Back to The AQL Ozone family
Aeroqual logo

Datasheet summary

SM 50

Engineering summary from PDF text extraction for SM 50. Verify every value with the OEM datasheet.

Summary

Product Identification

Aeroqual SM50 OEM Ozone Sensor Module

Software Options

Two software options are available:

  • Switching Device: For simple on/off switching of equipment. (Not installed)
  • Control Device: Maintains a specific gas concentration between user-defined levels by controlling external devices via the on-board relay. (Installed)

Software is loaded prior to delivery and can only be altered by returning to Aeroqual Limited.

Ozone Sensors

SensorCalibrated RangeAccuracyResolution0-5V Analog Output
Ozone ultra low (OZU)0 - 0.150 ppm<± 0.010 ppm0.001 ppm0 - 0.150 ppm
Ozone low (OZL)0 - 0.5 ppm<± 0.02 ppm (0 - 0.1 ppm)0.001 ppm0 - 0.5 ppm
<± 20% (> 0.1ppm)
Ozone high (OZH)0.5 - 20 ppm<± 20%0.01 ppm0 - 20 ppm

Operating Instructions

Relay Set Point Configuration

The Relay and Alarm Set Point is factory set by default, but can be adjusted using dip-switches. The table below shows example set points for different ozone sensor types.

Relay dipswitch (1 2 3 4)O₃ Ultra-Low (ppm)O₃ Low (ppm)O₃ High (ppm)
on on on on0.0000.0000
off on on on0.0100.0251
on off on on0.0200.0502
off off on on0.0300.0753
on on off on0.0400.1004
off on off on *0.0500.1255
on off off on0.0600.1506
off off off on0.0700.1757
on on on off0.0800.2008
off on on off0.0900.2259
on off on off0.1000.25010
off off on off0.1100.30012
on on off off0.1200.35014
off on off off0.1300.40016
on off off off0.1400.45018
off off off off0.1500.50020

* Default factory setting

Status LED / Diagnostics

  • Startup: Flashes green 2-6 times at 0.5-second intervals.
  • Warm-up (3-10 minutes): Flashes green at 2-second intervals.
  • Normal Operation: Solid green.
  • Sensor Failure: Flashes green rapidly at 0.3-second intervals.

Relay LED

Glows red when the gas concentration reaches the "Relay-Set-Point" and the relay is activated.

Using the Relay Output

The relay has volt-free contacts (NO, NC, COM) rated for a maximum of 30V / 2A. For higher loads, an external relay is required.

  • Normally Open (NO) Relay: Closes when the set point is reached, activating the external device. Switches off when the concentration drops below the set point. Fail-safe if power fails.
  • Normally Closed (NC) Relay: Opens when the set point is reached, deactivating the external device. Switches on when the concentration drops below the set point. Not fail-safe.

Operation as a Control Device

This mode requires specialized software loaded at the factory. The SM50 maintains gas concentration within a "dead band" (± 10% of the set point).

  • When concentration rises from below the "Lower Limit" to the "Upper Limit", the relay remains closed.
  • When concentration falls from above the "Upper Limit" to the "Lower Limit", the relay remains open.

Electrical

  • Input Power: 11 - 24 VDC
  • Power Consumption: 2.5 - 6 W max
  • Analog Output: 0-5V, 12-bit resolution.

Mechanical / Hydraulic

  • Board Size: 60 mm x 75 mm
  • Mounting: Screw or extrusion slot.
  • Fan: Onboard ball-bearing, 50,000 hours life.
  • Sensor Filter: Onboard.

Options

  • Display: LCD (3.5 digit) or VFD (2-line/8-column) on separate boards.
  • Piezo Alarm: Onboard or loose.
  • Communication: RS232 or RS485 (onboard option).
  • Temperature Sensor: Range -40 to 120 ºC, Accuracy ± 0.3 ºC, Resolution 0.01 ºC.
  • Humidity Sensor: Range 0 to 100% RH, Accuracy ± 2% RH, Resolution 1% RH.

Environmental

  • Operating Temperature: 0 to 50 ºC (or -20 to 50 ºC if enclosed).
  • Operating Humidity: 5 to 95% RH (non-condensating).

Warranty

1 year limited warranty (excludes calibration).


Disclaimer: This extraction may miss figures, footnotes, or revisions. Contractual data must match the OEM PDF revision used on the project.