Product identification

  • Product: iTEMP TMT31 Temperature transmitter
  • Type: Form B (flat face) head transmitter with 4 to 20 mA analog output.
  • Document: TI01613T/09/EN/01.21

Application

The iTEMP TMT31 is characterized by its reliability, long-term stability, high precision, and diagnostic functions. It is designed for maximum safety and availability and is intended for installation in a terminal head B.

Benefits

  • Diagnostics information according to NAMUR NE107.
  • Rapid, no-tools wiring due to optional spring terminal technology.
  • Increased safety with Ex approvals.
  • High accuracy and flexibility with Callendar-van-Dusen equation.

Measuring System

  • Input Type: 1 x RTD (Resistance Temperature Detector)
  • Configuration: 2-wire device with one measuring input and one analog output. Installable in a Form B (flat face) terminal head as per DIN EN 50446.

Diagnostics

Standard diagnostic functions include:

  • Cable open-circuit, short-circuit of sensor wires
  • Incorrect wiring
  • Internal device errors
  • Overrange and underrange detection
  • Device temperature overrange/underrange detection
  • Low voltage detection

Input

  • Measured Variable: Temperature (temperature-linear transmission behavior)

Measuring Range

StandardDesignationαMeasuring Range LimitsMin. Span
IEC 60751:2008Pt100 (1)0.00385–200 to +850 °C (–328 to +1 562 °F)10 K (18 °F)
Pt1000 (4)–200 to +250 °C (–328 to +482 °F)
JIS C1604:1984Pt100 (5)0.003916–200 to +510 °C (–328 to +950 °F)10 K (18 °F)
GOST 6651-94Pt100 (9)0.003910–200 to +850 °C (–328 to +1 562 °F)10 K (18 °F)
  • Pt100 (Callendar van Dusen): Measuring range limits are specified by entering limit values dependent on coefficients A to C and R₀.
  • Connection Type: 2-wire, 3-wire, or 4-wire connection.
  • Sensor Current: ≤ 0.3 mA.
  • Wire Resistance Compensation:
    • 2-wire circuit: Possible (0 to 30 Ω).
    • 3-wire and 4-wire connection: Up to max. 50 Ω per wire.

Output

  • Output Signal: 4 to 20 mA, 20 to 4 mA (inversely configurable).
  • Failure Information: As per NAMUR NE43.
    • Underranging: Linear drop from 4.0 to 3.8 mA.
    • Overranging: Linear increase from 20.0 to 20.5 mA.
    • Failure (e.g., sensor failure, short-circuit): ≤ 3.6 mA ("Low") or ≥ 21 mA ("High"), selectable.
  • Linearization/Transmission Behavior: Temperature-linear.
  • Filter: 1st order digital filter (0 to 120 s), Network frequency filter (50/60 Hz, not adjustable).
  • Switch-on Delay: ≤ 5 s. During switch-on delay, output current is ≤ 3.8 mA.

Power Supply

  • Supply Voltage: 10 V ≤ Vcc ≤ 36 V (standard, protected against polarity reversal). Values for hazardous areas are in Ex documentation.
  • Current Consumption: 3.5 to 22.5 mA.

Electrical Connection

  • Terminals: Choice of screw or push-in terminals for sensor and power supply cables.
Terminal DesignCable DesignCable Cross-section
Screw terminalsRigid or flexible≤ 1.5 mm² (16 AWG)
Spring terminalsRigid or flexible0.2 to 1.5 mm² (24 to 16 AWG)
Flexible with ferrules0.25 to 1.5 mm² (24 to 16 AWG)

Wire end ferrules are required for spring terminals and flexible cables ≤ 0.3 mm².

Performance Characteristics

  • Response Time: ≤ 0.5 s.
  • Reference Operating Conditions: Calibration temperature +25 °C ±3 K, Supply voltage 24 V DC, 4-wire circuit for resistance adjustment.

Maximum Measured Error

Errors are specified as ±2 s (Gaussian distribution) and include non-linearities and repeatability.

Transmitter VersionMeasured Error (±) (in entire measuring range)Measured Error (±) (in limited range –50 to +250 °C)
0.15 K or 0.07 % of span ¹)0.1 K or 0.07 % of span ¹)

¹⁾ whichever is higher.

Operating Influences

(Errors as ±2 s, Gaussian distribution)

DesignationStandardAmbient Temp. Influence (±) per 1 °C changeSupply Voltage Influence (±) per V change
Pt100 (1) IEC 60751:20080.02 °C (0.04 °F)0.01 °C (0.014 °F)
Pt1000 (4) IEC 60751:20080.01 °C (0.02 °F)0.01 °C (0.009 °F)
Pt100 (5) JIS C1604:19840.01 °C (0.03 °F)0.01 °C (0.011 °F)
Pt100 (9) GOST 6651-940.02 °C (0.04 °F)0.01 °C (0.014 °F)
  • Long-term drift (±)¹⁾:
    • After 1 year: 0.05 K or 0.03 % of span
    • After 3 years: 0.06 K or 0.04 % of span
    • After 5 years: 0.07 K or 0.05 % of span ¹⁾ whichever is higher.

Sensor Adjustment

  • Sensor-transmitter-matching: Utilizes Callendar-Van-Dusen coefficients to improve accuracy.
  • 1-point adjustment (offset): Shifts the sensor value.
  • Current output adjustment: Correction of 4 and/or 20 mA output values.

Mounting

  • Mounting Location:
    • Terminal head Form B (flat face) as per DIN EN 50446.
    • With DIN rail clip on DIN rail as per IEC 60715 (TH35).

Environment

  • Ambient Temperature: –40 to +85 °C (–40 to +185 °F).
  • Storage Temperature: –50 to +100 °C (–58 to +212 °F).
  • Operating Altitude: Up to 4000 m above sea level.
  • Humidity: Max. 95 % relative humidity (as per IEC 60751:2008), condensation permitted.
  • Climate Class: C1 according to IEC 60654-1.
  • Degree of Protection: IP 00 (screw terminals), IP 30 (spring terminals). Dependent on installed terminal head/housing.
  • Shock and Vibration Resistance: Vibration as per DNVGL-CG-0339 : 2015 and DIN EN 60068-2-27 (8.6 to 150 Hz at 3g). Shock as per KTA 3505.
  • Electromagnetic Compatibility (EMC): CE conformity. Interference immunity as per IEC/EN 61326 series, industrial requirements. Interference emission as per IEC/EN 61326 series (CISPR 11), Class B, group 1.

Measuring Category

  • Measuring category II as per IEC 61010-1.

Pollution Degree

  • Pollution degree 2 as per IEC 61010-1.

Mechanical Construction

  • Weight: 40 to 50 g.
  • Materials: All materials are RoHS-compliant.
    • Housing: Polycarbonate (PC).
    • Terminals: Nickel-plated brass (screw), Tin-plated brass (push-in), contact springs 1.4310, 301 (AISI).
    • Potting Compound: SIL gel.

Operability

  • Remote Operation: Device parameters are configured via the CDI interface using specialized configuration tools.
  • Write Protection: Software write protection using a password.

Certificates and Approvals

Current certificates and approvals are available via the Endress+Hauser Product Configurator online.

MTTF

  • Mean Time To Failure (MTTF): 418 years.

Ordering Information

Detailed ordering information is available via Endress+Hauser sales organizations or the Product Configurator online.

Accessories

Various accessories are available, including:

  • Device-specific: Adapter for DIN rail mounting, DIN rail clip, DIN mounting set, US M4 securing screws.
  • Communication-specific: Commubox FXA291, Configuration kit TXU10.
  • Service-specific: Applicator software, Configurator tool, DeviceCare SFE100, FieldCare SFE500.
  • System components: RN22/RN42 active barriers, RIA15 process display, RNB22 system power supply.

Documentation

  • Operating Instructions
  • Description of Device Parameters
  • Supplementary ATEX and CSA documentation

Disclaimer: This extraction may miss figures, footnotes, or revisions. Contractual data must match the OEM PDF revision used on the project.

Official datasheet (PDF)

Lenntech datasheet mirror

Curated from selected public technical reference material for discovery and preliminary comparison. This summary is not a substitute for a current certified manufacturer datasheet. Verify revisions and design limits before use.