Back to Endress + Hauser temperature measurement
Endress+Hauser logo

Datasheet summary

iTemp TMT127

Engineering summary from PDF text extraction for iTemp TMT127. Verify every value with the OEM datasheet.

Summary

Product identification

  • Product name: iTEMP® Pt100 TMT127
  • Description: Pt100 Temperature Transmitter for DIN rail mounting
  • Measuring principle: Electronic acquisition and conversion of input signals in industrial temperature measurement.
  • Measuring system: 2-wire transmitter with analogue output, measuring input for Pt100 in 2, 3, or 4-wire connection.

Operating limits

  • Input limits:
    • Pt100 as per IEC 60751
    • Temperature range: -200 to 850 °C (-328 to 1562 °F)
    • Minimum span: 10 K
    • Cable resistance: max. 40 Ω per cable
    • Sensor current: ≤ 0.6 mA
  • Output signal: Analogue 4 to 20 mA
    • Undershooting measuring range: linear decrease to 3.8 mA
    • Exceeding measuring range: linear increase to 20.5 mA
    • Sensor break/short-circuit: ≥ 21.0 mA (guaranteed > 21.5 mA)
  • Load: Max. (VSupply - 12V) / 0.022 A
  • Induced current requirement: ≤ 3.5 mA
  • Current limitation: ≤ 23 mA
  • Switch-on delay: 4 s (Ia = 3.8 mA during switch-on)
  • Galvanic isolation: U = 2 kV AC (input/output)

Feed water

Not applicable for this device.

Product water

Not applicable for this device.

Electrical

  • Supply voltage: Ub = 12 to 35 V (reverse polarity protection)
  • Residual ripple: Uss ≤ 3 V at Ub ≥ 15 V, fmax. = 1 kHz
  • Technology: Two-wire technology

Mechanical / hydraulic

  • Weight: approx. 90 g (3.18 oz)
  • Housing material: PC/ABS, UL 94V0
  • Terminals: Pluggable screw terminal, max. 2.5 mm² solid, or strand with wire end sleeve.
  • Width: 22.5 mm
  • Mounting: For 35 mm top hat DIN rail according to IEC 60715

Environmental conditions

  • Ambient temperature limits: -40 to +85 °C (-40 to 185 °F) (for Ex-area, see Ex-certificate)
  • Storage temperature: -40 to +100 °C (-40 to 212 °F)
  • Climate class: IEC 60654-1, class C
  • Ingress protection: IP 20
  • Shock resistance: 4g / 2 to 150 Hz as per IEC 60068-2-6
  • Vibration resistance: See "Shock resistance"
  • EMC: Shock resistance and interference emission as per EN 61326-1 (IEC 61326) and NAMUR NE 21
  • Condensation: Permitted

Accuracy

  • Response time: 1 s
  • Reference operating conditions: Calibration temperature: +25 °C (77 °F) ± 5 K (9 °F)
  • Measuring error:
    • Influence of supply voltage: ≤ ±0.01%/V deviation from 24 V (Percentages refer to the full scale value)
    • Influence of ambient temperature (temperature drift) for Pt100: Td= ±(15 ppm/K * (full scale value + 200) + 50 ppm/K * of set measuring range) * Δ ϑ
    • Influence of load: ± 0.02%/100 Ω (Values refer to the full scale value)
  • Long term stability: ≤ 0.1 K/year or ≤ 0.05%/year (Values under reference operating conditions. % refer to the set span. The highest value is valid.)
  • Accuracy specification:
    • Resistance thermometer RTD Pt100: 0.2 K (0.36 °F) or 0.08% ( % refer to the set span. The highest value is valid.)

Certificates and approvals

  • CE-Mark: Meets EC directives.
  • Hazardous area approvals: ATEX EEx ia, nA; CSA IS, NI; CSA GP; FM IS, NI. (See separate Ex documentation for details).
  • Ship building approval: GL Germanischer Lloyd
  • Other approvals: UL recognized component.
  • Other standards and guidelines: IEC 60529 (IP-Code), IEC 60751, IEC 61010, IEC 61326, NAMUR NE 21.

Disclaimer: This extraction may miss figures, footnotes, or revisions. Contractual data must match the OEM PDF revision used on the project.

Official datasheet (PDF)

Lenntech datasheet mirror

Curated from selected public technical reference material for discovery and preliminary comparison. This summary is not a substitute for a current certified manufacturer datasheet. Verify revisions and design limits before use.