Datasheet summary
RTD T13, T14, T15
Engineering summary from PDF text extraction for RTD T13, T14, T15. Verify every value with the OEM datasheet.
Summary
Product Identification
- Model Series: T13, T14, T15
- Description: Explosion proof RTD assemblies in thermowells with spring loaded insert and enclosure for process industry.
- Document: TI00126R/09/EN/16.23-00
Application
- Heavy duty applications in chemicals, petrochemicals, refineries, and offshore platforms.
- Protection Class: IP66/67
Operating Limits
- Measuring Range: -200 to 600 °C (–328 to 1,112 °F)
- Ambient Temperature Range (Terminal Head):
- Without mounted head transmitter: Depends on terminal head and cable gland.
- With mounted head transmitter: –40 to 85 °C (–40 to 185 °F)
- SIL mode (HART 7 transmitter): –40 to 70 °C (–40 to 158 °F)
- Ambient Temperature Range (Terminal Head with Display):
- With mounted head transmitter and display: –20 to 70 °C (–4 to 158 °F)
- Ambient Temperature Range (Field Transmitter):
- Without display: –40 to 85 °C (–40 to 185 °F)
- With display and/or integrated overvoltage protection: –40 to +80 °C (–40 to +176 °F)
- SIL mode: –40 to +75 °C (–40 to +167 °F)
Feed Water
- Measuring Principle: Resistance thermometer (RTD) using Pt100 sensor.
- Sensor Resistance: 100 Ω at 0 °C (32 °F).
- Temperature Coefficient: α = 0.003851 °C⁻¹.
- Sensor Types:
- Wire Wound (WW): Suitable for temperatures up to 600 °C. Relatively large and sensitive to vibrations.
- Thin Film (TF): Smaller size, better vibration resistance. Tolerance Class A as per IEC 60751 only up to 200 °C (392 °F).
- iTHERM StrongSens (TF): Vibration resistance > 60 g.
Input / Measurement Ranges
- T13/T14/T15 (Low Temperature Range): –50 to 200 °C (–58 to 392 °F)
- T13/T14/T15 (High Temperature Range): –200 to 600 °C (–328 to 1,112 °F)
- T13/T14/T15 (Pt100 TF iTHERM StrongSens): –50 to +500 °C (–58 to +932 °F)
Product Water
- Transmitter Outputs:
- Analog output 4 to 20 mA
- HART®
- PROFIBUS® PA
- FOUNDATION Fieldbus™
- Bluetooth® connectivity (optional)
- Transmitter Long-term Stability: ≤ 0.1 °C (0.18 °F) / year or ≤ 0.05 % / year.
Electrical
- Power Supply (for associated components like RN42): 24–230 V AC/DC, 0/50/60 Hz.
- Galvanic Isolation: 2 kV AC on most devices.
- Overvoltage Protection Module:
- Maximum continuous voltage: 42 VDC
- Nominal current: 0.5 A at 80 °C
- Lightning surge current (10/350 µs): Iimp = 1 kA (per wire)
- Nominal discharge current (8/20 µs): In = 5 kA (per wire) / 10 kA (total)
- Temperature range: –40 to +80 °C (–40 to +176 °F)
- Series resistance: 1.8 Ω per wire (±5 %)
Mechanical / Hydraulic
- Protection Class: IP66/67 (Ex-version).
- Housing Materials:
- Aluminum with polyester powder coating
- Stainless steel 316L
- Terminal Heads: Internal shape and size in accordance with DIN EN 50446.
- Weight (TA30H Head):
- Aluminum: approx. 640 g (22.6 oz)
- Stainless steel: approx. 2,400 g (84.7 oz)
- Weight (TA30H with Display):
- Aluminum: approx. 860 g (30.33 oz)
- Stainless steel: approx. 2,900 g (102.3 oz)
- Wire Specifications: 24 AWG, 19 strand silver plated copper with 0.025 mm PTFE.
- Lead Design: Flying leads 3" or 5½" with brass crimped sleeves.
- Grounding: Minimum cross-section of 4 mm² (13 AWG) for housing to ground connection.
Dimensions
- Dimensions are provided in inches and millimeters. Refer to the official PDF for detailed drawings and tables.
- Spring travel of the insert: ½"
Materials
- Thermowell Materials: AISI 316/1.4401, AISI 316L/1.4404/1.4435, A105/1.0460/C22.8.
- Recommended Max. Temperature for Continuous Use in Air:
- AISI 316/1.4401, 316L/1.4404/1.4435: 650 °C (1,202 °F) (Limited use up to 800 °C)
- A105/1.0460/C22.8: 450 °C (842 °F)
Response Time
- RTD assembly T15 (without thermowell):
- High temperature range: 3 s
- Low temperature range: 9 s
- RTD assemblies with thermowell (T13/T14):
- Stepped thermowell: 20 s (high temp), 25 s (low temp)
- Tapered thermowell: 25 s (high temp), 30 s (low temp)
- Straight thermowell (¾"): 30 s (high temp), 35 s (low temp)
Accuracy
- RTD Thermometer as per IEC 60751:
- Class A: ± (0.15 + 0.002 · |t|)
- Class B: ± (0.3 + 0.005 · |t|)
- Class AA: ± (0.1 + 0.0017 · |t|)
- Temperature Ranges for Tolerance Class Compliance:
- Wire Wound (WW): Cl. A up to 450 °C, Cl. AA up to 250 °C.
- Thin-Film (TF): Cl. A up to 300 °C, Cl. AA up to 150 °C.
- Pt100 (TF) iTHERM StrongSens: Cl. A up to 300 °C, Cl. AA up to 200 °C.
Calibration
- Manufacturer provides comparison temperature calibrations from –20 to +300 °C (–4 to +573 °F) on the ITS-90.
- Calibrations are traceable to NIST standards.
- Conforms to ASTM E220.
Installation
- Minimum Immersion Length (Thermowell):
- Stepped thermowell: 63.5 mm (2.5 in)
- Tapered thermowell: 114.3 mm (4.5 in)
- ¾" straight thermowell: 101.6 mm (4 in)
- Weld-in thermowell: 114.3 mm (4.5 in)
Disclaimer: This extraction may miss figures, footnotes, or revisions. Contractual data must match the OEM PDF revision used on the project.