Off-Gas System BMT 964 OG

Engineering summary from PDF text extraction for Off-Gas System BMT 964 OG. Verify every value with the OEM datasheet.

Summary

Product Identification

The OZONE-IN-OFF-GAS ANALYZER is an ozone analysing system for moist ozone off-gas, consisting of:

  • UV-photometric ozone analyser (BMT 964)
  • Peltier-electric sample gas dryer (DH5)
  • Ozone-resistant sample gas pump (SGP-5)

Features

  • Off-gas drying at the sample point
  • Prevents condensation in the sample line
  • Peltier-electric cooling
  • Automatic condensate removal
  • Ozone-resistant sample gas pump
  • Low flow warning
  • Ozone catalyst can be avoided
  • Measurement ranges down to 5 g/Nm³
  • Long-life mercury lamp
  • Warranty: 3 years (5 years on UV lamp)
  • High accuracy: error < 0.5%
  • Pressure and temperature compensated
  • Built-in sample gas filter
  • Built-in purge unit for automatic zeroing
  • Display units: g/Nm³, %wt/wt, or ppmv
  • Timing sequence for automatic zeroing
  • Outputs: 4-20 mA and 0-10 V isolated

Applications

  • Ozone measurement in moist sample gas
  • Municipal off-gas measurement
  • Ozone measurement at atmospheric or sub-atmospheric sample pressure

Specifications

Measurement Principle

  • Dual-beam UV photometer (254 nm), no moving parts

Reliability

  • MTBF (instrument incl. UV lamp): 65,000 h
  • MTBF (excl. UV lamp): 120,000 h

UV Lamp

  • Low pressure mercury lamp, long life design, burnt-in for 300 h

Display

  • 16 character alphanumeric backlit LCD

Concentration Ranges

  • Standard: 50, 100 g/Nm³
  • Optional: 5, 10, 20 g/Nm³
  • Selectable units: g/Nm³, %wt/wt, ppmv

Accuracy

  • Maximum error after zeroing: sum of 0.4% of measurement + 0.1% of scale
  • Repeatability error: 0.2% of measurement
  • Zero drift: typically 0.2% of range per day (after warm-up, non-cumulative)

Operating Limits

  • Maximum inlet pressure: 1 bar above pressure range
  • Ambient temperature (ozone analyser): 0 - 50°C (non-condensing)
  • Ambient temperature (dryer): Contact factory

Materials in Contact with Ozone

  • Quartz (cuvette windows)
  • Al₂O₃ (cuvette)
  • FFPM (window seals)
  • PTFE (tubing)
  • Stainless steel (fittings, cuvette spacer)

Gas Ports

  • Inlet gas port: for 3x5 mm (1/8" x 3/16") tubing; optional 4x6 mm, 1/8'', 1/4'', or 6mm Swagelok
  • Outlet gas port: for 4x6 mm tubing

Flow Rate

  • Typical: 0.5 l/min

Compensation

  • Temperature compensation: Standard
  • Pressure compensation: Built-in cuvette pressure transducer; absolute pressure range 1.15 bara (optional 1.5 or 2.0 bara); selectable units: bar, psi, Torr, MPa

Outputs and Controls

  • Signal outputs:
    • Concentration: 4-20 mA (isolated, active)
    • Concentration: 0-10 V (isolated)
  • Concentration alarms: High Alarm, Low Alarm (latching, non-latching)
  • Control input: Set to zero (24 V, 18 mA, isolated)
  • Control outputs (relay contacts, 28 V, 0.5 A, isolated): Lamp Low, Cuvette Dirty, High Alarm, Low Alarm
  • Error handling (Error Relay, 30 V, 1 A): Summarizes instrument failures
  • Warnings and errors: Lamp Low Warning/Error, Lamp Off Error, Cuvette Dirty Warning/Error, Overrange, Overpressure

Interface and Software

  • Serial interface: RS-232, bidirectional, 2400 - 38400 Baud
  • Automatic zeroing: With internal purge pump and solenoid valve
  • Software: BMT 964 Link (instrument configuration and log readout on Windows PC)

Power

  • Wide range input: 100 - 240 VAC
  • Power consumption: 200 VA

Mechanical

  • Dimensions (W x H x D): 400 x 300 x 120 mm
  • Weight: Approx. 6.5 kg
  • Cabinet: Wall mount aluminum cabinet, yellow powder coating

Compliance

  • CE-marked (EMC & safety)
  • cTUVus NRTL-listed

Sample Gas Pump (SGP-5)

  • Stroke: ± 0.7 mm
  • Motor: Brushless DC motor
  • Performance: 0.5 l/min at zero pressure head; 0.35 l/min against 100 mbar pressure head
  • Systemic pressure restriction: May not be lower than 100 mbar below atmospheric; may be up to 1 bar higher than atmospheric.

Sample Gas Flow Meter

  • Ozone resistant thermal mass flow meter
  • Low flow warning: Activated below approximately 0.3 l/min

Sample Gas Drying (DH5)

  • Drying is achieved by cooling the sample gas using a Peltier cooler.
  • Internal cooler temperature: Approximately +1°C.
  • Function: Removes water vapor to prevent condensation and can also reduce other vapors with dew points higher than the cooler temperature.

Disclaimer This extracted summary may omit figures, footnotes, or revisions. Contractual data must match the official OEM PDF revision used for the project.