Datasheet summary
TOC Analyzer CA78
Engineering summary from PDF text extraction for TOC Analyzer CA78. Verify every value with the OEM datasheet.
Summary
Product Identification
- Product Name: TOC Analyzer CA78
- Document Title: TOC Analyzer CA78
- Technical Information Document ID: TI01622C/07/EN/02.22-00
- Document Date: 2022-06-30
- Document Number: 71575760
Application
The analyzer is designed for the determination of total organic carbon (TOC) in ultrapure water applications with the following conditions:
- Conductivity < 10 µS/cm
- pH range: neutral
Measuring Principle
The TOC analyzer utilizes UV digestion to oxidize organic substances into CO₂. The resulting increase in conductivity due to the formation of hydrogen carbonate is measured by a pair of electrodes upstream and downstream of the UV radiation, which is then converted to TOC.
Performance Characteristics
- Measured variable: TOC (total organic carbon)
- Limit of detection (LOD): 0.1 µg/l (ppb)
- Response time t90: 50 s
- Number of measuring channels: 1 to 3, depending on the order version
Operating Limits
- Sample temperature: < 50 °C (122 °F)
- Process pressure: Max. 0.5 bar (7.25 psi); recommended 0.25 bar (3.62 psi)
- Maximum conductivity of sample: Max. 2 µS/cm (Order option: max. 10 µS/cm)
Feed Water Requirements
- Sample requirement: 14 ml/min.
- Sample quality: Particle-free
- Maximum conductivity of sample: Max. 2 µS/cm (Order option: max. 10 µS/cm)
Product Water
Not specified in the extract.
Electrical
- Supply voltage: 100/240 V AC, 47 - 63 Hz
- Power consumption: Max. 60 W
- Output signal (Measuring channel 1): 0/4 to 20 mA, galvanically isolated
- Output signal (Measuring channel 2, optional): 0/4 to 20 mA, galvanically isolated
- Load (Current outputs): Max. 500 Ω
- Transmission behavior: Configurable, in the measuring range of 4 to 20 mA (Standby: 3.8 mA)
- Current outputs, active: 0 to 20 mA; according to Namur NE43
Mechanical / Hydraulic
- Dimensions: 200 mm (width) x 500 mm (height) x 290 mm (depth) (approx. 7.87 in x 19.68 in x 11.42 in)
- Weight: Approx. 14 kg (30.86 lb)
- Materials: Stainless steel housing
- Hose specification: Sample hose 1/8 inch, 3.2 mm OD included in connection kit
Environment
- Ambient temperature range: 10 to 45 °C (50 to 113 °F)
- Storage temperature: 2 to 55 °C (35 to 131 °F)
- Relative humidity: 10 to 90 %, non-condensating
- Degree of protection: IP 42 (standard device), IP54 (order option)
- Pollution degree: 2
Operability
- Display: 8" touch screen monitor
- Operating language: English (US)
Maintenance
- Maintenance effort: 1 hour per month
- Maintenance intervals:
- Replacement of calibration solution - before every calibration
- Replacement of pump hose - every 6 months
- Replacement of UV reactor - every 6 months
- Replacement of UV reactor ballast - every 24 - 36 months
- Replacement of pump head - every 36 - 48 months
- Calibration interval: Recommended after replacing components in contact with the process.
Disclaimer: This extraction may miss figures, footnotes, or revisions. Contractual data must match the OEM PDF revision used on the project.