Datasheet summary
Ozone Analyzer BMT 964
Engineering summary from PDF text extraction for Ozone Analyzer BMT 964. Verify every value with the OEM datasheet.
Summary
Product Identification
- Product Name: OZONE ANALYZER BMT 964
- Manufacturer: BMT MESSTECHNIK GMBH
- Measurement Principle: Dual beam UV photometer (254 nm), no moving parts
Applications
- Monitoring of ozone generator output
- Monitoring of residual ozone after a process
- Potable water treatment
- Waste water treatment
- Industrial ozone processes
- Semiconductor ozone processes
Operating Limits
- Ambient Temperature: 0 - 50°C (non-condensing)
- Proof Pressure: 1 bar above pressure range
Feed Water / Gas
- Recommended flow rate: 0.1 to 1 l/min typical
- Pressure drop: approx. 3 mbar at 0.5 l/min (with sample gas filter)
- Carrier gas: Programmable as air or oxygen (PSA oxygen)
Product Water / Gas
- Concentration Ranges:
- Standard: 50, 100, 200, 300, 400, 500, 600 g/Nm³
- Optional: 2, 5, 10, 20 g/Nm³
- Selectable Units: g/Nm³, %wt/wt, ppmv
Performance
- Accuracy: After zeroing, the maximum error is the sum of 0.4% of measurement + 0.1% of scale.
- Repeatability error: 0.2% of measurement
- Response time: 0.03 s (analog output), 0.3 s (display)
- Zero drift: typ. 0.2% of range per day, after warm-up, non-cumulative
- MTBF:
- Instrument incl. UV lamp: 65,000 h
- Instrument excl. UV lamp: 120,000 h
- UV Lamp: Low pressure mercury lamp, long-life design, burnt-in for 300 h
- Warranty: 3 years, 5 years on the UV lamp
Mechanical / Hydraulic
- Dimensions (WxHxD): 144 x 72 x 230 mm (DIN 43 700)
- Panel cut-out (W x H): 139 x 67 mm
- Weight: 1.5 kg
- Gas ports: For PTFE tubing 3x5 mm (1/8" x 3/16"), optional 1/8" Swagelok, or 4x6 mm PTFE
- Materials in contact with ozone: Quartz (cuvette windows), Al₂O₃ (cuvette), FFPM (window seals), PTFE (tubing), stainless steel (fittings, cuvette spacer)
- Pressure Compensation: With built-in cuvette pressure transducer
- Absolute pressure range: 1.15 bar (optional 1.5 to 4.0 bar in steps of 0.5 bar)
- Minimum pressure: 0.4 bar abs
- Units selectable: bar, psi, Torr, Mpa
- Built-in sample gas filter: Standard
- Built-in purge unit: Optional (solenoid valve plus air pump) for automatic zeroing
Electrical
- Power Supply:
- 100 - 240 VAC, 15 VA
- Optional: 12 - 36 VDC, 15 W
- Signal Outputs:
- Concentration: 4 - 20 mA (isolated, active)
- Concentration: 0 - 10 V (isolated)
- Control Input: Set to zero (24 V, 18 mA, isolated)
- Control Outputs (Relay Contacts): 28 V, 0.5 A, isolated
- Lamp Low
- Cuvette Dirty
- High Alarm
- Low Alarm
- Purge Control
- Error Relay: 30 V, 1 A, summarizing instrument failures
- Serial Interface: RS-232, bidirectional, isolated, 2400 - 38400 Baud
Display and Control
- Display: 16 character alphanumeric backlit LCD
- Units Displayable: g/Nm³, %wt/wt, ppmv
- Alarms: High and low limit alarms, latching or non-latching
- Control: Key functions programmable via front panel or Windows PC
- Software: BMT 964 Link for instrument configuration and readout of Event and Error Logs
Error Handling
- Failure Warnings/Errors include: Lamp low, cuvette dirty, overrange, summary error.
- Specific errors include: Lamp Low Warning, Lamp Low Error, Lamp Off Error, Cuvette Dirty Warning, Cuvette Dirty Error, Overrange, Overpressure.
Disclaimer: This extraction may miss figures, footnotes, or revisions. Contractual data must match the OEM PDF revision used on the project.
Official datasheet (PDF)
Curated from selected public technical reference material for discovery and preliminary comparison. This summary is not a substitute for a current certified manufacturer datasheet. Verify revisions and design limits before use.