Summary
Product Identification
Product Family: AQL Ozone Model: Series 940 Transmitter / Controller
Base Unit Features
- Active sampling via Teflon tube with vacuum pump
- Water ingress protection
- Allows remote sampling from inaccessible areas (e.g., controlled atmosphere environments)
- Improved performance in fluctuating pressure and flow environments
Typical Applications
- Short distance remote monitoring for low oxygen environments (3-5% O₂)
- Indoor air quality – duct sampling
- Filtered air sampling
Electrical
- Power (user supplied): 24VDC, 500mA (range 22-26 VDC)
- Analogue Output: 4-20 mA (opto-isolated), 12-24 V
- External Signal Type: Transistor Output
- External Signals: 4
- External Signal Functions: Low Alarm, High Alarm, Control & Diagnostics (Optional)
- Communication: RS485 (Aeroqual proprietary protocol)
Inputs
- Standby toggle
- User configurable ID (1 - 255)
Parameters
- Alarm Set Points: 2 (user configurable)
- Control Set Point: 1 (user configurable)
Mechanical / Hydraulic
- Removable / Replaceable Sensor Head: Yes
- Sampling Pump: 12VDC, long life, brushless, diaphragm pump
- Enclosure Rating: IP41 & NEMA 2 equivalent
- Enclosure Size: 230W x 140H x 95D mm (9”W x 5½”H x 3¾”D in)
- Enclosure Casing: Fibre reinforced polycarbonate
- Weight: < 1100 g (< 1 lb 3 oz)
- Connectors: Screw
- Jumpers: J1, J2, J3 termination resistors
Sensor Head Specifications
Ozone Sensor Heads
| Specification | Low Range | High Range | Leak Range |
|---|---|---|---|
| Measurement Range | 0.000 - 0.500 ppm | 0.50 - 20.00 ppm | 0.00 - 50.00 ppm |
| Accuracy | < 0.008 ppm (0-0.100) | ± 10% (0.20 - 2.00) | ± 20%, 15s after Reset |
| ± 10% (0.100-0.500) | ± 15% (2.00 - 20.00) | ||
| Resolution | 0.001 ppm | 0.01 ppm | 0.01 ppm |
| T90 Response | < 60 seconds | < 35 seconds | < 10 seconds |
| Sensor Type | Gas-sensitive semiconductor | Gas-sensitive semiconductor | Gas-sensitive semiconductor |
- Controlled Atmosphere Version: Available on Request for all Ozone Sensor Heads.
VOC Sensor Head Specification*
- Measurement Range: 0 to 400 ppm toluene
- Accuracy: < +/- 10 ppm (0 – 200 ppm); < +/- 10% ( > 200 ppm)
- T90 Response: < 60 s
- Sensor Type: Gas-sensitive semiconductor
- Operating Temperature Range: -20°C to 60°C
- Operating Relative Humidity Range: 5 to 95% non-condensating
- * Specific Calibrations to other VOC’s available – contact Aeroqual Sales.
Ammonia Sensor Head Specification
- Measurement Range: 0 to 1000 ppm
- Accuracy: < +/- 5 ppm (0 – 100 ppm); < +/- 10% (100 – 1000 ppm)
- T90 Response: < 60 s
- Sensor Type: Gas-sensitive semiconductor
- Operating Temperature Range: -20°C to 60°C
- Operating Relative Humidity Range: 5 to 95% non-condensating
CO Sensor Head Specification
- Measurement Range: 0 to 2000 ppm carbon monoxide
- Accuracy: < +/- 10 ppm in the range 0 to 400 ppm
- T90 Response: < 150s
- Sensor Type: Gas-sensitive semiconductor
- Operating Temperature Range: 0°C to 70°C
- Operating Relative Humidity Range: 5 to 95% non-condensating
Environmental / Operating Limits
- Operating Temperature (Sensor Head & Base Unit): -5°C to 50°C (23°F to 122°F)
- Relative Humidity Limit (Sensor Head & Base Unit): 95% maximum
Approvals
- Part 15 of the FCC Rules
- EN 50082-1: 1997
- EN 50081-1: 1992
- UL 2034, BS 7860 (CO Sensor Head only)