SYCON 2702 en L

Engineering summary from PDF text extraction for SYCON 2702 en L. Verify every value with the OEM datasheet.

Summary

Product Identification

The SYCON 2702 is an online analyzer that automatically determines water parameters using the colorimetric titration method. It analyzes one parameter at a time, with the parameter and measuring range determined by the specific indicator used.

Operating Limits

  • Storage Temperature: 0 °C – 45 °C
  • Ambient Temperature: 10 °C – 45 °C
  • Measured Water Temperature: 5 °C – 40 °C
  • Air Humidity: 20 – 90 % RH (without ice or condensation)

Feed Water

  • Inlet Water Pressure:
    • Minimum: 0.5 bar
    • Maximum: 5 bar
    • Recommended: 1 - 2 bar
  • General Inlet Water Quality: Clear, colourless, free of solids, and without gas bubbles.
  • Specific Water Quality Requirements for Measuring Water Hardness:
    • pH: 4 - 10
    • Iron: < 3 ppm
    • Copper: < 0.2 ppm
    • Aluminium: < 0.1 ppm
    • Manganese: < 0.2 ppm
    • Acid capacity: Kₛ₄.₃ < 5 mmol/l
  • Note on Oxidising Agents: Oxidising agents (e.g., chlorine, ozone) above TrinkwV 2012 limits can interfere with measurements. An upstream activated carbon filter is recommended.

Product Water Analysis

  • Measurement Method: Titration with colour change.
  • Available Units: °dH, °f, ppm CaCO₃, mmol/l, °e, °Ж.
  • Measuring Ranges (Examples):
    • Total Hardness:
      • °dH: 0.012 - 30
      • °f: 0.021 - 53.6
      • ppm CaCO₃: 0.21 - 535.7
      • mmol/l: 0.0021 - 5.36
    • Carbonate Hardness / Alkalinity:
      • °dH: 0.3 - 22.5
      • °f: 0.5 - 40.2
      • ppm CaCO₃: 5.4 - 401.8
      • mmol/l: 0.11 - 8.04
  • Accuracy:
    • Measurement Accuracy: ± 5% of the maximum measured value of the indicator used.
    • Repeatability: ± 2.5% of the maximum measured value of the indicator used.
  • Indicator Consumption: Approx. 0.05 - 0.5 ml / analysis (depends on water hardness).
  • Measuring Time: Approx. 3 min (depends on water hardness and rinsing time).
  • Indicator Shelf Life: At least 2 years if stored properly (below 25 °C, in a dark place).
  • Water Consumption: Approx. 1 l/analysis at 2 bar (varies with inlet pressure and rinsing time).

Electrical

  • Power Supply:
    • 230 VAC version: 100 - 240 VAC (50/60 Hz)
    • 24 V AC/DC version: 21.6 – 26.4 V AC/DC (50 Hz)
  • Power Consumption:
    • Maximum: 25 VA
    • Measurement Mode: 8 VA
    • Standby: 3.5 VA
  • Relay Load Capacity:
    • With internal power supply (terminals 5 to 8): 1 A
    • With external power supply: 2.5 A
  • Relay Outputs: 2 potential-free outputs (NC/NO), max. 250 VAC / VDC; max 1 A. Functions include limit value alarm and device error.
  • Signal Input: Galvanically isolated contact input for start analysis, flow monitor, or interval reset.
  • Analogue Output: 0 – 20 mA / 4 – 20 mA, resolution < 100 μA, max. load 750 Ω.

Mechanical / Hydraulic

  • Protection:
    • Open wall mounting: IP43
    • Mounting in housing: IP54
  • Dimensions:
    • Without housing: 274 x 275 x 129 mm (W x H x D)
    • With housing: 332 x 345 x 190 mm (W x H x D)
  • Weight:
    • Without housing: approx. 1.9 kg
    • With housing: approx. 3.8 kg
  • Installation: Upright wall mounting in closed rooms.
  • Connections: Water inlet and drain connections for hose with 6 mm outer diameter.

Maintenance

  • Every 6 months: Cleaning the measuring chamber. Interval may need shortening at high ambient/water temperatures or with high organic content.
  • Every 30,000 analyses or after 2 years: Cleaning the measuring chamber, installation of maintenance kit (peristaltic pump cartridge and seals replacement).

Disclaimer: This extraction may miss figures, footnotes, or revisions. Contractual data must match the OEM PDF revision used on the project.