Back to Endress + Hauser temperature measurement
Endress+Hauser logo

Datasheet summary

iTemp TMT128

Engineering summary from PDF text extraction for iTemp TMT128. Verify every value with the OEM datasheet.

Summary

Product identification

  • Product Name: iTEMP® TC TMT128
  • Description: DIN rail mounted Temperature Transmitter for Thermocouples (TC)
  • Type: TMT128

Application

  • Temperature transmitter with fixed measuring ranges for converting TC input signals into a scalable 4 to 20 mA output signal.
  • Input: Thermocouples (TC)
  • Output: Analog, scalable 4 to 20 mA output signal

Features and Benefits

  • High accuracy: 0.08 % of span
  • Breakdown information in event of sensor break
  • Galvanic isolation: 2 kV (from sensor input to output)
  • Long-term stability: < 0.05 %/year
  • Electromagnetic compatibility to IEC 61326
  • Hazardous area approvals: ATEX, CSA, FM (EEx ia, nA, IS, NI, GP)
  • GL Germanische Lloyd / marine approval
  • UL recognized component to UL 3111-1

Function and System Design

  • Measuring Principle: Electronic acquisition and conversion of input signals.
  • Measuring System: Two-wire transmitter with analog output and measuring input for thermocouples.
  • Measured Variable: Temperature
  • Input Type: Thermocouples (TC)

Input Thermocouple Types and Measuring Ranges

DesignationStandardMeasuring Range LimitsMin. Span
BIEC 60584 part 10 to +1820 °C (32 to 3308 °F)500 K
CASTM E9880 to +2320 °C (32 to 4208 °F)500 K
DASTM E9880 to +2495 °C (32 to 4523 °F)500 K
EIEC 60584 part 1-270 to +1000 °C (-454 to 1832 °F)50 K
JIEC 60584 part 1-210 to +1200 °C (-346 to 2192 °F)50 K
KIEC 60584 part 1-270 to +1372 °C (-454 to 2501 °F)50 K
LDIN 43710-200 to +900 °C (-328 to 1652 °F)50 K
NIEC 60584 part 1-270 to +1300 °C (-454 to 2372 °F)50 K
RIEC 60584 part 1-50 to +1768 °C (-58 to 3214 °F)500 K
SIEC 60584 part 1-50 to +1768 °C (-58 to 3214 °F)500 K
TIEC 60584 part 1-270 to +400 °C (-454 to 752 °F)50 K
UDIN 43710-200 to +600 °C (-328 to 1112 °F)50 K

Output Signal

  • Analog Output: 4 to 20 mA
  • Breakdown Information: To NAMUR NE 43
  • Under-ranging: 3.8 mA (Standard)
  • Over-ranging: 20.5 mA (Standard)
  • Sensor Break: ≥ 21.0 mA (Standard), > 21.5 mA guaranteed if output setting is ≥ 21.0 mA.

Source Impedance

  • Max. (V<sub>power supply</sub> - 12V) / 0.022 A

Transmission Behavior

  • Temperature linear

Cold Junction Compensation

  • Internal cold junction (Pt100)
  • Accuracy of cold junction: ±1 K
  • Sensor current: ≤ 350 nA

Electrical

  • Supply Voltage (Ub): 12 to 35 V, reverse polarity protection
  • Permitted Residual Ripple (Uss): ≤ 3V at Ub ≥ 15V, f<sub>max</sub> = 1 kHz
  • Induced Current Requirement: ≤ 3.5 mA
  • Current Limitation: ≤ 23 mA
  • Switch-on Delay: 4 s (during switch-on procedure Ia ≤ 3.8 mA)
  • Response Time: 1 s

Accuracy

  • Reference Operating Conditions: Calibration temperature: +25 °C (77 °F) ± 5 K

  • Maximum Measured Error:

    • Influence of supply voltage: ≤ ±0.01%/V deviation from 24 V
    • Influence of load: ±0.02%/100 Ω
    • Influence of ambient temperature (temperature drift) for Thermocouples (TC): Td = ±(50 ppm/K * max. measuring range + 50 ppm/K * of set measuring range) * Δϑ Δϑ = deviation of the ambient temperature from the reference operating condition (25 °C (77 °F) ± 5 K).
  • Accuracy Table (Values refer to set span):

    Thermocouple Type (TC)Accuracy (Typical)
    K, J, T, E, L, U0.5 K or 0.08%
    N, C, D1.0 K or 0.08%
    S, B, R2.0 K or 0.08%

Long-Term Stability

  • ≤ 0.1 K/year or ≤ 0.05%/year (Values under reference operating conditions; larger value applies)

Installation Conditions

  • Installation Location: No restrictions

Environmental Conditions

  • Ambient Temperature Limits: -40 to +85 °C (-40 to 185 °F) (for Ex-area, see Ex-certificate)
  • Storage Temperature: -40 to +100 °C (-40 to 212 °F)
  • Climate Class: As per IEC 60654-1, Class C
  • Degree of Protection: IP 20
  • Shock Resistance: 4g / 2 to 150 Hz as per IEC 60068-2-6
  • Vibration Resistance: See 'Shock resistance'
  • Electromagnetic Compatibility (EMC): Shock resistance and interference emission as per IEC 61326 and NAMUR NE 21.
  • Condensation: Permitted

Mechanical / Hydraulic

  • Weight: approx. 90 g (3.18 oz)
  • Housing Material: PC/ABS, UL 94V0
  • Terminals: Pluggable screw terminal, max. 2.5 mm² (16 AWG) solid, or strand with wire end sleeve.

Human Interface

  • Display Elements: Illuminated yellow LED (2 mm, 0.08 in) signals device operation.
  • Operating Elements: None on the device.

Certificates and Approvals

  • CE-Mark: Device meets EC directives.
  • Hazardous Area Approvals: ATEX, CSA, FM, TIIS, NEPSI (Specific details and Ex versions available in separate documentation).
  • Ship Building Approval: GL (Germanischer Lloyd)
  • Other Standards and Guidelines: IEC 60529 (IP-Code), IEC 61010-1 (Safety), IEC 61326-1 (EMC), NAMUR.
  • UL: Recognized component to UL 3111-1.

Disclaimer: This extraction may miss figures, footnotes, or revisions. Contractual data must match the OEM PDF revision used on the project.