Datasheet summary
Testomat 2000 CLF
Engineering summary from PDF text extraction for Testomat 2000 CLF. Verify every value with the OEM datasheet.
Summary
Product Identification
- Product Name: Testomat 2000® CL T, Testomat 2000® CL F
- Type: Process Photometer for total chlorine (CL T) or free chlorine (CL F)
Operating Limits
- Total Chlorine Measuring Range: 0 – 2.5 ppm
- Free Chlorine Measuring Range: 0 – 2.5 ppm
- Operating Pressure Range: 0.3 to 8 bar (with flow controller valve body removed for 0.3 to 1 bar range)
- Measuring Water Temperature: 10 °C to 40 °C (for temperatures above 40 °C, a KCN type cooler is recommended)
Feed Water
- Requirements: Must be clear, colourless, and free of undissolved particles.
- Maximum CO₂ Concentration: Incorrect evaluations may occur if CO₂ concentration exceeds 20 mg/l.
Product Water
- Analysis Principle: Wet-chemical online monitor using the photometric analysis principle, based on the DPD method according to EN ISO 7393-2.
- Measurement Time: Approximately 1 minute reaction time, plus analysis and flush times.
- Reagents: Requires specific Heyl Testomat 2000® reagents (A and B for free chlorine; A, B, and C for total chlorine).
Electrical
- Power Supply: Requires an interference-free and stable power supply. Mains voltage specified on the rating plate (e.g., 24 V / 115 V / 230 V).
- Relay Outputs: Max. 240 VAC, 4 A. Used for:
- External flush valve (Flush OUT)
- Limit value output 1 (LV1 OUT)
- Limit value output 2 (LV2 OUT)
- Measuring point switch-over (M. point. 1/2 OUT)
- Universal output (AUX OUT)
- Fault message output (Alarm OUT)
- Maintenance message output (Maintenance OUT)
- Inputs:
- External analysis start (Start IN)
- External analysis stop (Stop IN)
- Universal input 1 (IN1 IN)
- Universal input 2 / water meter (IN2 IN)
- Optional Interface: 0/4-20 mA or 0/2-10 V, or RS 232 for log printer.
Mechanical / Hydraulic
- Water Inlet Connection: Standard plug connector for plastic hoses 6/4 x 1 (6 mm outer diameter / 4 mm inner diameter). Maximum inlet hose length of 5 m recommended.
- Water Outlet Connection: Requires an opaque outlet hose with an internal diameter of 14 mm. Hose must be laid without backwater or siphoning effect.
- Reagent Storage: 500 ml reagent bottles.
- Internal Design: Includes terminal strips, mains switch, dosing pump, controller/filter receiver, and measuring chamber.
Operation
- Modes: Time control, quantity control via water meter, dynamic analysis start, and external analysis start.
- Analysis Interval: Programmable from 0 to 99 minutes.
- Limit Value Monitoring: Two independent limit values (LV1, LV2) with programmable hysteresis (1, 2, or 3 bad analyses).
- Display: 4-line text display showing measured values, statuses, and programming data. Includes status indicators for limit values, alarm, and maintenance.
- Reagent Handling: Requires careful insertion and extraction of reagents.
- Settings: Menu-driven operation via a keypad and display. Basic program requires a four-digit password (current time in reverse order).
Disclaimer: This extraction may miss figures, footnotes, or revisions. Contractual data must match the OEM PDF revision used on the project.