UV-H

Engineering summary from PDF text extraction for UV-H. Verify every value with the OEM datasheet.

Summary

Product identification

The Aeroqual UV-H Photometer Instrument is a portable, single-beam UV photometer designed for ozone measurement.

Operating limits

  • Measurement range: 0-200 ppm (or 0-400 mg/m³) Ozone
  • Environmental Temperature: 0-40°C
  • Relative Humidity: 0-95% RH (non-condensating)

Feed water

  • Sampling method: Brushless DC diaphragm pump
  • Sample flow rate: 0.4 ±0.05 LPM
  • Inlet filter: 5 µm pore size, 37 mm PTFE filter
  • Pneumatic connections: Kynar ¼” compression fittings
  • Wetted materials: Kynar, Teflon PFA, quartz, aluminium, Viton

Product water

  • Precision: <±5% of reading above 0.2 ppm; ±0.01 ppm below 0.2 ppm
  • Resolution: 0.01 ppm
  • T₉₀: < 60s

Electrical

  • Power supply: 24 VDC Switch mode power supply (Input: 90-260V AC; 47-63Hz)
  • Digital interface: RS232 as standard
  • Optional outputs: 0-5V DC analog output (12 bit), setpoint-controlled 30V DC 2A relay

Mechanical / hydraulic

  • Display: 4-line x 20-character VFD
  • Data logging: 2 GB SD card
  • Enclosure: Metal instrument case, dimensions 378 x 236 x 132 mm
  • Software: PC data logging software supplied