Datasheet summary
Omnigrad-S TR65, TC65
Engineering summary from PDF text extraction for Omnigrad-S TR65, TC65. Verify every value with the OEM datasheet.
Summary
Product Identification
- Models: Omnigrad S TR65 (resistance insert - RTD), TC65 (thermocouple insert - TC)
- Type: Modular thermometer, optionally with adjustable compression fitting.
Applications
- Fine chemicals
- Petrochemicals
- Power stations
- Environmental engineering
Operating Limits
- Measuring Range:
- Resistance insert (RTD): –200 to 600 °C (–328 to 1,112 °F)
- Thermocouple (TC): –40 to 1,100 °C (–40 to 2,012 °F)
- Static Pressure Range: Up to 75 bar (depending on process connection)
- Degree of Protection: Up to IP68
Feed Water
See the official PDF for exact tabulated values regarding permitted flow velocity.
Product Water
N/A
Electrical
- Head Transmitter Outputs and Protocols:
- Analog output 4 to 20 mA
- HART®
- PROFIBUS® PA
- FOUNDATION Fieldbus™
Mechanical / Hydraulic
- Thermometer Design: Modular design with standard terminal heads as per DIN EN 50446.
- Process Connection:
- Thread NPT ½", NPT ¾": Max. 75 bar
- Compression fitting (metal clamping ring): 40 bar
- Compression fitting (PTFE clamping ring): 5 bar
- Materials:
- AISI 316/1.4401, AISI 316L/1.4404, 1.4435: Max. 650 °C (1,202 °F) continuous use
- Alloy 600/2.4816: Max. 1,100 °C (2,012 °F) continuous use
Performance Characteristics
- Operating Conditions - Ambient Temperature Range:
- Terminal head (without mounted head transmitter): Depends on terminal head and cable gland/fieldbus connector.
- With mounted head transmitter: –40 to 85 °C (–40 to 185 °F)
- With mounted head transmitter and display: –20 to 70 °C (–4 to 158 °F)
- Shock and Vibration Resistance:
- RTD: Exceeds IEC 60751 requirements (3 g from 10 to 500 Hz).
- iTHERM StrongSens Pt100 (TF, vibration-resistant): 600 m/s² (60 g)
- Thin-film sensor (TF): >4 g
- Wire wound sensor (WW): >3 g
- Thermocouple TC: 4G / 2 to 150 Hz as per IEC 60068-2-6
- RTD: Exceeds IEC 60751 requirements (3 g from 10 to 500 Hz).
- Accuracy:
- RTD Resistance Thermometer (as per IEC 60751):
- Class A: ± (0.15 + 0.002 · |t| °C)
- Class AA: ± (0.1 + 0.0017 · |t| °C)
- Class B: ± (0.3 + 0.005 · |t| °C)
- Tolerance Class Temperature Ranges:
- Wire wound sensor (WW): Cl. A (–100 to +450 °C), Cl. AA (–50 to +250 °C)
- Thin-film version (TF): Standard Cl. A (–30 to +300 °C), Cl. AA (0 to +150 °C); iTHERM StrongSens Cl. A (–30 to +300 °C), Cl. AA (0 to +150 °C)
- Thermocouples (as per IEC 60584 / ASTM E230/ANSI MC96.1): Standard and Special tolerance classes available. See the official PDF for detailed deviation limits.
- RTD Resistance Thermometer (as per IEC 60751):
- Response Time (insert without transmitter, ambient ~23 °C in running water):
- Resistance Thermometer (Pt100, TF/WW):
- 6 mm diameter: t50 3.5 s, t90 8 s (Conical tip); t50 2 s, t90 5 s (Straight tip)
- 3 mm diameter: t50 2 s, t90 5 s (Conical tip); t50 1 s, t90 2.5 s (Straight tip)
- Thermocouple (not grounded):
- 6 mm diameter: t50 2.5 s, t90 7 s
- 3 mm diameter: t50 1 s, t90 2.5 s
- Thermocouple (grounded):
- 6 mm diameter: t50 2 s, t90 5 s
- 3 mm diameter: t50 0.8 s, t90 2 s
- Resistance Thermometer (Pt100, TF/WW):
- Insulation Resistance: ≥ 100 MΩ at ambient temperature.
- Self-heating: Negligible when an Endress+Hauser iTEMP temperature transmitter is connected.
- Calibration: Comparison temperature calibration from –80 to +1,400 °C (–110 to +2,552 °F) based on ITS90, traceable to national/international standards.
Installation Conditions
- Orientation: No restrictions.
- Minimum Immersion Depth (Error ≤0.1 K):
- Thin-film sensor (TF), iTHERM StrongSens (6 mm): ≥ 40 mm
- Thin-film sensor (TF) (3 mm): ≥ 30 mm
- Thin-film sensor (TF) (6 mm): ≥ 50 mm
- Wire wound sensor (WW) (3 mm): ≥ 30 mm
- Wire wound sensor (WW) (6 mm): ≥ 60 mm
Certificates and Approvals
- CE mark
- Hazardous area approvals (ATEX, CSA, FM etc. - refer to specific Ex documentation)
Disclaimer: This extraction may miss figures, footnotes, or revisions. Contractual data must match the OEM PDF revision used on the project.
Official datasheet (PDF)
Curated from selected public technical reference material for discovery and preliminary comparison. This summary is not a substitute for a current certified manufacturer datasheet. Verify revisions and design limits before use.