Datasheet summary
Cerabar M PMP51B
Engineering summary from PDF text extraction for Cerabar M PMP51B. Verify every value with the OEM datasheet.
Summary
Product identification
- Product: Cerabar PMP51B
- Type: Digital pressure transmitter with metallic process membrane
Applications
- Pressure measuring ranges: up to 400 bar (6,000 psi)
- Process temperatures: up to 400 °C (752 °F) with diaphragm seal
- Accuracy: up to ±0.055%
- Process pressure and level measurement in liquids or gases
Performance characteristics
- Reference operating conditions: As per IEC 62828-2, with constant ambient temperature, humidity, and ambient pressure.
- Total performance: Calculated based on reference accuracy and ambient temperature effect.
- Reference accuracy: Varies by measuring cell and turn down (TD) ratio. For example:
- Standard device, 400 mbar measuring cell, TD 1:1: ±0.075%
- Platinum device, 400 mbar measuring cell, TD 1:1: ±0.055%
- Temperature effect (E2):
- E2M - Main temperature error
- E2E - Electronics error (0.2% for analog output, 0% for digital outputs)
- Resolution: Current output: < 1 μA
- Long-term stability:
- 1 year: ±0.10%
- 5 years: ±0.20%
- 10 years: ±0.25%
- Response time T63 and T90:
- Standard device (without diaphragm seal), analog electronics: Dead time max. 50 ms, T63 max. 40 ms, T90 max. 90 ms.
- Standard device (without diaphragm seal), HART electronics: Dead time max. 50 ms, T63 max. 85 ms, T90 max. 200 ms.
- Warm-up time (according to IEC62828-4): ≤ 5 s
Input
Measured variable
- Absolute pressure
- Gauge pressure
Measuring range
- Absolute pressure: Measuring cells range from 400 mbar (6 psi) to 400 bar (6,000 psi).
- Maximum working pressure (MWP) up to 400 bar (6,000 psi).
- Overpressure limit (OPL) up to 600 bar (9,000 psi).
- Gauge pressure: Measuring cells range from 400 mbar (6 psi) to 400 bar (6,000 psi).
- Maximum working pressure (MWP) up to 400 bar (6,000 psi).
- Overpressure limit (OPL) up to 600 bar (9,000 psi).
Output
- Output signal:
- 4 to 20 mA analog, 2-wire
- 4 to 20 mA with HART communication protocol, 2-wire
- PROFINET with Ethernet-APL: 10BASE-T1L, 2-wire 10 Mbit
- Signal on alarm: In accordance with NAMUR recommendation NE 43.
- 4 to 20 mA analog: Signal over-range > 20.5 mA, Signal under-range < 3.8 mA.
- 4 to 20 mA HART: Max alarm (settable 21.5 to 23 mA), Min alarm < 3.6 mA (factory setting).
- PROFINET with Ethernet-APL: According to "Application layer protocol for decentralized periphery", Version 2.4; diagnostics according to PROFINET PA Profile 4.02.
- Load:
- Analog/HART: Load resistance (RLmax) up to 1065 Ω.
- Supply voltage: 10.5 to 35 VDC.
Power supply
- Analog/HART:
- Ex d, Ex e, non-Ex: 10.5 to 35 VDC
- Ex i: 10.5 to 30 VDC
- PROFINET over Ethernet-APL: 9.6 to 15 VDC, 540 mW.
Mechanical construction
- Materials in contact with process: See the official PDF for exact tabulated values.
- Materials not in contact with process: See the official PDF for exact tabulated values.
- Weight: See the official PDF for exact tabulated values.
Mounting
- Orientation:
- A position-dependent zero point shift can be corrected.
- Diaphragm seals can also shift the zero point depending on installation position.
- Use of shutoff devices and/or water pocket pipes is recommended.
- Installation instructions:
- Standard devices without diaphragm seals mounted similar to pressure gauges (DIN EN837-2).
- Adjust housing and local display for optimal readability.
- Mounting brackets available for pipes or walls.
- Use flushing rings for flanges if buildup or clogging is expected.
- Prevent moisture penetration into the housing during mounting, connection, and operation.
- Route cables and plugs downwards.
Environment
- Ambient temperature range: See the official PDF for exact tabulated values.
- Storage temperature: See the official PDF for exact tabulated values.
- Operating altitude: See the official PDF for exact tabulated values.
- Climate class: See the official PDF for exact tabulated values.
- Atmosphere: See the official PDF for exact tabulated values.
- Degree of protection: See the official PDF for exact tabulated values.
Certificates and approvals
- CE mark
- RCM-Tick marking
- Ex approvals
- Corrosion test
- EAC conformity
- Drinking water approval
- Functional safety SIL/ IEC 61508 Declaration of Conformity (optional)
- Marine approval
- Radio approval
- CRN approval
- Pressure Equipment Directive 2014/68/EU (PED)
- Oxygen application
- China RoHS symbol
- RoHS
- Certification PROFINET with Ethernet-APL
Disclaimer: This extraction may miss figures, footnotes, or revisions. Contractual data must match the OEM PDF revision used on the project.
Official datasheet (PDF)
Curated from selected public technical reference material for discovery and preliminary comparison. This summary is not a substitute for a current certified manufacturer datasheet. Verify revisions and design limits before use.