Datasheet summary
RTD TR88, TC88
Engineering summary from PDF text extraction for RTD TR88, TC88. Verify every value with the OEM datasheet.
Summary
Product Identification
- Product Name: TR88 (Resistance Thermometer - RTD) / TC88 (Thermocouple - TC)
- Description: Modular thermometer with extension neck and threaded connection for installation in an existing thermowell. Available with optional head transmitter.
Application
- Universal range of application
- Suitable for installation in existing thermowells
- Degree of protection up to IP68
Operating Limits
Measuring Range
- Resistance insert (RTD): –200 to 600 °C (–328 to 1 112 °F)
- Thermocouple (TC): –40 to 1 100 °C (–40 to 2 012 °F)
Ambient Temperature Range
- Terminal head without mounted head transmitter: Depends on terminal head, cable gland, or fieldbus connector.
- Terminal head with mounted head transmitter: –40 to 85 °C (–40 to 185 °F)
- Terminal head with mounted head transmitter and display: –20 to 70 °C (–4 to 158 °F)
Process Temperature Range
- Maximum: –200 to +1 100 °C (–328 to +2 012 °F), depending on sensor type and thermowell material.
Measuring Principle
Resistance Thermometer (RTD)
- Utilizes Pt100 temperature sensors (100 Ω at 0 °C).
- Types:
- Wire wound (WW): Larger size, good long-term stability up to 600 °C, sensitive to vibrations.
- Thin film (TF): Smaller size, better vibration resistance, potential deviation from standard at high temperatures.
Thermocouples (TC)
- Based on the Seebeck effect.
- Measure temperature differences; absolute temperature requires cold junction compensation.
- Standardized in IEC 60584 and ASTM E230/ANSI MC96.1.
Measuring System
- Modular design allowing for easy replacement or calibration of the insert without process interruption.
- Designed for installation in existing thermowells.
Input
Measured Variable
- Temperature (temperature-linear transmission behavior).
Measuring Range (Sensor Type Dependent)
- Pt100 thin-film: –50 to +400 °C (–58 to +752 °F)
- Pt100 thin-film, iTHERM StrongSens: –50 to +500 °C (–58 to +932 °F)
- Pt100 wire wound, extended: –200 to +600 °C (–328 to +1 112 °F)
- Thermocouple TC, type J: –40 to +750 °C (–40 to +1 382 °F)
- Thermocouple TC, type K: –40 to +1 100 °C (–40 to +2 012 °F)
Output
Output Signal
- Directly-wired sensors.
- Via Endress+Hauser iTEMP temperature transmitters with various protocols:
- Analog output 4 to 20 mA
- HART®
- PROFIBUS® PA
- FOUNDATION Fieldbus™
Head Transmitters (iTEMP)
- Improve accuracy and reliability compared to directly wired sensors.
- Configurable via PC with Endress+Hauser software.
- Features: Dual or single sensor input, optional pluggable display, mathematical functions, sensor monitoring (drift, backup, diagnostics), sensor-transmitter matching.
Power Supply
- Details on sensor connection types for RTD and Thermocouple are provided with diagrams.
Performance Characteristics
Accuracy
- Thermocouples: Permissible deviation limits as per IEC 60584 and ASTM E230/ANSI MC96.1 (Class 1, Class 2).
- RTD: Max. tolerances according to IEC 60751 (Class A, AA, B).
Self-Heating
- Negligible when using an Endress+Hauser iTEMP temperature transmitter.
Response Time
- Tested in accordance with IEC 60751 in flowing water. Values provided for different insert diameters, sensor types (TF, WW, TC), and grounding (grounded/not grounded).
Insulation Resistance
- RTD: > 100 MΩ at 25 °C (between terminals and sheath).
- TC: > 1 GΩ at 20 °C, > 5 MΩ at 500 °C (between terminals and sheath).
Dielectric Strength
- Tested at room temperature for 5 s:
- ≥ 1 000 V DC for Ø6 mm inserts (between terminals and sheath).
- ≥ 250 V DC for Ø3 mm inserts (between terminals and sheath).
Calibration
- Methods include calibration at fixed-point temperatures or against a reference thermometer.
- Requires specific immersion lengths for accurate calibration, especially at high temperatures or with head transmitters.
- Sensor-transmitter matching service available to reduce conversion errors.
- Standard calibrations offered from –80 to +600 °C based on ITS90.
Mounting
Orientation
- No restrictions.
Installation Instructions
- Designed for installation in an existing or separately ordered thermowell.
- Insertion length (ML) varies based on thermowell type, with a selectable range of 100 to 5000 mm.
Environment
Shock and Vibration Resistance
- Inserts exceed IEC 60751 requirements (3g, 10-500 Hz).
- Vibration resistance varies by sensor type:
- Pt100 (WW): > 30 m/s² (3g)
- Pt100 (TF), enhanced: > 40 m/s² (4g)
- iTHERM StrongSens Pt100 (TF): > 600 m/s² (60g)
- Thermocouple inserts: > 30 m/s² (3g)
Electromagnetic Compatibility (EMC)
- Depends on the head transmitter used.
Process
Process Pressure Range
- Maximum process pressure depends on the thermowell. Permitted flow velocity reduces with immersion depth.
Mechanical Construction
Weight
- 0.5 to 2.5 kg (1 to 5.5 lbs) for standard options.
Material
- Extension neck, insert, and process connection:
- AISI 316L / 1.4404, 1.4435: Recommended max. temp. 650 °C (1 202 °F). Good general corrosion resistance, especially in acidic, non-oxidizing atmospheres.
- AISI 316Ti / 1.4571: Recommended max. temp. 700 °C (1 292 °F). Similar to 316L with added titanium for intergranular corrosion resistance.
- Alloy 600 / 2.4816: Recommended max. temp. 1 100 °C (2 012 °F). Good resistance to aggressive, oxidizing, and reducing atmospheres, and chlorinated media.
Process Connection
- Threaded connection: Available in M, G, and NPT threads (e.g., M14x1.5, G 1/2", NPT 1/2").
- Threaded compression fitting (TA50): With SS316 clamping ring, max. process temp. 500 °C (932 °F), max. process pressure 40 bar at 20 °C.
Terminal Heads
- TA30A: IP66/68, –50 to +150 °C, Aluminum.
- TA30A with display: IP66/68, –50 to +150 °C, Aluminum, with TID10 display.
- TA30D: IP66/68, –50 to +150 °C, Aluminum, accommodates two head transmitters.
- TA30P: IP65, –40 to +120 °C, Polyamide, antistatic, for hazardous locations (Ex ia).
- TA20B: IP65, –40 to +80 °C, Polyamide.
- TA21E: IP65, –40 to 130 °C (silicone seal), Aluminum alloy.
Extension Neck
- Influences temperature in the terminal head; must stay within specified limits.
Insert
- RTD Inserts:
- TPR100: Pt100, 3- or 4-wire, mineral insulated. Measuring range –50 to +400 °C (Class A/AA). Diameters 3 mm, 6 mm.
- iTHERM TS111: Pt100, 3- or 4-wire, mineral insulated. Enhanced vibration resistance (> 60g). Measuring range –50 to +500 °C (Class A/AA). Diameter 6 mm.
- TPR100 (wire wound): Pt100, 3- or 4-wire, mineral insulated. Measuring range –200 to +600 °C (Class A/AA). Diameters 3 mm, 6 mm.
- TC Inserts:
- TPC100: Type K (Alloy 600) or Type J (316L). Measuring range: Type K -40 to 1200 °C (IEC), 0 to 1250 °C (ASTM). Type J -40 to 750 °C (IEC/ASTM). Diameters 3 mm, 6 mm.
Certificates and Approvals
- Available via the Product Configurator on the Endress+Hauser website.
Other Standards and Guidelines
- IEC 60529 (IP code)
- IEC/EN 61010-1
- IEC 60751 (RTDs)
- IEC 60584 and ASTM E230/ANSI MC96.1 (Thermocouples)
- DIN EN 50446 (Terminal heads)
Accessories
- Configuration kits (TXU10)
- Commubox FXA195 (HART)
- Wireless HART adapter (SWA70)
System Components
- RN22: Barrier for 0/4 to 20 mA signals.
- RIA15: Loop-powered 4 to 20 mA process indicator.
Supplementary Documentation
- Technical Information sheets for head transmitters, inserts, and thermowells.
- Supplementary ATEX and IECEx documentation.
Disclaimer: This extraction may miss figures, footnotes, or revisions. Contractual data must match the OEM PDF revision used on the project.
Official datasheet (PDF)
Curated from selected public technical reference material for discovery and preliminary comparison. This summary is not a substitute for a current certified manufacturer datasheet. Verify revisions and design limits before use.