Summary

Product Identification

  • Product Name: Unifit CPA842
  • Type: Hygienic process assembly for safe installation of 12 mm sensors (Pg 13.5)
  • Intended Use: Food, beverage, and life sciences industries.

Application

  • Permanently installed assembly for 12 mm sensors (e.g., pH, conductivity, ISFET pH, oxygen).
  • Suitable for CIP (cleaning in place) and SIP (sterilization in place).
  • Versions in stainless steel 1.4435 (AISI 316 L) and Alloy C22 are autoclavable.

Your Benefits

  • Robust assembly meeting 3-A standard 74-05, EHEDG, and ASME BPE criteria.
  • Suitable for in-place cleanability and in-line steam sterilizability.
  • Flush-mounted, gap-free electrode seal.
  • Electropolished surface available with Ra = 0.38 μm or 0.76 μm (stainless steel 1.4435 (AISI 316 L)).
  • Hygienic process connections with various versions for adaptation.

Measuring System

A complete measuring system comprises:

  • Unifit CPA842 assembly
  • 12 mm sensor (120 mm shaft length), e.g., Memosens CPS61E, CLS82E, or COS81E
  • Transmitter, e.g., Liquiline CM44x
  • Measuring cable, e.g., CYK10

Dependability

  • Manufactured to Endress+Hauser's certified quality criteria.
  • Made from high-grade materials: Stainless steel 1.4435 as per BN2 (AISI 316 L).
  • Assembly can be cleaned and sterilized inline in accordance with EHEDG criteria.

Mechanical Construction

Dimensions

DimensionWithout sensor protectionWith sensor protection (PAL)With sensor protection (cover)
L1 (mm/in)110 (4.33)--
L2 (mm/in)-137.5 (5.41)351 (13.81)
D (Diameter)25 (1)25 (1)61 (2.40)

Design

  1. Process connection
  2. Coupling nut
  3. O-ring
  4. Sensor protection
  5. Process connection
  6. Molded seal

Process Connections

  • DN25 standard
  • DN25 B.Braun port
  • Clamp 1.5" short (OD 50.5 mm / 1.99 in)
  • Clamp 1.5" long (OD 50.5 mm / 1.99 in)
  • Clamp 2" (OD 64 mm / 2.52 in)
  • Clamp 1.5" angled at 15° (OD 50.5 mm / 1.99 in)
  • Dairy fitting DN50 DIN 11851
  • Aseptic DN50 threaded DIN11864-1A
  • Varivent flange N (DN40 to 125)

Immersion Depth

Process connectionFeatureX0 (mm/in)X1 (mm/in)X2 (mm/in)D (mm/in)XP (mm/in)X3 (mm/in)
DN25 standardAA37.5 (1.46)61 (2.4)65 (2.6)25 (1)11 (0.43)29 (0.1)
DN25 B.Braun portAB57 (2.24)80.5 (3.17)84.5 (3.33)25 (1)11 (0.43)49 (0.16)
Clamp 1.5" shortAC6 (0.24)29.5 (1.16)33.5 (1.32)25 (1)7 (0.27)-
Clamp 1.5" longAD39 (1.53)62.5 (2.46)66.5 (2.61)25 (1)7 (0.27)-
Clamp 2"AE59 (2.23)82.5 (3.25)86.5 (3.4)25 (1)6 (0.24)-
Clamp 1.5" angledAF17.8 (0.7)41.3 (1.63)--25 (1)6 (0.24)-
Dairy fitting DN50AG41 (1.61)64.5 (2.53)68.5 (2.7)25 (1)19.5 (0.77)-
Aseptic DN50AK41 (1.61)64.5 (2.53)68.5 (2.7)25 (1)19.5 (0.77)-
Varivent N 68mmAH6 (0.24)29.5 (1.16)33.5 (1.32)25 (1)16.5 (0.65)-

Leakage Monitoring

Leakage monitoring is available as an optional order specification for 3-A and EHEDG variants or can be ordered separately. It indicates potential molded seal defects.

Weight

  • Assembly (various process connections): 0.3 to 1.4 kg (0.7 to 3.1 lbs)
  • Protection cover: approx. 0.2 kg (0.4 lbs)

Materials

  • In contact with medium:
    • Seals: EPDM, FKM, Silicone, FFKM (all FDA-compliant, USP Class VI)
    • Assembly: Stainless steel 1.4435 (AISI 316 L) with surface roughness Ra ≤ 0.76 µm or Ra ≤ 0.38 µm.
  • Not in contact with medium: Stainless steel 1.4308 (AISI 304H) or 1.4404 (AISI 316L), PP137 conductive (protection cover).
  • Lubricant: Klüber Paraliq GTE703 USP87 Class VI, FDA 21CFR 178.3570.

Environment

  • Ambient temperature: –15 to 70 °C (5 to 158 °F)
  • Storage temperature: –15 to 70 °C (5 to 158 °F)

Process

  • Process temperature: –15 to 140 °C (5 to 280 °F) (Pay attention to electrode specification)
  • Pressure: 16 bar (232 psi) up to 140 °C (284 °F) (Pay attention to electrode specification)
  • Flow velocity: Excessive flow velocities may damage sensors. Avoid cavitation: flow velocities should be < 7.5 m/s (24.6 ft/s) at 1 bar and 20°C (68°F).

Certificates and Approvals

  • ADI-free
  • Biological reactivity (USP Class VI)
  • EHEDG (optional)
  • ASME BPE
  • FDA
  • Regulation (EC) 1935/2004
  • 3-A Sanitary (optional)
  • 3.1 Inspection certificate (optional)
  • cGMP
  • DGRL-2014/68/EU / PED-2014/68/EU
  • CRN (optional)

Ordering Information

  • Detailed configuration is available via the Product Configurator on the Endress+Hauser website.
  • CAD and 2D drawings may be downloadable.

Scope of Delivery

  • Assembly in the ordered version
  • Sensor seal (mounted)
  • Process seal (mounted) for DN25 standard and DN25 B. Braun port
  • Dust caps
  • Operating Instructions

Accessories

Installation accessories

  • Welding socket, straight (DN25 standard, DN25 B. Braun)
  • Welding socket, angled (DN25 standard, DN25 B. Braun)

Seals

  • Kits for EPDM, FKM, FFKM, and Silicone seals.

Sensors (selection)

A range of Memosens sensors are compatible, including:

  • pH sensors (CPS61E, CPS11E, CPS12E, CPS41E, CPS76E, CPS16E, CPS96E, CPS77D)
  • Oxygen sensors (COS81E, COS22E)
  • Conductivity sensors (CLS82E)
  • NIR absorption sensor (OUSBT66)