Summary
Product identification
- Product Name: Series 100 Ozone Controller
- Manufacturer: Aeroqual Limited
- Series: 100
Operating limits
- Operating Temperature Range: -5°C to 50°C
- Relative Humidity Range: 5% to 95%
- Storage Temperature: Do not store or turn off below 10°C.
- Hazardous Environments: Do not use in or near volatile fuel or chemicals.
Feed water
The device is designed to measure ambient ozone concentration. It should not be placed directly in an ozone stream.
Product water
Not applicable.
Electrical
- Power Requirements: 12VDC, 400mA
- Permissible Voltage Range: 11 VDC (min) to 15 VDC (max)
- Power supply should be sized to account for voltage drop across the Cat 5 cable.
- Typical resistance of Cat 5 cable: 9.38 Ohms per 100m at 20ºC.
- Unit Interface (power, control signal): RJ45 connector
- External signal for control & switching: 12 VDC, 150 mA
Mechanical / hydraulic
- Sensor Heads: Removable / replaceable
- High Concentration Ozone Sensor Head:
- Measurement Range: 0.00 to 5.00 ppm
- Accuracy: ±10% between 0.50 ppm and 5.00 ppm
- T90 Response Time: 30 seconds
- Note: Exposure to concentrations higher than 5.00 ppm may affect calibration.
- Low Concentration Ozone Sensor Head:
- Measurement Range: 0.000 to 0.500 ppm
- Accuracy: ±0.010 ppm from 0 to 0.100 ppm; ±10% from 0.100 to 0.500 ppm
- Resolution: 0.001 ppm
- T90 Response Time: 80 seconds
- Note: Exposure to concentrations higher than 0.500 ppm may affect calibration.
- High Concentration Ozone Sensor Head:
- Controller Dimensions:
- Height (with sensor head): 64 mm
- Diameter of base: 130 mm
- Components: Series 100 controller base, Ozone sensor head, User guide, 3 rubber pads, 4 Velcro dots.
Specifications – Series 100
- Sensor type: Gas Sensitive Semiconductor (GSS)
- High/ low control or set points: 16 position rotary switch