Back to The AQL Ozone family
Aeroqual logo

Datasheet summary

SM 60

Engineering summary from PDF text extraction for SM 60. Verify every value with the OEM datasheet.

Summary

Product identification

Product: SM60 Ozone Controller | Switch Family: AQL Ozone family

Features

  • Accurate control of ozone in air
  • Ideal controller for ozone generators
  • Low sensor drift via active sampling
  • Large LED status indicators
  • Adjustable relay set point
  • Relay switch outputs
  • Analog 0-5V output
  • RS 232 | RS 485 outputs
  • Robust and reliable

Specifications

  • Sensor type: Analytic GSS Technology ® Gas Sensitive Semiconductor
  • Sampling method: Active sampling via fan
  • Operational mode: Continuous control
  • Warm up time: 10 minutes (max. accuracy)
  • LED status indicators: Green = sensor ON/OFF | Amber = relay energized (i.e. O₃ generator ON/OFF)
  • Relay set point: Adjustable via dipswitch (OZL default = 0.1 ppm)
  • Control band: 10% hysteresis (avoids O₃ generator “hunting”)
  • Digital outputs: RS 232 | RS 485
  • Analog output: 0-5 VDC (8 bit)
  • Relay outputs: Normally Open | Normally Closed | Common
  • Primary relay rating: 24V | 2A (use with secondary relay for switching high voltage loads)
  • Power supply: 12 VDC | 800 mA | Plug-in AC power adaptor supplied
  • Enclosure rating: IP20 & NEMA 1 equivalent
  • Enclosure dimensions: 130 W x 94 H x 57 D (mm) | 5 ⅛ W x 3 ¾ H x 2 ¼ D (in)
  • Enclosure casing: Flame resistant thermoplastic PS
  • Enclosure mounting: Screw fix
  • Weight: 225 g | 8 oz (excludes AC power adaptor)
  • Approvals: Part 15 of FCC Rules, EN 61000-6-3: 2001, EN 61000-6-1: 2001

Ozone Sensors

SensorCalibrated RangeMaximum Exposure0-5V Output ScaleAccuracyOutput ResolutionResponse Time T90Operating Conditions
OZL0-0.5 ppm1 ppm0-0.500<± 15%0.002 ppm< 60 s-5°C to 40°C
OZH0.5-20 ppm25 ppm0-50<± 15%0.2 ppm< 60 s5 to 95% RH*

* Non-condensing

Disclaimer: This extraction may miss figures, footnotes, or revisions. Contractual data must match the OEM PDF revision used on the project.

Official datasheet (PDF)

Lenntech datasheet mirror

Curated from selected public technical reference material for discovery and preliminary comparison. This summary is not a substitute for a current certified manufacturer datasheet. Verify revisions and design limits before use.