 % of current output range (output).
- Repeatability: Conductivity: Max. 0.2 % of measured value ± 2 digits.
- Cell constant: 6.3 cm⁻¹.
- Measuring frequency: 2 kHz.
- Temperature compensation range: -10 to +150 °C (14 to 302 °F).
- Reference temperature: 25 °C (77 °F).
Installation
- Sensor must be fully immersed; avoid air bubbles.
- Install at least 1 m (3.3 ft) downstream from pipe bends.
- Installation factor: Accounts for wall effects in confined installations; can be entered in the transmitter. Factor can be disregarded if wall distance
a > 15 mm(from DN 65). - Air set: Must be performed before installation to compensate for residual coupling.
Environmental Conditions
- Ambient temperature:
- Compact version/electronics: 0 to +55 °C (32 to 131 °F).
- Sensor (remote): -20 to +60 °C (-4 to 140 °F).
- Storage temperature: -25 to +70 °C (-13 to 158 °F).
- Electromagnetic compatibility: As per EN 61326-1 and EN 61326-2-3.
- Degree of protection: IP67 / Type 4.
- Relative humidity: 10 to 95 % (non-condensing).
- Vibration resistance: 10 to 500 Hz, 0.15 mm deflection, 19.6 m/s².
Process Conditions
- Process temperature:
- Remote version: Max. 125 °C (257 °F) at 70 °C ambient.
- Compact version: Max. 125 °C (257 °F) at 35 °C ambient; Max. 55 °C (131 °F) at 55 °C ambient.
- Sterilization: Max. 150 °C (302 °F) for max. 60 minutes (conditions vary by version and ambient temperature).
- Absolute process pressure:
- Up to 90 °C (194 °F): 13 bar (188.5 psi) abs.
- At 125 °C (257 °F): 9 bar (130.5 psi) abs.
- CRN environment: 1 to 6 bar (14.5 to 87 psi) abs.
- Negative pressure: Down to 0.1 bar (1.45 psi) absolute.
- Flow velocity: Max. 5 m/s (16.4 ft/s) for low-viscosity media in DN65 pipes.
Mechanical Construction
- Dimensions: See drawings for remote and compact versions.
- Remote version (with mounting plate): 225 x 142 x 109 mm.
- Compact version (MV5, CS1, AA5, SMS): 225 x 142 x 255 mm.
- Compact version (VA4, BC5): 225 x 142 x 213 mm.
- Weight:
- Remote transmitter: Approx. 2.5 kg (5.5 lb).
- CLS54 sensor: 0.3 to 0.5 kg (0.66 to 1.1 lb).
- Compact version: Approx. 3 kg (6.6 lb).
- Materials:
- CLS54 sensor (medium contact): Virgin PEEK.
- CLS54 sensor (other): PPS-GF40, Stainless steel 1.4404 (AISI 316L), FKM, EPDM.
- Transmitter housing: Stainless steel 1.4301 (AISI 304).
- Front window: Polycarbonate.
- Chemical Resistance (PEEK sensor): See table for specific concentrations and temperatures of NaOH, HNO₃, H₃PO₄, H₂SO₄, and Peracetic acid.
- Process Connections:
- Sanitary connection DIN 11851 (DN 50)
- Aseptic coupling DIN 11864-1 form A
- Clamp ISO 2852 (2")
- SMS coupling 2"
- Varivent N (DN 40 - 125)
- NEUMO BioControl D50 (DN 40, 2")
- Additional connections available on request.
Operability
- Display and operating elements: LC display, four operating keys, LED for alarm.
- Operation: Via keypad, HART handheld terminal, PC with FieldCare, or PROFIBUS PA/DP.
- Menu structure: Clear menu with access code protection.
Certificates and Approvals
- Hygiene: FDA listed materials, EHEDG Type EL - Class I certified, 3-A Standard 74- certified.
- Biological Reactivity: Optional USP class VI certification.
- EC Regulation: Meets requirements of EC Regulation No. 1935/2004.
- Pressure Approval: Canadian pressure approval (ASME B31.3).
Accessories
- Cable extension: CLK6 measuring cable, VBM junction box.
- Desiccant pouch: For VBM junction box.
- Post mounting kit: For remote version mounting.
- Software upgrade: Function upgrade for MRS and temperature coefficient determination (Order No. 51501643).
- Optoscope: Interface for service purposes with Scopeware software (Order No. 51500650).
- Calibration solutions: CLY11 series (various conductivity values).
Disclaimer: This extraction may miss figures, footnotes, or revisions. Contractual data must match the OEM PDF revision used on the project.