Datasheet summary
S900
Engineering summary from PDF text extraction for S900. Verify every value with the OEM datasheet.
Summary
Product identification
- Series: 900 Transmitter / Controller
- Sensor Heads: Ozone (Low, High, Leak), VOC, Ammonia, CO
Operating limits
- Operating temperature range (Sensor head and base unit): -5°C to 50°C; 23°F to 122°F
- Relative humidity limit (Sensor head and base unit): 95% maximum
Ozone Sensor Heads
- Low Measurement range: 0.000 - 0.500 ppm
- Low Accuracy: < 0.008 ppm (0 - 0.100 ppm); ± 10% (0.100 – 0.500 ppm)
- Low Resolution: 0.001 ppm
- Low T90 Response: < 60 seconds
- High Measurement range: 0.50 - 20.00 ppm
- High Accuracy: ± 10% (0.20 - 2.00 ppm); ± 15% (2.00 - 20.00 ppm)
- High Resolution: 0.01 ppm
- High T90 Response: < 35 seconds
- Leak Measurement range: 0.00 - 50.00 ppm
- Leak Accuracy: ± 20%, 15 seconds after Reset
- Leak Resolution: 0.01 ppm
- Leak T90 Response: < 10 seconds
- Sensor type: Gas-sensitive semiconductor
VOC Sensor Head Specification
- Measurement range: 0 to 400 ppm toluene
- Accuracy: < +/- 10 ppm (0 – 200 ppm); < +/- 10% ( > 200 ppm)
- T90 response: < 60 s
- Sensor type: Gas-sensitive semiconductor
- Operating temperature range: -20°C to 60°C
- Operating relative humidity range: 5 to 95% non-condensating
Ammonia Sensor Head Specification
- Measurement range: 0 to 1000 ppm
- Accuracy: < +/- 5 ppm (0 – 100 ppm); < +/- 10% (100 – 1000 ppm)
- T90 response: < 60 s
- Sensor type: Gas-sensitive semiconductor
- Operating temperature range: -20°C to 60°C
- Operating relative humidity range: 5 to 95% non-condensating
CO Sensor Head Specification
- Measurement range: 0 to 2000 ppm carbon monoxide
- Accuracy: < +/- 10 ppm in the range 0 to 400 ppm
- T90 response: < 150s
- Sensor type: Gas-sensitive semiconductor
- Operating temperature range: 0°C to 70°C
- Operating relative humidity range: 5 to 95% non-condensating
Electrical
- Power (user supplied): Regulated 12 VDC, 800 mA
- Analogue output: 4-20 mA (opto-isolated), 10-30 V
- External signal type: Transistor Output
- External signals: 4
- External signal functions: Low Alarm, High Alarm, Control & Diagnostics
- Inputs: Standby toggle
- Communication: RS485
- RS485 protocol: Aeroqual proprietary product
Mechanical / hydraulic
- Removable / replaceable sensor head: Yes
- Enclosure rating: IP20 & NEMA 1 equivalent
- Enclosure size (with base unit): 64 H x 130 Ø (mm); 2½ H x 5⅛Ø (in)
- Mounting: Screw fixture
- Weight: < 200 g; < 7 oz
Controls and Configuration
- Jumpers: J1, J2, J3 termination resistors
- Connectors: Screw
- ID: 1 (Default); User configurable from 1 - 255
- Alarm set points (2): User configurable
- Control set point: User configurable
Approvals
- Base Unit: Part 15 of the FCC Rules, EN 50082-1: 1997, EN 50081-1: 1992
- CO Sensor Head: UL 2034, BS 7860
Disclaimer: This extraction may miss figures, footnotes, or revisions. Contractual data must match the OEM PDF revision used on the project.
Official datasheet (PDF)
Curated from selected public technical reference material for discovery and preliminary comparison. This summary is not a substitute for a current certified manufacturer datasheet. Verify revisions and design limits before use.