Datasheet summary
CDC PHC LDO MTC series Electrodes/Sensors
Engineering summary from PDF text extraction for CDC PHC LDO MTC series Electrodes/Sensors. Verify every value with the OEM datasheet.
Summary
Product Identification
The Hach HQD series instruments are digital electrochemistry meters designed for measuring pH, conductivity, and dissolved oxygen (O₂). The system utilizes INTELLICAL digital electrodes which store calibration data within the electrode itself. Available models include:
- HQ 11D: For pH measurements.
- HQ 14D: For conductivity measurements.
- HQ 30D Flexi: A 1-channel instrument for pH, conductivity, and O₂.
- HQ 40D Multi: A 2-channel instrument for pH, conductivity, O₂, and Ion Selective Electrodes (ISE).
Operating Limits
Dissolved Oxygen (LDO)
- Range: 0.00 – 20.0 mg/l; 0 – 200 %
- Resolution: 0.01 or 0.1 mg/l; 0.1 % saturation
- Accuracy: ±1 % of the measuring range
- Air pressure compensation: Automatic
pH
- Range: 0–14
- Resolution (selectable): 0.1 / 0.01 / 0.001
- Accuracy: ±0.002
- Temperature compensation: Automatic
ORP (Oxidation Reduction Potential)
- Range: ±1500 mV
- Resolution: 0.1 mV
- Accuracy: ±0.1 mV
Ion concentration (ISE)
- Range: Depends on the ISE electrode
- Resolution (selectable): Max. 5 places; 0.1 / 0.01 / 0.001
- Accuracy: ±0.1 mV
Temperature
- Range: -10 to +110 °C
- Resolution: 0.1 °C
- Accuracy: ±0.3 °C
Conductivity
- Range: 0.01 µS /cm – 200 mS /cm (HQ 14D) or 0.01 µS /cm – 400 mS /cm (HQ 30D, HQ 40D)
- Resolution: Max. 5 places, 2 decimal places, if possible
- Accuracy: ±0.5 % (1 µS /cm – 200 mS/cm for HQ 14D) or ±0.5 % (1µS /cm – 400 mS / cm for HQ 30D, HQ 40D)
- Temperature compensation: Non-linear (natural water in conformity with DIN 38404 and EN ISO 7888), non-linear (NaCl), linear coefficient [numeric value] % / °C, no compensation.
Resistivity
- Range: 2.5 Ω cm – 49 MΩ cm
- Resolution: Max. 5 figures
- Accuracy: ±0.5 %
TDS (Total Dissolved Solids)
- Range: 0.0 – 50,000 mg/l
- Resolution: Max. 3 figures
- Accuracy: ±0.5 in the measuring range
Salinity
- Range: 0 - 42 (g/kg, ‰, no unit)
- Resolution: Up to 0.01 ppt
- Accuracy: ±0.1 mg/l at < 8 mg/l
Features
- Autoread: Supported on all models.
- Autocal: Automatic buffer recognition for pH.
- Calibration:
- pH: IUPAC (1.679; 4.005; 7.000; 10.012) and DIN (1.09; 4.65; 9.23) standards, colour coded (4, 7, 10). Max. 4 calibration points.
- Conductivity: Demal (1 D; 0.1 D; 0.01 D), Molar (0.1 M; 0.01 M; 0.001 M), NaCl (0.05 %; 25 µS /cm; 1000 µS /cm; 18 mS /cm). User defined; standard sea water. 1 calibration point.
- O₂: 1 calibration point.
- Sensor Status Indicator: Supported on all models.
- Interfaces: Waterproof USB port for printer, PC, keyboard and USB stick.
- Password Protection: Supported on all models.
- Data Management: Basic, detailed, total (GLP).
- Data Memory: 500 readings; data can be saved manually or automatically.
- Sample/Operator ID: Alphanumeric, max. 12 characters; 12 sample names and 20 user names. Automatic logging of sample numbers (0 – 999).
- Measurement Mode: Manual, interval, continuous; analytical methods editable.
- Display: Backlit graphic display; 240 × 160 pixels; automatic switch-off in economy mode. Date and time display. Simultaneous display of two parameters on HQ 40D.
- Power Supply: 115 V / 250 V (power unit optional).
- Battery Operation: 4 AA batteries or rechargeable batteries (battery charger needed).
- Protection Class: IP 67 for instrument, outdoor electrodes, and connections.
- Dimensions, Weight: 95 × 197 × 36 mm (H × W × L), 323 g (without batteries).
Electrodes / Sensors
All INTELLICAL electrodes/sensors are watertight to a depth of 3 metres for 24 hours, including the temperature sensor. Outdoor electrodes/sensors are watertight to a depth of 30 metres for 24 hours, with a steel housing and reinforced cable.
- pH: Standard (liquid or gel electrolyte), rugged outdoor, and ULTRA (for low ionic strengths). Cable lengths from 1 m to 30 m.
- Conductivity: Standard and rugged outdoor 4-pin graphite electrodes. Cable lengths from 1 m to 30 m.
- LDO (Dissolved Oxygen): Standard and rugged outdoor sensors. Cable lengths from 1 m to 30 m. Includes a specific sensor for BOD measurement.
- ORP: Standard and rugged outdoor electrodes with gel or liquid electrolyte. Cable lengths from 1 m to 30 m.
- Ion Selective Electrodes (ISE): Sodium, Chloride, and Fluoride ISE standard electrodes. Available with 1 m or 3 m cables.
Buffer and Conductivity Standard Solutions
A range of certified pH buffer solutions (IUPAC and DIN standards) and conductivity standard solutions (KCl and NaCl) are available in various quantities and concentrations, traceable to standard reference materials.
- pH Buffer Solutions: Certified standards and quality buffer solutions in bottles, powder pillows, and single pouches.
- Conductivity Standard Solutions: Certified solutions including KCl and NaCl in various concentrations.
Accessories
- Outdoor Kit
- Electrode holder
- Cases (for standard and outdoor electrodes)
- Cable markers
- Electrode markers
- USB-/AC-adaptor
- USB stick (128 MB capacity)
- Keyboard with USB connector
- LDO sensor cap
- Stand for HQD meters
- Probe holder
- BOD manager PC software
Disclaimer: This extraction may miss figures, footnotes, or revisions. Contractual data must match the OEM PDF revision used on the project.