Back to Endress + Hauser temperature measurement
Endress+Hauser logo

Datasheet summary

iTherm Moduline TM401

Engineering summary from PDF text extraction for iTherm Moduline TM401. Verify every value with the OEM datasheet.

Summary

Product Identification

  • Product Name: iTHERM ModuLine TM401
  • Product Type: Resistance thermometer for hygienic and aseptic applications
  • Technology: Basic technology, metric version

Applications

  • Hygienic and aseptic applications in the Food & Beverages and Life Sciences industries.
  • Suitable for use in non-hazardous areas.

Operating Limits

  • Measuring Range: –50 to +200 °C (–58 to +392 °F)
  • Pressure Range: Up to 50 bar (725 psi) - Note: Maximum pressure is dependent on the specific process connection.
  • Protection Class: Up to IP69K

Feed Water

  • Medium - State of Aggregation: Gaseous or liquid (including high viscosity media like yogurt).

Product Water

  • Not applicable for this product.

Measuring Principle

  • Resistance thermometers (RTD) using a Pt100 temperature sensor according to IEC 60751.
  • The sensor has a resistance of 100 Ω at 0 °C (32 °F) and a temperature coefficient α = 0.003851 °C⁻¹.
  • Available sensor types include thin-film platinum resistance thermometers (TF).

Output & Communication

  • Output Signal:
    • 4 to 20 mA analog output, HART®
    • HART® SIL transmitter (optional)
    • PROFINET® over Ethernet-APL
    • IO-Link®
  • Transmitters offer enhanced measurement accuracy and reliability.

Electrical

  • Power Supply: Details provided in wiring diagrams; dependent on transmitter type.
  • Overvoltage Protection: HAW562 and HAW569 surge arresters are available.

Performance Characteristics

  • Maximum Measurement Error (RTD):
    • Class A: ± (0.15 + 0.002 · |t|) °C
    • Class AA: ± (0.1 + 0.0017 · |t|) °C
    • Class B: ± (0.3 + 0.005 · |t|) °C
    • Note: |t| is the absolute temperature value in °C.
  • Self-heating: Negligible when using Endress+Hauser iTEMP temperature transmitters.
  • Response Time (t50 / t90):
    • 1x Pt100 thin-film sensor, Ø6 mm: 5 s / 11 s
    • Reduced tip 4.5 mm x 18 mm: 3.5 s / 9 s
    • Ø8 mm, reduced tip 5.3 mm x 20 mm: 5 s / 10.5 s
    • Note: Response time without transmitter.
  • Insulation Resistance: ≥ 100 MΩ at ambient temperature, measured between terminals and outer jacket with ≥ 100 VDC.

Mechanical Construction

  • Material: AISI 316L (corresponds to 1.4404 or 1.4435) with a recommended max. temperature for continuous use in air of 650 °C (1202 °F).
  • Surface Roughness (wetted surfaces):
    • Standard surface, mechanically polished: Ra ≤ 0.76 µm (30 µin)
    • Mechanically polished, buffed: Ra ≤ 0.38 µm (15 µin)
  • Weight: 0.5 to 2.5 kg (1 to 5.5 lbs) for standard options.

Terminal Heads

  • Various terminal head types (TA30A, TA30R, TA30S) available with different degrees of protection, temperature ranges, and materials.
  • TA30A: IP66/68, –50 to +150 °C, aluminum.
  • TA30R: IP69K, –50 to +130 °C, stainless steel 316L.
  • TA30S: IP65, –40 to +85 °C, polypropylene (PP), FDA-compliant.

Process Connections

  • Sanitary Connection (DIN 11851): Available for DN25, DN32, DN40, DN50. Maximum pressure up to 40 bar (580 psi) for DN25-DN40, and 25 bar (363 psi) for DN50.
  • Metal Sealing System: M12x1.5, G3/8", G1/2" threads. Maximum pressure 16 bar (232 psi) for thermowell diameters 6 mm and 8 mm.
  • Clamp (ISO 2852): Specific dimensions and properties detailed in the full document.
  • Other connections: Varivent®, Liquiphant-M weld-in adapter.

Certificates and Approvals

  • Hygiene standards: 3-A, EHEDG, ASME BPE, FDA, TSE.

Disclaimer: This extraction may miss figures, footnotes, or revisions. Contractual data must match the OEM PDF revision used on the project.