Deltabar S PMD78B
Engineering summary from PDF text extraction for Deltabar S PMD78B. Verify every value with the OEM datasheet.
Summary
Product identification
- Product Name: Deltabar PMD78B
- Type: Digital differential pressure transmitter with metallic process membrane
- Application: Differential pressure, level, and flow measurement in liquids or gases.
Operating limits
- Pressure measuring ranges: up to 40 bar (600 psi)
- Process temperatures: up to 400 °C (752 °F) with diaphragm seal
- Static pressure: up to 160 bar (2 400 psi)
- Accuracy: up to ±0.075%
Feed water
- Measured variable: Differential pressure
Product water
- Not applicable for this product.
Electrical
- Output signal:
- 4 to 20 mA with HART communication protocol (2-wire)
- PROFINET with Ethernet-APL: 10BASE-T1L, 2-wire, 10 Mbit/s
- Supply voltage:
- Analog/HART (Ex d, Ex e, non-Ex): 10.5 to 35 VDC
- Analog/HART (Ex i): 10.5 to 30 VDC
- PROFINET with Ethernet-APL: 9.6 to 15 VDC (540 mW)
- Signal on alarm: According to NAMUR recommendation NE 43 for HART; according to "Application layer protocol for decentralized periphery" Version 2.4 for PROFINET.
Mechanical / hydraulic
- Measuring principle: Differential pressure measuring cell with metallic membrane, utilizing fill fluid and a resistance bridge (semiconductor technology).
- Diaphragm seal options: Single-sided or double-sided with temperature isolators and/or capillaries.
- Materials in contact with process: See the official PDF for exact tabulated values.
- Weight: See the official PDF for exact tabulated values.
- Dimensions: See the official PDF for exact tabulated values.