Back to Endress + Hauser temperature measurement
Endress+Hauser logo

Datasheet summary

RTD TH11, TH12

Engineering summary from PDF text extraction for RTD TH11, TH12. Verify every value with the OEM datasheet.

Summary

Product identification

RTD TH11 and TH12 general purpose RTD with terminal head or conductor extension lead wires for process and laboratory applications.

Areas of application

The TH11 and TH12 temperature sensors are RTDs designed for use in the process industry or factory automation. They are made up of a measurement probe with an RTD, insulation, and sheath.

Applications include:

  • Heat exchangers
  • Power & recovery area
  • Carbon regeneration furnace
  • Dryers
  • Flue Gas
  • Compressor stations
  • Process reactors

Measuring principle

These resistance thermometers use a Pt100 temperature sensor according to IEC 60751. The sensor has a resistance of 100 Ω at 0 °C (32 °F) and a temperature coefficient of α = 0.003851 °C⁻¹.

Two types of platinum resistance thermometers are available:

  • Wire wound (WW): Platinum wire in a ceramic support, sealed with a protective layer. Suitable for temperatures up to 600 °C (1112 °F). Relatively large and sensitive to vibrations.
  • Thin film (TF): Thin platinum layer on a ceramic substrate. Smaller size and better vibration resistance. Tight limit values of tolerance category A as per IEC 60751 are observable up to approximately 300 °C (572 °F). Generally used for temperatures below 400 °C (932 °F).

Equipment architecture

The single element RTDs are designed for process and laboratory applications within temperature ranges of -200 to 600 °C (-328 to 1112 °F).

  • Low range thin film RTDs: -50 to 200 °C (-58 to 392 °F). Constructed with silver plated copper internal leads and PTFE wire insulation.
  • High range RTDs: -200 to 600 °C (-328 to 1112 °F). Constructed with nickel internal leads inside swaged MgO insulated cable.

Measurement range

  • Low temperature range (thin film): TH11/TH12 models can cover -50 to 200 °C (-58 to 392 °F).
  • High temperature range (wire wound): TH11/TH12 models can cover -200 to 600 °C (-328 to 1112 °F).

Transmitter specifications

iTEMP® TMT18x

  • Housing material: PC
  • Potting material: PUR
  • Terminals: Cable up to max. 2.5 mm² / 16 AWG or with wire end ferrules.
  • Degree of protection: NEMA Type 4x Encl.

iTEMP® TMT8x

  • Housing material: PC (RoHS-compliant)
  • Potting material: PU
  • Terminals: Screw terminals (up to 2.5 mm²/16 AWG) or spring terminals (0.25 mm² to 0.75 mm²/24 AWG to 18 AWG).
  • Degree of protection: NEMA Type 4x Encl.

iTEMP® TMT162 field transmitter

  • Housing material: Die-cast aluminum (AlSi10Mg/AlSi12) with powder coating.
  • Cable entry: 2 x ½" NPT.
  • Degree of protection: NEMA Type 4x Encl. (IP67).
  • Display: Brilliant blue backlit display.

iTEMP® TMT162 (hygienic applications)

  • Housing material: Stainless steel 1.4435 (AISI 316L).
  • Cable entry: 2 x ½" NPT.
  • Degree of protection: IP69K.
  • Display: Brilliant blue backlit display.

iTEMP® HART® TMT142

  • Housing material: Die-cast aluminum (AlSi10Mg/AlSi12) with powder coating.
  • Cable entry: 3 x ½" NPT.
  • Degree of protection: NEMA Type 4x Encl. (IP67).
  • Display: Brilliant blue backlit display.

Housing

TA30A

  • Degree of protection: IP66/68.
  • Temperature range: -50 °C to +150 °C (-58 °F to +302 °F) without cable gland.
  • Material: Aluminum, polyester powder coated.
  • Cable entry: ½" NPT and M20x1.5.
  • Weight: 420 g (14.81 oz).

TA30H

  • Flameproof (XP) version, explosion-protected.
  • Degree of protection: IP66/68.
  • Temperature range: -50 °C to +150 °C (-58 °F to +302 °F) for rubber seal without cable gland.
  • Material: Aluminum, polyester powder coated.
  • Cable entry glands: ½" NPT, ¾" NPT, M20x1.5.
  • Weight: 860 g (30.33 oz).

Terminal head Al

  • Material: Die-cast Aluminium.
  • Sensor connection: ½" NPT Female.
  • Cable entry: ½" NPT or ¾" NPT Female.
  • Degree of protection: NEMA Type 4x Encl. (IP 66/68).

Terminal head plastic

  • Material: Polypropylene, FDA compliant.
  • Sensor connection: ½" NPT Female.
  • Cable entry: ½" NPT or ¾" NPT Female.
  • Degree of protection: NEMA Type 4x Encl.

Terminal head stainless steel, TA20J style

  • Material: Deep drawn stainless steel AISI 316L SS (hygienic design).
  • Sensor connection: ½" NPT female.
  • Cable entry: ½" NPT female.
  • Degree of protection: NEMA Type 4x Encl. (IP66).

Compression fitting

  • One time adjustable fitting: 316 Stainless Steel.
  • Re-adjustable fitting: 316 Stainless Steel with TFE ferrule.

Performance characteristics

Response time

Sheath Ø (T)High temp. rangeLow temp. range
1/16"2 s3 s
1/8"2 s7 s
3/16"3 s9 s
¼"not available6 s

Maximum measured error

RTD corresponding to IEC 60751. See the official PDF for exact tabulated values.

Transmitter long-term stability

  • ≤ 0.1 °C/year (≤ 0.18 °F / year) or ≤ 0.05% / year.

Insulation resistance

  • ≥100 MΩ at 25 °C (77 °F)
  • ≥10 MΩ at 300 °C (572 °F)

Sensor element tolerance

Sensor elementClassTemperature rangeMax. Tolerances (°C)
Thin film (TF)A-50 °C to +200 °C (-58 °F to +392 °F)± (0.15 + 0.002 · |t|)¹
B-50 °C to +200 °C (-58 °F to +392 °F)± (0.3 + 0.005 · |t|)¹
Wire wound (WW)A-100 °C to +450 °C (-148 °F to +842 °F)± (0.15 + 0.002 · |t|)¹
B-200 °C to +600 °C (-328 °F to +1112 °F)± (0.3 + 0.005 · |t|)¹

¹ |t| = absolute value °C.

Measurement accuracy with transmitters

TransmitterMeasurement accuracySensor currentGalvanic isolation (input/output)
TMT82 HART® / TMT84 PA / TMT85 FF± typ. 0.25 °C (0.45 °F)I ≤ 0.3 mAU = 2 kV AC
TMT180 Pt100 PCP / TMT181 multifunctional PCP0.2 °C (0.36 °F), optionalI ≤ 0.6 mA-
TMT182 HART®0.1 °C (0.18 °F) or 0.08%¹I ≤ 0.2 mAU = 2 kV AC
TMT162 HART® Field transmitter0.5 °C (0.9 °F) or 0.08%¹I ≤ 0.3 mAU = 2 kV AC
TMT142≤ 0.105 °C (0.19 °F)

¹ % is related to the adjusted measurement range (the larger value applies).

Wire specifications

  • TH11: 24AWG, 19 strand silver plated copper with 0.010” PTFE extruded outer. 1/8” sensors have 28AWG seven strand wires with the same extrusion.
  • TH12: 24AWG, seven strand silver plated copper with 0.010” PTFE then 0.015” FEP outer jacket. 1/8” sensors are 28AWG, seven strand SPC, 0.010” PTFE, 0.015” FEP white outer jacket. Flex armor is 0.272” nominal OD, 304SS 0.010” thick, square lock style.
  • Maximum temperature for extension cable: 200 °C (392 °F).

Installation conditions

Environmental conditions

Ambient temperature

  • Without mounted head transmitter: Depends on the terminal head used and the cable gland or fieldbus connector.
  • With mounted head transmitter: -40 to 85 °C (-40 to 185 °F).
  • With mounted head transmitter and display: -20 to 70 °C (-4 to 158 °F).

Shock and vibration resistance

  • 4G / 2 to 150 Hz as per IEC 60068-2-6.

Process conditions

Process pressure limits

See the official PDF for exact tabulated values and p/T load curves.

Mechanical construction

Material

  • Wetted parts: 316 SS.

Immersion length

  • TH11: 4", 6", 9", 12".
  • TH12: 6", 12", 18", 24", 48", 72", 120".

Sheath diameter

  • 1/8 in 316 SS
  • 3/16 in 316 SS
  • ¼ in 316 SS
  • 3/8 in (reduced to 3/16 in) 316 SS

Certificates and approvals

  • CE Mark
  • UL recognized components to UL 3111-1 (iTEMP® Series).
  • Complies with IEC 60529 (IP-Code), IEC 61010, ASTM E644, NEMA - ANSI / NEMA 250, IEC 60571.

Disclaimer: This extraction may miss figures, footnotes, or revisions. Contractual data must match the OEM PDF revision used on the project.

Official datasheet (PDF)

Lenntech datasheet mirror

Curated from selected public technical reference material for discovery and preliminary comparison. This summary is not a substitute for a current certified manufacturer datasheet. Verify revisions and design limits before use.