Datasheet summary
TOC Analyzer CA79
Engineering summary from PDF text extraction for TOC Analyzer CA79. Verify every value with the OEM datasheet.
Summary
Product identification
- Product Name: TOC Analyzer CA79
- Manufacturer: Endress+Hauser Liquid Analysis
- Document Title: TOC Analyzer CA79
- Technical Information (TI) Number: TI01623C/07/EN/02.22-00
- Document Date: 2022-06-30
- Article Number: 71575761
Application
This analyzer is designed for determining total organic carbon (TOC) in ultrapure water applications that meet the following conditions:
- Conductivity < 2 µS/cm
- pH range: neutral
Measuring principle
The analyzer uses UV digestion to oxidize organic substances into CO₂. The resulting increase in conductivity due to CO₂ formation is measured by upstream and downstream sensors. This conductivity change is then converted into a TOC value.
Performance characteristics
- Measured Variable: TOC (total organic carbon)
- Limit of Detection (LOD): 0.1 µg/l (ppb)
- Response Time (t90): 50 s
- Number of measuring channels: 1 to 3, depending on the order version
- Calibration Interval: Calibrated on delivery. Recalibration recommended after replacing components in contact with the process (e.g., pump hose, UV reactor).
Feed water
- Maximum conductivity of sample: 2 µS/cm
- Particle content: Particle-free
- Sample temperature: < 50 °C (122 °F)
- Process pressure: Max. 0.5 bar (7.25 psi); recommended 0.25 bar (3.62 psi)
Product water
- Sample outlet: Depressurized
Electrical
- Supply voltage: 100/240 V AC, 47 - 63 Hz
- Power consumption: Max. 60 W
- Output signal (Measuring channel 1): 0/4 to 20 mA, galvanically isolated
- Output signal (Measuring channel 2, optional): 0/4 to 20 mA, galvanically isolated
- Load (Current outputs): Max. 500 Ω
- Controller input (order option): 24 V (for starting measurement, 1-channel devices only)
Mechanical / hydraulic
- Dimensions: 500 (19.68) x 200 (7.87) mm (Height x Width)
- Weight: Approx. 14 kg (30.86 lb)
- Materials: Stainless steel housing
- Hose specification: Sample hose 1/8 inch, 3.2 mm OD included in connection kit
- Sample flow rate: 14 ml/min.
Environment
- Ambient temperature range: 10 to 45 °C (50 to 113 °F)
- Storage temperature: 2 to 55 °C (35 to 131 °F)
- Relative humidity: 10 to 90 %, non-condensating
- Degree of protection: IP 42 (standard), IP54 (order option)
- Electromagnetic compatibility: EN 61326-1:2013, Class A for Industry
- Electrical safety: EN/IEC 61010-1:2010, Class 1 equipment
- Pollution degree: 2
Operability
- Display: 8" touch screen monitor
- Operating language: English (US)
- Data security: Software meets requirements of 21 CFR, Part 11.
Maintenance
- Maintenance effort: 1 hour per month
- Pump hose replacement: every 6 months
- UV reactor replacement: every 6 months
- UV reactor ballast replacement: every 24 - 36 months
- Pump head replacement: every 36 - 48 months
Disclaimer: This extraction may miss figures, footnotes, or revisions. Contractual data must match the OEM PDF revision used on the project.