Back to BMT Ozone Measurement
BMT logo

Datasheet summary

Ozone Monitor BMT 932

Engineering summary from PDF text extraction for Ozone Monitor BMT 932. Verify every value with the OEM datasheet.

Summary

Product identification

  • Product Name: OZONE MONITOR BMT 932
  • Manufacturer: BMT MESSTECHNIK GMBH
  • Type: Ambient Ozone Monitor, Low Concentration

Features

  • Dual beam UV photometer
  • 3 years warranty
  • Long-life mercury lamp (5 years warranty)
  • High accuracy, error less than 1%
  • Resolution: 0.001 ppmv
  • Pressure and temperature compensated
  • Available ranges: 1 to 250 ppmv
  • 19'' rack mount, or cabinet versions
  • Traceable to international standard (NIST)
  • Available in 1, 3, or 6 channel versions
  • Two individual threshold alarm levels per channel
  • Three configurable EMO outputs
  • Cycle time: 20 s per channel
  • Dual scrubber system (utility + reserve)
  • Integrated ozone generator for automatic utility scrubber testing (every 24 hours)
  • Automatic reserve scrubber takeover upon utility scrubber failure
  • Sample gas flow: 1 l/min, automatically controlled via an electronic flow meter
  • Configuration via front panel or Windows PC software
  • Complete error and event log with timestamps

Applications

  • Monitoring of ozone in ambient air
  • TLV monitoring (Threshold Limit Value)
  • Ambient monitoring in ozone plants
  • Vent-Gas monitoring

Specifications

Measurement principle

  • Dual-beam UV photometer (254 nm)

UV lamp

  • Low pressure mercury lamp, long life design, burnt-in for 300 h

Display

  • Alphanumeric LCD 2x20 char.

Concentration ranges

  • 1 ppmv (1.000 ppmv), 2.000, 5.000, 10.000, 20.00, 100.0 and 250.0 ppmv

Noise

  • ±0.001 ppmv

Minimum detectable concentration

  • 0.002 ppmv

Accuracy

  • Maximum error (traceable to NIST) is the sum of: 0.8% of measurement + 0.2% of scale

Response time

  • 20 s (0 - 95 %)

Temperature and pressure compensation

  • Standard

Flow rate

  • 1 l/min, automatically controlled by an electronic flow meter

Sample ports

  • 1, 3 or 6 (automatic and manual selection)

Scrubber life

  • 8 years (safety monitor operation)

Threshold alarms

  • Two individual alarm thresholds per channel
  • 3 EMO outputs controlled by any combination of alarms

Error relay

  • Indicates instrument failure, including:
    • Lamp low (easy lamp replacement)
    • Low flow
    • Scrubber failure (tested every 24 h, reserve scrubber takes over automatically)

Warmup time

  • Maximum 7 min., typical 4 min.

Gas ports

  • Compression type for ¼" PFA/FEP/PTFE tubing (Cabinet: ¼" One-Touch Fitting)
  • Safety catalyst for off gas included

Signal outputs

  • Concentration: 4-20 mA (isolated, active)
  • Concentration: 0-1 V, 0-10 V (isolated)
  • Sample port identification by binary output
  • Option: 932-RELAY with contacts identifying each port plus its low/high alarm, accessed by 50-pole D-Sub connector

Control output

  • Lamp low (relay contact, 30 V, 1 A)

Digital interface

  • RS-232, isolated, showing concentration and error messages
  • Link Mode commands for full control from Windows PC or PLC

Accessories kit (included)

  • One filter holder & one mounting fitting for filter holder per sample port
  • 200 filter inserts

Spare parts (available)

  • UV lamp, scrubber, filter holder

Power

  • Universal line voltage: 100 - 240 VAC, 35 VA, 50/60 Hz
  • Optional: 18-36 VDC, 25W (19” rack mount)

Ambient temperature

  • 0 - 45°C (non-condensing)

Cabinet version

  • IP65, NEMA 4X

Physical dimensions and weight

  • 19" rack mount version:
    • Dimensions (H x D): 132 x 300 mm
    • Weight: 9 kg
  • Cabinet version (BMT 932 C):
    • Dimensions (W x H x D): 357 x 440 x 133 mm
    • Weight: 7 kg
    • Mounting: Wall mount with brackets (four mounting holes 6 mm ID, spaced 307 x 397 mm, W x H)

Compliance

  • CE-marked (EMC & safety)
  • cTUVus NRTL-listed
  • RoHS

Filters

  • Filter inserts are made of fine pure glass fibre felt to remove aerosol particles without affecting the ozone.
  • Filters should be replaced on a regular basis, with the optimum interval determined by the user for their specific situation.

Disclaimer: This extracted summary may not include all figures, footnotes, or revisions. Contractual data must match the official OEM PDF revision used on the project.

Official datasheet (PDF)

Lenntech datasheet mirror

Curated from selected public technical reference material for discovery and preliminary comparison. This summary is not a substitute for a current certified manufacturer datasheet. Verify revisions and design limits before use.