Datasheet summary

LT200 Thermostat

Engineering summary from PDF text extraction for LT200 Thermostat. Verify every value with the OEM datasheet.

Summary

Product identification

  • Product: DRY THERMOSTAT DRB200
  • Type: DRB200-1, DRB200-2

Operating limits

  • Temperature range: 37–150 °C
  • Time range: 1– 480 min.
  • Heating rate: From 20 °C to 150 °C in 10 min.
  • Temperature stability: ±1 °C in conformity with EN, ISO, EPA methods

Applications

The DRB200 thermostat is designed for digestions preceding analysis of parameters such as COD, Total nitrogen, Total phosphorus, Metals, Total chromium, TOC, and Trihalogenmethane.

Standard Digestion Parameters

ParameterTemp. [°C]Time [min.]
COD150120
Total nitrogen100 / 105*60 / 30
Total phosphorus100 / 105*60 / 30
Metals10060
Total chromium10060
TOC105120
Trihalogenmethane1008

* Digestion temperature and time depend on the used reagents.

Mechanical / hydraulic

DRB200-1

  • Number of cuvettes: 9 × 16 mm diameter
  • Number of reaction vessels: 2 × 20 mm diameter

DRB200-2

  • Number of cuvettes: 15 × 16 mm diameter or 21 × 16 mm diameter

  • Number of reaction vessels: 4 × 20 mm diameter or 30 × 16 mm diameter

  • Dimensions: 250 × 145 × 310 mm (W×H×D)

  • Weight: 2 kg (DRB200-1), 2.8 kg (DRB200-2)

  • Splash protection: Two transparent lids

Electrical

  • Power input (DRB200-1): 115 V / 300 VA or 230 V / 450 VA
  • Power input (DRB200-2): 115 V / 600 VA or 230 V / 900 VA