Back to The AQL Ozone family
Aeroqual logo

Datasheet summary

Series 930 - 935

Engineering summary from PDF text extraction for Series 930 - 935. Verify every value with the OEM datasheet.

Summary

Product identification

  • Product line: AQL Ozone family
  • Series: 930 | 935 Transmitter & Controller
  • Specific model mentioned: Series 935 Ozone Monitor with temperature & humidity sensor

Features

  • Active sampling for higher accuracy
  • Tamperproof & water resistant enclosure
  • Multiple analog & digital outputs
  • Real-time network capability
  • Large LED local display option
  • Temperature & humidity sensor (S935)
  • PC data logging & networking option

Applications

  • Aggressive ambient environments
  • Gas alarm & control applications
  • Indoor air quality monitoring
  • Real-time network monitoring

Ozone Sensor Heads

Sensor HeadCalibrated RangeMaximum ExposureLowest Detection LimitAccuracy of CalibrationResolutionResponse Time (T90)
Ultra low0 - 0.150 ppm0.25 ppm1 ppb<± 0.005 ppm0.001 ppm< 70 s
Low0 - 0.5 ppm1.0 ppm1 ppb<± 0.008 < 0.1 ppm; <± 10% 0.1 - 0.5 ppm0.001 ppm< 60 s
High0.5 - 20 ppm25 ppm10 ppb<± 10% 0.2 - 2 ppm; <± 15% 2 - 20 ppm0.01 ppm< 60 s

Operating Conditions

  • Temperature: -5°C to 40°C
  • Humidity: 5 to 95% RH (non-condensating)

Electrical

  • Power: 24VDC, 500mA (range 22-26 VDC) - user supplied
  • Analogue output: 4-20mA (opto-isolated), 12-24 V
  • Communication: RS485 (Aeroqual proprietary protocol)
    • Jumpers J1, J2, J3 for termination resistors
    • ID: 1 (Default), User configurable from 1 - 255

Mechanical / hydraulic

  • Removable & replaceable sensor head: Yes
  • Temperature & humidity sensor (Series 935):
    • Temperature Range: -40°C to 124°C (-40°F to 255°F)
    • Humidity Range: 0 to 100% RH
  • Enclosure rating: IP41 & NEMA 2 equivalent
  • Enclosure casing: Fibre reinforced polycarbonate
  • Enclosure size: 180W x 110H x 90D (mm); 7⅛W x 4¾H x 3½D (in)
  • Weight: < 850 g; < 30 oz

Other features

  • External signal type: Transistor output
  • External signals (4): Low Alarm | High Alarm | Control | Diagnostics
  • LED Display: Optional
  • Inputs: Standby toggle
  • Alarm set points (2): User configurable
  • Control set point: User configurable
  • Software (standard): PC Configuration Program CD
  • Software (optional): PC Data Logging & Networking CD (code: R60)
  • Approvals: Part 15 of FCC Rules, EN 50082-1: 1997, EN 50081-1: 1992

Disclaimer: This extraction may miss figures, footnotes, or revisions; contractual data must match the OEM PDF revision used on the project.