Datasheet summary
Series 130s
Engineering summary from PDF text extraction for Series 130s. Verify every value with the OEM datasheet.
Summary
Product identification
Series 130s Ozone Controller
Features
- Active sampling for accuracy
- LED ozone level indicator
- Low service maintenance
- Replaceable sensor head
- Tamperproof enclosure
- Water & dust resistant
- Alarm & control outputs
- Simple plug & play
Applications
- Ozone generator control
- Industrial monitoring
- Health & safety
- Alarm systems
Specifications
Ozone Sensor Head
| Feature | Low Range | High Range |
|---|---|---|
| Calibrated Range | 0-0.5 ppm | 0-10 ppm |
| Maximum Exposure | 1 ppm | 15 ppm |
| Lowest Detection Limit | 0.001 ppm | 0.01 ppm |
| Accuracy of Calibration | <±0.008 ppm (0-0.1 ppm); <±10% (0.1-0.5 ppm) | <±0.1 ppm (0-1 ppm); <±15% (1-10 ppm) |
| Response Time (T90) | < 60 s | < 60 s |
Operating Conditions
- Temperature: -5 to 40°C
- Humidity: 5 to 95% RH non-condensing
Electrical
- Power supply: 12VDC (switching power adaptor 100-240VAC supplied)
- I/O connectors (power, control signal): Screw terminals and cable gland
- Setting control and alarms: Dip switches (Alarm/Control Mode and relay/alarm set point)
- External signal for control and switching
- Relays: 2 relays, 28V max, 150mA max
- External signal functions: Low Alarm, High Alarm, Control
Mechanical / Hydraulic
- Removable / replaceable sensor head: Yes (sensor head can also be refurbished)
- Audible alarm: Piezoelectric (active in Alarm Mode)
- LED status indicator (tri-colour): Displays Alarm levels, Control band, Diagnostics
- Yellow: Ozone concentration is below low alarm limit
- Green: Ozone concentration is between low and high limits
- Red: Ozone concentration is above high alarm limit
- Enclosure casing and rating: Fibre reinforced plastic; IP41 & NEMA 2 equivalent
- Enclosure size: 180W x 110H x 90D mm (7⅛W x 4¾H x 3½D in)
- Weight: < 850 g (< 30 oz)
Approvals
- Part 15 of FCC Rules
- EN 50082-1: 1997
- EN 50081-1: 1992
Disclaimer: This extracted summary may miss figures, footnotes, or revisions. Contractual data must match the OEM PDF revision used on the project.