Datasheet summary
SYCON 2802 en L
Engineering summary from PDF text extraction for SYCON 2802 en L. Verify every value with the OEM datasheet.
Summary
Product identification
The SYCON 2802 is an online analyzer based on the SYCON series, utilizing the colorimetric titration method for automatic determination of water parameters. It can determine one parameter at a time, with the specific parameter and measuring range dependent on the indicator used.
General specifications
- Power:
- 230 VAC version: 100 – 240 VAC (50/60 Hz)
- 24 V AC/DC version: 21.6 – 26.4 V AC/DC (50 Hz)
- Power consumption:
- Maximum: 25 VA
- Measurement mode: 8 VA
- Standby: 3.5 VA
- Load capacity:
- Relays with internal power supply (terminals 1 to 4): 1 A
- Relays with external power supply: 2.5 A
- Protection:
- Open wall mounting: IP43
- Mounting in housing: IP54
- Storage temperature: 0 °C – 45 °C
- Ambient temperature: 10 °C – 45 °C
- Measured water temperature: 5 °C – 40 °C
- Air humidity: 20 – 90 % RH (without ice or condensation)
- Inlet water pressure: min: 0.5 bar - max: 5 bar (recommendation 1 - 2 bar)
- Inlet water general: Clear, colourless, free of solids, without gas bubbles.
Water quality requirements for measuring water hardness
- pH: 4 - 10
- Iron: < 3 ppm
- Copper: < 0.2 ppm
- Aluminium: < 0.1 ppm
- Manganese: < 0.2 ppm
- Acid capacity: KS₄.₃ < 5 mmol/l
Note on oxidising agents: Oxidising agents above permitted limits can affect the indicator dye and interfere with measurements. An activated carbon filter upstream of the device can remove these agents. The filter requires regular replacement.
Technical data
Installation
- Type: Wall mounting in closed rooms
- Dimensions (without housing): 274 x 275 x 129 mm (W x H x D)
- Dimensions (with housing): 332 x 345 x 190 mm (W x H x D)
- Weight (without housing): approx. 2.0 kg
- Weight (with housing): approx. 4.0 kg
Analysis characteristics
- Measurement method: Titration with colour change
- Measuring range: The measuring range is defined by the indicator used.
- Total hardness:
- °dH: 0.012 - 30
- °f: 0.021 - 53.6
- ppm CaCO₃: 0.21 - 535.7
- mmol/l: 0.0021 - 5.36
- °e: 0.015 - 38.6
- °Ж: 0.0043 - 10.7
- Carbonate hardness:
- °dH: 0.3 - 22.5
- °f: 0.5 - 40.2
- ppm CaCO₃: 5.4 - 401.8
- mmol/l: 0.11 - 8.04
- °e: 0.39 - 28.9
- °Ж: 0.21 - 16.1
- Total hardness:
- Accuracy:
- Measurement accuracy: ± 5% of the maximum measured value of the indicator used
- Repeatability: ± 2.5% of the maximum measured value of the indicator used
- Current interface: ± 0.3 mA
- Indicator consumption: Approx. 0.05 - 0.5 ml / analysis (depending on water hardness)
- Measuring time: Approx. 3 min (depending on water hardness and rinsing time)
- Number of analyses per indicator bottle: Approx. 10,000 analyses / 500 ml indicator for soft water. Consumption depends on hardness and indicator.
- Shelf life of indicators: At least 2 years if stored properly (below 25 °C, in a dark place).
- Water consumption: Approx. 1 l/analysis at 2 bar. Varies with inlet pressure and rinsing time.
Inputs/outputs (24 V AC/DC and 230 VAC versions)
- Relay outputs: 4 relays, max. 250 VAC / VDC 1 A (potential-free, NC/NO). Functions include limit value alarms, device error, analysis active, indicator message, sample pump, and maintenance interval.
- Signal inputs: 2 galvanically isolated contact inputs. Functions include start analysis, water meter, flow monitor, interval reset, and error reset.
- Analogue output: 0 – 20 mA / 4 – 20 mA (Resolution: < 100 μA, Max. load: 750 Ω).
- RS-485: Modbus RTU protocol.
Maintenance intervals
- Every 6 months: Cleaning the measuring chamber. Intervals may need shortening with high ambient/water temperatures or high organic matter.
- Every 30,000 analyses or after 2 years of operation: Cleaning the measuring chamber, installation of maintenance kit (peristaltic pump cartridge and seals replacement), and resetting the maintenance counter.
Indicators for monitoring total hardness
See the official PDF for exact tabulated values. Available units for analysis value include: °dH, °f, ppm CaCO₃, mmol/l, °e, °Ж.
Indicators for monitoring carbonate hardness
See the official PDF for exact tabulated values. Available units for analysis value include: °dH, mmol/l, °f.
Note: The measured value should be close to the middle of the indicator's measuring range. One bottle of indicator is sufficient for approx. 10,000 analyses in the lower range of the indicator type.
Mounting
The SYCON 2802 must be installed upright.
- Without housing: Wall bracket has four 6 mm holes.
- With housing: Supplied with 4 mounting brackets with 6.5 mm holes. Required space for opening: at least 450 x 350 mm (W x H).
Connections:
- A: Cable glands for electrical connection
- B: Water inlet connection (hose with 6 mm outer diameter)
- C: Water drain connection (hose with 6 mm outer diameter)
Disclaimer: This extraction may miss figures, footnotes, or revisions. Contractual data must match the OEM PDF revision used on the project.