Back to Endress + Hauser temperature measurement
Endress+Hauser logo

Datasheet summary

iTherm Compactline TM311

Engineering summary from PDF text extraction for iTherm Compactline TM311. Verify every value with the OEM datasheet.

Summary

Product identification

  • Product: iTHERM CompactLine TM311
  • Type: Metric/imperial, compact RTD 4 to 20 mA/IO-Link thermometer for industrial and hygienic applications
  • Document ID: TI01439T/09/EN/05.25-00
  • Part Number: 71703224

Applications

  • Universal use in hygienic and aseptic applications within the food & beverages and pharmaceutical industries.
  • Standardization for machine and skid builders.

Operating limits

  • Measuring range: –50 to +200 °C (–58 to +392 °F)
  • Pressure range: Up to 50 bar (725 psi)
  • Protection class: IP69

Feed water

The document does not specify feed water parameters.

Product water

The document does not specify product water parameters.

Electrical

  • Output (Without electronics): Pt100 (4-wire connection)
  • Output (With electronics): IO-Link, 4 to 20 mA, 1 × PNP switch output (depending on connection type)
  • Supply voltage (Electronic version): Ub = 10 to 30 VDC, protected against reverse polarity. IO-Link communication requires at least 15 VDC.
  • Switching capacity (PNP): Ia ≤ 200 mA (ON), Ia ≤ 10 µA (OFF).
  • Switch cycles: > 10,000,000
  • Voltage drop (PNP): ≤ 2 V
  • Input current required: ≤ 3.5 mA for 4 to 20 mA; ≤ 9 mA for IO-Link.
  • Maximum current consumption: ≤ 23 mA for 4 to 20 mA.
  • Switch-on delay: 2 s
  • IO-Link: Version 1.1, COM2 (38.4 kBaud), Minimum cycle time 10 ms.

Mechanical / hydraulic

  • Design: Compact, made entirely of stainless steel.
  • Connection: M12 connection with IP69 protection.
  • Thermowell diameter: 6 mm (¹⁄₄ in)
  • Material: AISI 316L (corresponds to 1.4404 or 1.4435). Recommended max. temperature for continuous use in air: 650 °C (1202 °F).
  • Surface roughness: Ra ≤ 0.76 µm (30 µin) for standard surface, mechanically polished.
  • Weight: 0.2 to 2.5 kg (0.44 to 5.5 lbs) for standard versions.
  • Process connections: Various types available including compression fittings, weld-in adapters, clamps, and milk pipe connections. See the official PDF for exact tabulated values.

Performance characteristics

  • Measured error (Pt100 Cl. A, without electronics): ≤ 0.55 °C (0.99 °F) maximum.
  • Measured error (Pt100 Cl. A, with electronics): ≤ 0.48 °C (0.86 °F) maximum digital.
  • Measured error (with sensor-transmitter-matching): ≤ 0.14 °C (0.25 °F) maximum digital.
  • Long-term drift (IO-Link): ± 9 mK (1 month) to ± 31 mK (5 years).
  • Long-term drift (Current output): ± 2.5 µA (1 month) to ± 8.8 µA (5 years).
  • Response time T63 (iTHERM TipSens, direct contact): 1 s.
  • Response time T90 (iTHERM TipSens, direct contact): 1.5 s.
  • Electronics response time: Max. 1 s.

Environment

  • Ambient temperature range: –40 to +85 °C (–40 to +185 °F)
  • Storage temperature: –40 to +85 °C (–40 to +185 °F)
  • Operating altitude: Up to 2,000 m (6,600 ft) above sea level.
  • Degree of protection: IP69.
  • Shock and vibration resistance: Meets IEC 60751 requirements (3 g in the 10 to 500 Hz range).
  • Electromagnetic compatibility (EMC): In accordance with IEC/EN 61326 series and NAMUR Recommendation NE21.

Certificates and approvals

  • 3-A mark
  • EHEDG certification
  • ASME BPE conformity
  • FDA
  • EC 1935/2004
  • EN 2023/2006
  • TSE/ADI
  • GB4806-2016 and GB9685-2016
  • Marine approval according to DNV GL
  • Device safety according to EN 61010-1 and cCSAus
  • Electromagnetic compatibility as per NAMUR NE21
  • Diagnostics information according to NAMUR NE43

Disclaimer: This extraction may miss figures, footnotes, or revisions. Contractual data must match the OEM PDF revision used on the project.

Official datasheet (PDF)

Lenntech datasheet mirror

Curated from selected public technical reference material for discovery and preliminary comparison. This summary is not a substitute for a current certified manufacturer datasheet. Verify revisions and design limits before use.