
Datasheet summary
Ozone-in-water Sensor BMT 964 AQ-LC
Engineering summary from PDF text extraction for Ozone-in-water Sensor BMT 964 AQ-LC. Verify every value with the OEM datasheet.
Summary
Product identification
- Product Name: OZONE-IN-CLEAN-WATER SENSOR BMT 964 AQ-LC
- Product Line: BMT Ozone Measurement
- Sensor Type: Low Concentration Ozone Sensor for Clear and Clean Water
Features
- Dual beam UV photometer
- Stainless steel cuvette
- Long-life mercury lamp
- Warranty: 3 years (instrument), 5 years (UV lamp)
- High accuracy, error less than 0.5%
- Selectable concentration ranges from 0.5 to 15 g/m³ (ppmw)
- Simple cleaning of cuvette windows
- Relay contact for control of automatic zeroing
- Failure warnings: lamp low, cuvette dirty, overrange, summary error
- Key functions programmable via Windows PC
- High and low limit alarms
- Timing sequence for automatic zeroing
- Isolated signal outputs: 4-20mA and 0-10V
- Isolated RS-232 interface (bidirectional)
- Optional Remote Display (BMT 964 RD) available
Applications
- Monitoring of ozone in clear and clean water systems, or water with constant turbidity.
- Pharmaceutical
- Biotechnology
- Food, baby food
- Beverages
Specifications
Measurement Principle
- Dual-beam UV photometer (254 nm)
MTBF
- Instrument including UV lamp: > 65,000 hours
- Instrument excluding UV lamp: > 120,000 hours
UV Lamp
- Low pressure mercury lamp, long-life design, burnt-in for 300 hours
Concentration Ranges
- Selectable: 0.5, 0.75, 2, 5, 10 and 15 g/m³ (ppmw)
- Selectable units: g/m³ and ppmw
Accuracy
- After zeroing, max. error is the sum of 0.4% of measurement + 0.1% of scale.
Repeatability Error
- 0.2% of measurement
Response Time
- 0.03 s (analog output)
- 0.3 s (remote display)
Zero Drift
- Typically 0.2% of range per day, after warm-up, non-cumulative.
Operating Limits
- Max. Inlet Pressure: 10 bar
- Ambient Temperature: 0 - 50°C (non-condensing)
Materials in Contact with Ozonated Water
- Stainless steel (SS), PVDF, PFA, quartz, FFPM
Hydraulic
- Fluid Ports: PVDF compression fittings for 4 x 6 mm (5/32'' x 1/4'') PTFE tubing or 1/4" SS Swagelok.
- Recommended Flow Rate: 0.05 to 0.1 l/min (typical)
- Pressure Drop: Approx. 10 mbar at 0.1 l/min (without valve)
Signal Outputs
- Concentration: 4 - 20 mA (isolated, active)
- Concentration: 0 - 10 V (isolated)
- Concentration Alarms: High Alarm, Low Alarm (latching, not latching)
Control Input
- Set to zero (24 V, 18 mA, isolated)
Control Outputs (Relay Contacts, 28 V, 0.5 A, isolated)
- Lamp Low
- Cuvette Dirty
- High Alarm
- Low Alarm
- Purge Control
Error Handling
- Error Relay: 30 V, 1 A, summarizing instrument failures.
- Warnings and Errors: Lamp Low Warning/Error, Lamp Off Error, Cuvette Dirty Warning/Error, Overrange, Overpressure.
Serial Interface
- RS-232, bidirectional, isolated
- Baud Rates: 2400 - 38400
Automatic Zeroing
- Possible via the purge contact.
Software
- Name: BMT 964 Link
- Function: Instrument configuration and readout of Event and Error Logs on a Windows PC.
Electrical
- Power Input: Wide range 100 - 240 VAC, 15 VA
- Optional DC Power: 12 - 36 VDC, 15 W
Mechanical
- Dimensions (W x H x D): 360 x 160 x 91 mm
- Weight: Approx. 4 kg
- Enclosure: Splash proof aluminium (IP 65, NEMA 4X) with sea water resistant coating (RAL 5009, azure).
- Mounting: Four mounting brackets provided.
- Connectors: All electric connectors are waterproof.
Disclaimer: This extracted summary may omit figures, footnotes, or revisions. Contractual data must match the OEM PDF revision used on the project.
Official datasheet (PDF)
Curated from selected public technical reference material for discovery and preliminary comparison. This summary is not a substitute for a current certified manufacturer datasheet. Verify revisions and design limits before use.