Testomat 2000 THCL

Engineering summary from PDF text extraction for Testomat 2000 THCL. Verify every value with the OEM datasheet.

Summary

Product identification

The Testomat® 2000 THCl is a combined process photometer for total chlorine (0 - 2.5 ppm) and a process titrator for water hardness (0.25 - 2.50 °dH).

Short description

This instrument monitors total chlorine concentration using a colorimetric analysis principle and determines water hardness through titration. It is suitable for applications such as protecting reverse osmosis membranes from hardness and chlorine damage.

Parameter / Measuring range

ParameterUnitRangeResolution
Water hardness°dH0.25 - 2.50(0.05)
°f0.45 - 4.48(0.1)
ppm CaCO₃4.47 - 44.7(0.9)
mmol/l0.04 - 0.45(0.01)
Total Chlorinemg/l (ppm)0.0 - 2.5(0.1)

Operating limits

  • Temperature: Sample water must be between 10 °C and 40 °C.
  • Heavy-metal ions:
    • Iron above 0.5 mg/l
    • Copper above 0.1 mg/l
    • Aluminum above 0.1 mg/l
  • Carbon Dioxide (CO₂): Concentration of more than 100 mg/l can affect analysis.
  • pH range: 4 - 10.5.

Feed water

  • Sampling tube: Maximum length of 5 meters. Must be equipped with a hand-operated shut-off valve.
  • Connection: Branch line connection should be taken vertically from the top of the main soft water line.
  • Hose connection (standard): Opaque plastic hoses 6/4 x 1 (external diameter 6 mm/internal diameter 4 mm).
  • Pressure: If operating with a pressure range of 0.1 to 1 bar, remove the valve body from the regulator/filter housing.
  • Cooler: KCN type cooler must be installed in the branch line if water temperatures are higher than the operating limit.

Product water (Drain)

  • Hose connection: Internal diameter 14 mm.
  • Drainage: Ensure water can run freely to the drain. Use a light-impervious hose for the drain pipe to discourage algae formation.

Electrical

  • Mains input: 230 - 240 V AC (or 24 V AC).
  • Relay outputs: Maximum load 240 V AC, 4 A.
  • Signal inputs: Connect only with volt-free contacts.
  • Optional Interface Output: 0/4 - 20 mA or serial interface (RS 232).

Mechanical / hydraulic

  • Dimensions: See dimensioned sketch (Page 3).
  • Mounting: Unit should be fixed vertically. Avoid tension on the housing.
  • Hose connections: Standard plug connector for opaque plastic hoses 6/4 x 1. Optional adapter for quick-acting coupling available.

Controls and displays

  • Power switch: Located on the right-hand side panel.
  • Display: 4-line LC-Display for current analysis, status results, and programming data.
  • Status LEDs: Signal current status of analysis and unit.
  • Programming keys: Cursor block with ENTER key for value entry.
  • Function keys: Manual, STANDBY, Horn, i-key, M-Taste.

Relay outputs

  • Flush: External flush valve control.
  • LV1 (Limit Value 1): For residual hardness, volt-free change-over contact, programmable functions (Continuous, Impulse, Interval, Hysteresis).
  • LV2 (Limit Value 2): For total chlorine, volt-free change-over contact, programmable functions (Continuous, Impulse, Interval, Hysteresis).
  • Measuring point switch-over (1/2): For monitoring two measuring points, volt-free relay output. Terminal 9 for Measuring point 1 (Residual hardness), Terminal 10 for Measuring point 2 (Total chlorine).
  • AUX (Programmable Function Output): Volt-free relay output, programmable for contact before, during, or after analysis.
  • Alarm: Fault message output, volt-free change-over relay contact.
  • Maintenance: Maintenance message output, volt-free change-over relay contact.

Signal inputs

  • Start: External analysis start (volt-free normally open contact).
  • Stop: External analysis stop (volt-free normally open/closed contact).
  • IN1 (Universal Input 1): External measuring point switch-over (suppression of measuring point 2) (volt-free normally open/closed contact).
  • IN2 (Universal Input 2): Water meter input (volt-free normally open contact) for quantity recording.

Maintenance

  • Interval: Programmable maintenance interval in days.
  • Work: Includes cleaning the measuring chamber, sight-glass windows, and filter housing.
  • Reagent Level Monitoring: Low level indicators for reagents A, B, C, and indicator TH.

Disclaimer

This summary is based on the provided text extract. It may not capture all figures, footnotes, or revisions present in the original OEM PDF datasheet. Contractual data must always match the specific OEM PDF revision used for the project.

Official datasheet (PDF)

Lenntech datasheet mirror