Back to Endress + Hauser temperature measurement
Endress+Hauser logo

Datasheet summary

iTemp TMT188

Engineering summary from PDF text extraction for iTemp TMT188. Verify every value with the OEM datasheet.

Summary

Product identification

  • Product Name: iTEMP TC TMT 188
  • Type: Temperature head transmitter for thermocouples (TC)
  • Application: For installation in a sensor head Form B. Fixed adjusted temperature head transmitter for converting TC input signals into a 4 to 20 mA analogue output signal.

Features and benefits

  • Fixed adjusted measurement range for thermocouples (TC)
  • 2 wire technology, 4 to 20 mA analogue output
  • High accuracy in total ambient temperature range
  • Fault signal on sensor break or short circuit, presettable to NAMUR NE 43
  • EMC to NAMUR NE 21, CE
  • UL recognized component to UL 3111-1
  • GL ship building approval
  • Ex-Certification:
    • ATEX Ex ia and dust zone 22 in compliance with EN 50281-1
    • FM IS
    • CSA IS
  • Galvanic isolation

Input

Thermocouples (TC)

TypeMeasurement RangeMin. Measurement Range
B (PtRh30-PtRh6)0 to +1820 °C (32 to 3308 °F)500 K (900 °F)
C (W5Re-W26Re)0 to +2320 °C (32 to 4208 °F)500 K (900 °F)
D (W3Re-W25Re)0 to +2495 °C (32 to 4523 °F)500 K (900 °F)
E (NiCr-CuNi)-200 to +915 °C (-328 to 1679 °F)50 K (90 °F)
J (Fe-CuNi)-200 to +1200 °C (-328 to 2192 °F)50 K (90 °F)
K (NiCr-Ni)-200 to +1372 °C (-328 to 2501 °F)50 K (90 °F)
L (Fe-CuNi)-200 to +900 °C (-328 to 1652 °F)50 K (90 °F)
N (NiCrSi-NiSi)-270 to +1300 °C (-454 to 2372 °F)50 K (90 °F)
R (PtRh13-Pt)0 to +1768 °C (32 to 3214 °F)500 K (900 °F)
S (PtRh10-Pt)0 to +1768 °C (32 to 3214 °F)500 K (900 °F)
T (Cu-CuNi)-200 to +400 °C (-328 to 752 °F)50 K (90 °F)
U (Cu-CuNi)-200 to +600 °C (-328 to 1112 °F)50 K (90 °F)
MoRe5-MoRe410 to +2000 °C (32 to 3632 °F)500 K (900 °F)
  • Cold junction: Internal (Pt100)
  • Cold junction accuracy: ± 1 K
  • Sensor current: 30 mA

Output

  • Output signal: 4 to 20 mA
  • Transmission: Temperature linear
  • Max. load: (Vpower supply - 8 V) / 0.025 A
  • Input current required: ≤ 3.5 mA
  • Current limit: ≤ 25 mA
  • Switch on delay: 4 s (during power up Ia = 3.8 mA)
  • Reply time: 1 s
  • Failure signal (fault monitoring):
    • Measurement range undercut: Linear drop to 3.8 mA
    • Exceeding measurement range: Linear rise to 20.5 mA
    • Sensor breakage; Sensor short circuit: ≥ 21.0 mA

Accuracy

Reference conditions

  • Calibration temperature: 23 °C ± 5 K
  • Thermocouple (TC) Measurement accuracy:
    • K, J, T, E, L, U: typ. 0.5 K
    • N, C, D: typ. 1.0 K
    • S, B, R, MoRe5-MoRe41: typ. 2.0 K
  • Influence of the internal reference junction: Pt100 DIN IEC 751 Cl. B

Application conditions

  • Influence of power supply: ≤ ±0.01%/V deviation from 24 V
  • Load influence: ≤ ±0.02%/100 Ω
  • Temperature drift:
    • Thermocouple (TC): Td = ± (15 ppm/K * max. meas. range + 50 ppm/K * preset meas. range) * ∆ϑ
    • ∆ϑ = Deviation of the ambient temperature according to reference condition.
  • Long term stability: ≤ 0.1 K / Year or ≤ 0.05% / Year

Electrical

  • Power supply: Ub = 8 to 35 V, polarity protected
  • Galvanic isolation (In/out): Û = 3.75 kV AC
  • Allowable ripple: Uss ≤ 5 V at Ub ≥ 13 V, fmax. = 1 kHz

Mechanical / hydraulic

  • Installation angle: No limit
  • Installation area: Connection head according to DIN 43 729 Form B; TAF 10 field housing
  • Dimensions: ∅44 mm housing diameter, 22.5 mm height
  • Weight: approx. 40 g
  • Materials:
    • Housing: PC
    • Potting: PUR
  • Terminals: Cable up to max. 1.75 mm² (secure screws)
  • Terminal connections: Power supply and current output, SETUP socket

Ambient conditions

  • Ambient temperature: -40 to +85 °C (for Ex-areas see Ex-certification)
  • Storage temperature: -40 to +100 °C
  • Climatic class: To EN 60 654-1, Class C
  • Moisture condensation: Allowable
  • Ingress protection: IP00 / IP66 installed
  • Vibration protection: 4g / 2 to 150 Hz according to IEC 60 068-2-6

Certification

  • CE mark: Fulfils EU regulations.
  • Ex-certification: Various options available (ATEX, CSA, FM). Refer to specific Ex documentation for hazardous area data.
  • UL recognized: Component to UL 3111-1.
  • GL approval: Germanische Lloyd ship building approval.

Disclaimer

Extraction may miss figures, footnotes, or revisions; contractual data must match the OEM PDF revision used on the project.