TU5 Turbidimeter

Engineering summary from PDF text extraction for TU5 Turbidimeter. Verify every value with the OEM datasheet.

Summary

Product Identification

The TU5 Series Turbidimeters feature a groundbreaking 360° x 90° Detection Technology for improved precision and sensitivity. They are suitable for applications including drinking water, power, beverage, and pharmaceutical industries. The instruments meet EPA Method 10258 and ISO 7027-1:2016 standards for regulatory reporting.

TU5200 Benchtop Laser Turbidimeters

TU5300sc / TU5400sc Online Laser Turbidimeters

Technical Data

TU5200

  • Light source: Class 2 laser product, 650 nm (EPA) or Class 1 laser product, 850 nm (ISO).
  • Measuring range:
    • EPA: 0 - 700 NTU / FNU / TE/F / FTU; 0 - 100 mg/L; 0 - 175 EBC
    • ISO: 0 - 1000 NTU / FNU / TE/F / FTU; 0 - 100 mg/L; 0 - 250 EBC
  • Accuracy: ±2% plus 0.01 NTU from 0 - 40 NTU; ±10% of reading from 40 - 1000 NTU (based on Formazin primary standard).
  • Resolution: 0.0001 NTU / FNU / TE/F / FTU / EBC / mg/L
  • Repeatability:
    • <40 NTU: Better than 1% of reading or ±0.002 NTU.
    • 40 NTU: Better than 3.5% of reading.

  • Stray light: <10 mNTU
  • Units: NTU, FNU, TE/F, FTU, EBC; mg/L (if calibrated with Degrees calibration curve).
  • Operating temperature range: 10 - 40 °C
  • Operating humidity: 80% at 30 °C (non-condensing)
  • Sample temperature: 4 - 70 °C
  • Storage conditions: -30 - 60 °C
  • Power requirements: 100 - 240 V AC, 50/60 Hz
  • Certifications: CE compliant, US FDA accession numbers provided for EPA and ISO versions, Australian ACMA Marking.
  • Dimensions (H x W x D): 195 mm x 409 mm x 278 mm
  • Weight: 2.4 kg
  • Warranty: 2 years

TU5300sc / TU5400sc

  • Light source: Class 2 laser product, 650 nm (EPA) or Class 1 laser product, 850 nm (ISO).
  • Measuring range:
    • EPA: 0 - 700 NTU / FNU / TE/F / FTU; 0 - 100 mg/L; 0 - 175 EBC
    • ISO: 0 - 1000 NTU / FNU / TE/F / FTU; 0 - 100 mg/L; 0 - 250 EBC
  • Accuracy: ±2% or 0.01 NTU from 0 - 40 NTU; ±10% of reading from 40 - 1000 NTU (based on Formazin primary standard).
  • Resolution: 0.0001 NTU / FNU / TE/F / FTU / EBC
  • Repeatability: Better than 1% of reading or ±0.002 NTU (TU5300sc) or ±0.0006 NTU (TU5400sc).
  • Stray light: <10 mNTU
  • Units: NTU, FNU, TE/F, FTU, EBC
  • Signal average time: TU5300sc: 30 - 90 seconds; TU5400sc: 1 - 90 seconds
  • Response time: TU5300sc: T90 <45 seconds at 100 mL/min; TU5400sc: T90 <30 seconds at 100 mL/min
  • Sample temperature: 2 - 60 °C
  • Sample pressure: 6 bar maximum (compared to air at sample temperature).
  • Flow rate: 100 - 1000 mL/min; optimal flow rate: 200 - 500 mL/min
  • Operating temperature range: 0 - 50 °C
  • Operating humidity: 5 - 95% RH (non-condensing).
  • Storage conditions: -40 - 60 °C
  • Enclosure waterproof rating: Electronic compartment IP55; other functional units IP65.
  • Certifications: CE compliant, US FDA accession numbers provided for EPA and ISO versions, Australian ACMA Marking.
  • Dimensions (H x W x D): 249 mm x 268 mm x 190 mm
  • Weight: 2.7 kg (5.0 kg with accessories)
  • Warranty: 2 years

Note: Technical data is subject to change without notice.

Principle of Operation

The TU5 Series turbidimeters utilize a 360° x 90° optical design to measure turbidity. A laser is directed into the sample, causing suspended particles to scatter light. The light scattered at a 90° angle is collected by a detector after being reflected by a conical mirror. The amount of scattered light is proportional to the sample's turbidity. This design is optimized for high accuracy at low turbidity ranges and does not include ratio technology, which is typically used for high turbidity applications with color and large particle interference.

Accessories and Calibration

  • Calibration: Stablcal calibration sets (with and without RFID), glass rod secondary turbidity standard.
  • Vials: Replacement vials for TU5300sc/TU5400sc, sample vials for TU5200.
  • Accessories: Automatic cleaning module, flow sensor, desiccant cartridge, maintenance kit, SIP10 sipper unit, manual vial wiper.
  • Note: An SC controller is required for operation of the TU5300sc or TU5400sc.