Back to Endress + Hauser temperature measurement
Endress+Hauser logo

Datasheet summary

RTD MLTTS01

Engineering summary from PDF text extraction for RTD MLTTS01. Verify every value with the OEM datasheet.

Summary

Product Identification

  • Product Name: MLTTS01 Modular thermometer
  • Application: Universal range, specifically mentioned for the Oil & Gas industry.

Measuring Principle

The MLTTS01 can utilize two types of temperature sensing elements:

  • Resistance Thermometer (RTD):

    • Uses a Pt100 sensor according to IEC 60751.
    • Resistance is 100 Ω at 0 °C (32 °F).
    • Temperature coefficient α = 0.003851 °C⁻¹.
    • Types of RTD elements:
      • Wire Wound (WW): Platinum wire coils in ceramic support, good for up to 600 °C (1 112 °F), but larger and more sensitive to vibrations.
      • Thin Film (TF): Thin platinum layer on a ceramic substrate, smaller and more vibration-resistant. Limit for tight tolerances (Cl. A) is approximately 300 °C (572 °F). Generally used below 400 °C (752 °F).
  • Thermocouple (TC):

    • Uses the Seebeck effect.
    • Standardized in IEC 60584 and ASTM E230/ANSI MC96.1.

Operating Limits

  • Measuring Range:
    • Resistance insert (RTD): –50 to +500 °C (–58 to +932 °F)
    • Thermocouple (TC): –40 to +1 100 °C (–40 to +2 012 °F)
  • Pressure Range: Up to 63 bar (914 psi) depending on process connection.
  • Degree of Protection: Up to IP68.
  • Ambient Temperature (Terminal Head):
    • Without mounted head transmitter: Depends on terminal head and cable gland/fieldbus connector.
    • With mounted head transmitter: –40 to +85 °C (–40 to +185 °F).
    • With mounted head transmitter and display: –20 to +70 °C (–4 to +158 °F).
  • Shock and Vibration Resistance:
    • RTD: 3G / 10 to 500 Hz (IEC 60751).
    • TC: 4G / 2 to 150 Hz (IEC 60068-2-6).

Feed Water (Process Conditions)

  • Process Pressure: Maximum process pressure depends on process temperatures and flow velocity. Specific limits for flanges:
    • DN25 PN40: 40 bar (580 psi) at 20 °C (68 °F).
    • DN25 PN63: 63 bar (914 psi) at 20 °C (68 °F).
  • Permitted Flow Velocity: Diminishes with increasing immersion length. A Thermowell sizing tool is available on the Endress+Hauser website.

Product Water (Output and Communication)

  • Head Transmitter Outputs:
    • Analog output 4 to 20 mA
    • HART®
    • PROFIBUS® PA
    • FOUNDATION Fieldbus™

Electrical

  • Power Supply (for transmitter): 24V DC / 30 mA (example using RN221N). Universal power supply 20 to 250 V DC/AC, 50/60 Hz.
  • Insulation Resistance:
    • RTD: > 100 MΩ at 25 °C (IEC 60751).
    • TC: > 1 GΩ at 20 °C, > 5 MΩ at 500 °C (IEC 1515).
  • Dielectric Strength: ≥1 000 V DC between terminals and insert sheath.
  • Self-heating: Negligible when using an Endress+Hauser iTEMP® temperature transmitter.

Mechanical / Hydraulic

  • Design: Modular design with standard terminal heads (DIN EN 50446).
  • Thermowell Material:
    • AISI 316L (1.4404/1.4435): Max. continuous use 650 °C (1 202 °F).
    • AISI 316Ti (1.4571): Max. continuous use 700 °C (1 292 °F).
    • Inconel 600 (2.4816): Max. continuous use 1 100 °C (2 012 °F).
  • Process Connections: Flange connections according to EN 1092-1 (e.g., DN25 PN40, DN25 PN63).
  • Dimensions: See diagrams in the original document for detailed dimensions (extension neck length E, immersion length IL, insertion length L).
  • Weight: From 1 to 4 kg (2.2 to 8.8 lbs) for standard options.

Accuracy

  • RTD Resistance Thermometer (IEC 60751):
    • Class AA (former 1/3): ± (0.1 + 0.0017 · |t|)
    • Class A: ± (0.15 + 0.002 · |t|)
    • Class B: ± (0.3 + 0.005 · |t|)
  • Temperature Ranges for Tolerance Classes:
    • Wire Wound (WW): Cl. A (–100 to +450 °C), Cl. AA (–50 to +250 °C).
    • Thin Film (TF): Cl. A (–30 to +300 °C), Cl. AA (–30 to +300 °C).
  • Thermocouples: Permissible deviation limits from standard characteristics are provided for IEC 60584 (Class 1 & 2) and ASTM E230/ANSI MC96.1 (Standard and Special tolerance).

Response Time

  • Reduced tip, Pt100 TF (16 mm diameter): t₅₀ = 30 s, t<0xE2><0x82><0x90> = 116 s.
  • Reduced tip, Thermocouple (ungrounded, 16 mm diameter): t₅₀ = 40 s, t<0xE2><0x82><0x90> = 130 s. (Values calculated at approx. 23 °C ambient, immersed in running water at 0.4 m/s with 10 K excess temperature).

Calibration

  • Endress+Hauser provides comparison temperature calibration from –80 to +1 400 °C (–110 to +2 552 °F) traceable to national and international standards. Calibration is performed on the replaceable insert.

Certifications and Approvals

  • CE Mark
  • Hazardous area approvals (ATEX, CSA, FM - details in separate documentation).
  • PED approval (compliant with Pressure Equipment Directive 97/23/EC).
  • Material certification (e.g., 3.1 according to EN 10204) available on request.
  • Various tests available for thermowells (Helium leak test, Pressure test, PMI, Load capacity calculation, Dye penetration test, Radiographic test).

Accessories

  • Communication-specific: Configuration kit TXU10, Commubox FXA195, Commubox FXA291, Wireless HART adapter SWA70, Fieldgate FXA320/FXA520, Field Xpert SFX100.
  • System components: Field display unit RIA16, Active barrier RN221N, Supply unit RNS221.
  • Software: Applicator, W@M, FieldCare.

Disclaimer: This extraction may miss figures, footnotes, or revisions. Contractual data must match the OEM PDF revision used on the project.

Official datasheet (PDF)

Lenntech datasheet mirror

Curated from selected public technical reference material for discovery and preliminary comparison. This summary is not a substitute for a current certified manufacturer datasheet. Verify revisions and design limits before use.