Summary

Product Identification

The Cerabar PMP23 is a pressure transducer designed for measuring absolute and gauge pressure in gases, vapors, liquids, and dust, particularly in applications requiring hygienic standards. It is suitable for international use due to a wide range of approvals and process connections.

Key Features:

  • High reproducibility and long-term stability
  • Reference accuracy: up to 0.3%
  • Customized measuring ranges
  • Turn down ratio up to 5:1
  • Sensor for measuring ranges up to 40 bar (600 psi)
  • Housing and process isolating diaphragm made of 316L
  • IP69 protection rating
  • Fully-welded process connections
  • Suitable for CIP/SIP cleaning
  • Optional IO-Link communication

Operating Limits

  • Maximum Working Pressure (MWP): Max. 0 to +160 bar (0 to +2400 psi)
  • Over Pressure Limit (OPL): Max. 0 to +160 bar (0 to +2400 psi) (depends on measuring range)
  • Process Temperature Range (at process connection): –10 to +100 °C (+14 to +212 °F), with a maximum of +135 °C (+275 °F) for one hour during SIP cleaning.
  • Ambient Temperature Range:
    • General: –40 to +85 °C (–40 to +185 °F)
    • For hazardous areas: –40 to +70 °C (–40 to +158 °F)
    • With IO-Link: –40 to +70 °C (–40 to +158 °F)
  • Storage Temperature Range: –40 to +85 °C (–40 to +185 °F)

Feed Water / Process Media

  • Measured Variable: Gauge pressure or absolute pressure
  • Suitable Media: Gases, vapors, liquids, and dust.

Product Water / Output

  • Output Signal:
    • 4 to 20 mA (2-wire)
    • IO-Link: 4 to 20 mA (3-wire or 4-wire)
    • IO-Link (SSP Ed. 2 V1.1), 4 to 20 mA
  • Signal Range (4 to 20 mA): 3.8 to 20.5 mA
  • Signal on Alarm (4 to 20 mA): Factory setting for MAX alarm is >21 mA. Complies with NAMUR NE 43.
  • Dynamic Behavior:
    • Dead time (t₁): 6 ms (analog), 7 ms (IO-Link)
    • Time constant (T₆₃), t₂: 10 ms (analog), 11 ms (IO-Link)
    • Time constant (T<0xE2><0x82><0x90>₀), t₃: 15 ms (analog), 16 ms (IO-Link)
  • Switch Output Response Time: ≤ 20 ms

Electrical

  • Supply Voltage: 10 to 30 V DC (IO-Link requires minimum 18 V DC)
  • Current Consumption:
    • 4 to 20 mA output: ≤ 26 mA
    • IO-Link: Maximum ≤ 300 mA
  • Overvoltage Protection: Meets requirements of EN 61000-4-5 (1kV wire/ground).
  • Electrical Connections: M12 plug (IP65/67 or IP66/69), Valve plug (ISO4400 M16 or NPT ½). Cable connections available (5m, 10m, 25m).
  • Degree of Protection: IP66/68 NEMA Type 4X/6P (cable), IP65 NEMA Type 4X (valve plug), IP65/67 NEMA Type 4X (M12 plug), IP66/69 NEMA Type 4X (metal M12 plug).

Mechanical / Hydraulic

  • Housing Material: Stainless steel 316L
  • Process Isolating Diaphragm Material: 316L (1.4435)
  • Process Connections Material: 316L (1.4404 or 1.4435)
  • Process Connections Available: Thread ISO 228 G, Thread M24 x 1.5, DIN 11851, SMS, Clamp ISO 2852, Tri-Clamp ISO 2852, Varivent.
  • Flush Mount Diaphragm Diameter: 17.2 mm (0.68 in)
  • Weight: Varies based on electrical connection and process connection. For example, M12 plug (plastic housing cap) is 0.012 kg (0.03 lbs), and a G 1" A thread connection is 0.270 kg (0.60 lbs).
  • Cleaning: Cleaned from oil and grease (option HA). Suitable for CIP/SIP.

Performance Characteristics

  • Reference Accuracy: Up to 0.3% of calibrated span.
    • Non-linearity: ±0.3%
    • Non-repeatability: ±0.1%
  • Turn Down: Up to 5:1 (preset at factory, not changeable).
  • Long-term Stability:
    • Analog/IO-Link: ±0.2% of URL (1 year), ±0.4% of URL (5 years), ±0.45% of URL (8 years).
  • Thermal Change: Varies with temperature range and pressure range. For example, from –20 to +85 °C (–4 to +185 °F): <1% of calibrated span for TD 1:1 on <1 bar ranges.
  • Resolution: Minimum 1.6 μA for current output.
  • Switch-on Time: ≤ 2 s.

Environment

  • Degree of Protection: Varies by connection type, up to IP66/69.
  • Vibration Resistance: Guaranteed for 5 to 2000 Hz: 0.05g²/Hz (IEC 60068-2-64).
  • Electromagnetic Compatibility (EMC): Interference emission as per EN 61326-1 equipment B, Interference immunity as per EN 61326-1 (industrial environment).