Datasheet summary
Testomat 2000 CLT
Engineering summary from PDF text extraction for Testomat 2000 CLT. Verify every value with the OEM datasheet.
Summary
Product Identification
- Product Name: Testomat 2000® CLT self clean
- Type: Process Photometer for total chlorine
Operating Limits
- Total Chlorine Measurement Range: 0 – 2.5 ppm (or mg/l)
- Operating Pressure: 0.3 to 8 bar (factory setting), adjustable to 0.3 to 1 bar by removing flow controller valve body.
- Water Temperature: 10 °C to 40 °C. For temperatures above 40 °C, a KCN type cooler should be installed.
Feed Water
- Must be clear, colorless, and free of undissolved particles.
- Maximum recommended CO₂ content: 20 mg/l.
Product Water
- The instrument monitors total chlorine concentration.
- Applications include monitoring disinfectant dosing in the drinking water field.
Electrical
- Mains Voltage Input: 24 V / 115 V / 230 V AC.
- Power Consumption: Not specified.
- Relay Outputs: Max. 240 VAC, 4 A.
Mechanical / Hydraulic
- Water Inlet Connection: Standard plug connector for plastic hoses 6/4 x 1 (external diameter 6 mm / internal diameter 4 mm).
- Water Outlet Hose Diameter: Internal diameter 12 mm.
- Reagent Bottle Storage: 500 ml.
Performance Specifications
- Measurement Principle: Wet-chemical online monitor using the photometric analysis principle, based on the DPD method (EN ISO 7393-2).
- Analysis Process: Involves adding three reagents, with a reaction time of approximately 1 minute (excluding flush time).
- Cleaning Function: Measuring chamber is cleaned after a pre-set number of analyses.
- Key Features:
- Menu-driven operation with plain text display.
- Freely selectable units (ppm or mg/l).
- Precise piston-dosing pump.
- Analysis initiation options: Automatic interval, external control, volume controlled (water meter).
- Two independent limit values with hysteresis.
- Monitoring of two measuring points (via external solenoid valves).
- Internal error documentation.
- Programmable service address and maintenance interval.
- Optional interfaces: 0/4-20 mA or 0/2-10 V, or RS 232 for a log printer.
Safety Information
- Read operating instructions carefully and completely before working with the instrument.
- Adhere to hazard warnings and safety information for reagents and chemicals, consulting safety data sheets.
- Disconnect power before installation and connection/disconnection.
- Only connect to the specified mains voltage.
- Protect the instrument from moisture, condensation, and splash water.
- Do not perform modifications not described in the instructions.
- Maximum electrical load capacity of relay outputs must not be exceeded.
- Immediately switch off and contact service staff if malfunctioning occurs. Unauthorized tampering voids warranty.
- Use only a dry, lint-free cloth for cleaning.
- Dispose of the instrument in accordance with national regulations.
- Assembly and commissioning require electrical and process engineering knowledge and should be performed by specialists.
Installation
- Install in a location protected against dripping/splash water, dust, and aggressive substances (e.g., switch cabinet or wall).
- Install vertically, without mechanical stress, and at a vibration-free site.
- Water inlet hose should be as short as possible (max. 5 m).
- Ensure sufficient space on the left side for the door to open.
- Water pressure must be between 0.3 bar and 8 bar, avoiding strong fluctuations.
- Water inlet connection should be laid vertically upwards to prevent dirt ingress.
- Install a manually operated shut-off valve in the branch to the instrument.
- Use an opaque plastic hose (6/4 x 1) for the water inlet.
- Water outlet hose (internal diameter 12 mm) should be laid without backwater or siphoning effect.
- Connecting cables should be short and installed separately from power cables.
- Connect the instrument to the protective earth conductor (for 230/115 VAC).
- Protect against interference voltages using a mains filter if necessary.
Commissioning
- Use only Heyl Testomat 2000® reagents for guaranteed trouble-free operation.
- Ensure reagent bottles are correctly assembled with the intake hose.
- Fill the intake and transport hoses with reagent from the pump up to the measuring chamber.
- Slowly open the manual shut-off valve for the water inlet to prevent overflowing; allow 2-6 seconds to fill the measuring chamber.
Instrument Settings and Data Input
- Date/Time/Weekday: Configurable via the programming menu.
- Password Protection: A four-digit password (current time in reverse order) is required for basic program settings.
- Operating Mode: Options include Time Controlled, Volume Interval (via water meter), Dynamic Analysis Start, and External Analysis Start.
- Analysis Interval (Time Control): Determined by interval pause (0-99 minutes), flush times, and analysis duration.
- Volume Control: Analysis triggered after a programmed water quantity is measured (1-9999 litres).
- Water Meter Type: Configurable constant (e.g., 100 litres/impulse).
- Displayed Unit: Programmable as mg/l or ppm.
- Parameter Selection: Can measure Free Chlorine or Total Chlorine. For total chlorine, reagents A, B, and C are used. For free chlorine, reagent C is disconnected.
- Flushing Times:
- Internal Flushing: Recommended > 60 seconds if the plant is idle or intervals are long. Minimum 10 seconds for connections > 3 m with 6 mm internal hose diameter.
- External Flushing: Recommended for short analysis intervals or long/large diameter supply lines.
- Interval Pause: 0 to 99 minutes between analyses when using timer triggering.
- Limit Value Monitoring: Two limit values (LV1, LV2) can be programmed. Outputs react based on programmed switch functions.
- Hysteresis: Suppression of the first, second, or third "bad" analysis result (setting 1, 2, or 3) to improve evaluation reliability.
Maintenance
- Refer to the "Maintenance" section and the "Maintenance manual of Testomat 2000®/Testomat ECO®".
- Regular visual inspection is recommended, checking reagent expiry, hose connections, air in hoses, water connections, instrument cleanliness, and measuring chamber/drain.
Technical Data
- See the official PDF for exact tabulated values.
Disclaimer
Extraction may miss figures, footnotes, or revisions; contractual data must match the OEM PDF revision used on the project.
Official datasheet (PDF)
Curated from selected public technical reference material for discovery and preliminary comparison. This summary is not a substitute for a current certified manufacturer datasheet. Verify revisions and design limits before use.