Summary

Product identification

  • Product Name: OZONE-IN-CLEAN-WATER SENSOR BMT 964 AQ-LC
  • Product Line: BMT Ozone Measurement
  • Sensor Type: Low Concentration Ozone Sensor for Clear and Clean Water

Features

  • Dual beam UV photometer
  • Stainless steel cuvette
  • Long-life mercury lamp
  • Warranty: 3 years (instrument), 5 years (UV lamp)
  • High accuracy, error less than 0.5%
  • Selectable concentration ranges from 0.5 to 15 g/m³ (ppmw)
  • Simple cleaning of cuvette windows
  • Relay contact for control of automatic zeroing
  • Failure warnings: lamp low, cuvette dirty, overrange, summary error
  • Key functions programmable via Windows PC
  • High and low limit alarms
  • Timing sequence for automatic zeroing
  • Isolated signal outputs: 4-20mA and 0-10V
  • Isolated RS-232 interface (bidirectional)
  • Optional Remote Display (BMT 964 RD) available

Applications

  • Monitoring of ozone in clear and clean water systems, or water with constant turbidity.
  • Pharmaceutical
  • Biotechnology
  • Food, baby food
  • Beverages

Specifications

Measurement Principle

  • Dual-beam UV photometer (254 nm)

MTBF

  • Instrument including UV lamp: > 65,000 hours
  • Instrument excluding UV lamp: > 120,000 hours

UV Lamp

  • Low pressure mercury lamp, long-life design, burnt-in for 300 hours

Concentration Ranges

  • Selectable: 0.5, 0.75, 2, 5, 10 and 15 g/m³ (ppmw)
  • Selectable units: g/m³ and ppmw

Accuracy

  • After zeroing, max. error is the sum of 0.4% of measurement + 0.1% of scale.

Repeatability Error

  • 0.2% of measurement

Response Time

  • 0.03 s (analog output)
  • 0.3 s (remote display)

Zero Drift

  • Typically 0.2% of range per day, after warm-up, non-cumulative.

Operating Limits

  • Max. Inlet Pressure: 10 bar
  • Ambient Temperature: 0 - 50°C (non-condensing)

Materials in Contact with Ozonated Water

  • Stainless steel (SS), PVDF, PFA, quartz, FFPM

Hydraulic

  • Fluid Ports: PVDF compression fittings for 4 x 6 mm (5/32'' x 1/4'') PTFE tubing or 1/4" SS Swagelok.
  • Recommended Flow Rate: 0.05 to 0.1 l/min (typical)
  • Pressure Drop: Approx. 10 mbar at 0.1 l/min (without valve)

Signal Outputs

  • Concentration: 4 - 20 mA (isolated, active)
  • Concentration: 0 - 10 V (isolated)
  • Concentration Alarms: High Alarm, Low Alarm (latching, not latching)

Control Input

  • Set to zero (24 V, 18 mA, isolated)

Control Outputs (Relay Contacts, 28 V, 0.5 A, isolated)

  • Lamp Low
  • Cuvette Dirty
  • High Alarm
  • Low Alarm
  • Purge Control

Error Handling

  • Error Relay: 30 V, 1 A, summarizing instrument failures.
  • Warnings and Errors: Lamp Low Warning/Error, Lamp Off Error, Cuvette Dirty Warning/Error, Overrange, Overpressure.

Serial Interface

  • RS-232, bidirectional, isolated
  • Baud Rates: 2400 - 38400

Automatic Zeroing

  • Possible via the purge contact.

Software

  • Name: BMT 964 Link
  • Function: Instrument configuration and readout of Event and Error Logs on a Windows PC.

Electrical

  • Power Input: Wide range 100 - 240 VAC, 15 VA
  • Optional DC Power: 12 - 36 VDC, 15 W

Mechanical

  • Dimensions (W x H x D): 360 x 160 x 91 mm
  • Weight: Approx. 4 kg
  • Enclosure: Splash proof aluminium (IP 65, NEMA 4X) with sea water resistant coating (RAL 5009, azure).
  • Mounting: Four mounting brackets provided.
  • Connectors: All electric connectors are waterproof.