Datasheet summary
Testomat Eco
Engineering summary from PDF text extraction for Testomat Eco. Verify every value with the OEM datasheet.
Summary
Product Identification
Testomat ECO® is an online analysis instrument for residual total hardness (water hardness).
Intended Use
The instrument is designed for the automatic determination and monitoring of residual total hardness in water with a pH level above 4. The required measuring range is determined by indicator selection and user programming.
Operating Limits
- Operating Pressure: 0.3 to 8 bar (factory setting). Can be adjusted to 0.3 to 1 bar by modifying the flow controller.
- Water Temperature: 10 °C to 40 °C. For temperatures above 40 °C, a KCN type cooler should be installed.
- Water pH Level: Above 4.0. For pH < 4.0, a special valve block (Art. No. 40018) may be required.
Feed Water Properties
- Heavy Metal Ions: High concentrations of heavy metal ions can influence color reaction. Iron above 0.5 mg/l, copper above 0.1 mg/l, and aluminum above 0.1 mg/l can cause a brownish-red color display.
- Carbonic Acid (CO₂): If the measuring water contains more than 20 mg/l CO₂, erroneous evaluations cannot be excluded. An aerator type R may be necessary.
Performance Specifications
- Menu-driven operation and programming via plain text display.
- Determines total water hardness.
- Freely selectable hardness units: °dH, °f, ppm CaCO₃, or mmol/l.
- High measuring accuracy with a precise piston-dosing pump.
- Extended operating periods with a 500 ml indicator storage bottle.
- Analysis initiation:
- Automatic interval operation (interval pause settable from 0-99 minutes).
- Quantity-dependent via impulse water meter/turbine.
- External analysis stop.
- Two independent limit values with hysteresis (1, 2, or 3 bad analyses) and settable switch functions via two change-over contacts.
- Fault message output via a change-over contact.
- 0/4 - 20 mA current output.
Indicators for Testomat ECO®
| Parameter/Indicator Type | Unit | Range | Resolution |
|---|---|---|---|
| Water hardness (TH 2005) | °dH | 0.05 - 0.50 | 0.01 |
| °f | 0.09 - 0.89 | 0.02 | |
| ppm CaCO₃ | 0.89 - 8.93 | 0.2 | |
| mmol/l | 0.01 - 0.09 | 0.002 | |
| Water hardness (TH 2025) | °dH | 0.25 – 2.50 | 0.05 |
| °f | 0.45 – 4.48 | 0.1 | |
| ppm CaCO₃ | 4.5 – 44.8 | 0.9 | |
| mmol/l | 0.04 – 0.45 | 0.01 | |
| Water hardness (TH 2100) | °dH | 1.0 – 10.0 | 0.2 |
| °f | 1.8 – 17.9 | 0.4 | |
| ppm CaCO₃ | 18 - 179 | 3.8 | |
| mmol/l | 0.18 – 1.79 | 0.04 | |
| Water hardness (TH 2250) | °dH | 2.5 – 25.0 | 0.5 |
| °f | 4.5 – 44.8 | 1.0 | |
| ppm CaCO₃ | 45 - 448 | 10 | |
| mmol/l | 0.45 – 4.48 | 0.1 |
Electrical
- Mains Voltage Input: 24 V / 115 V / 230 V (refer to rating plate).
- Mains Output for Consumers: Max. 4 A.
- Relay Outputs (LV1, LV2, Alarm): Max. 240 VAC, 4 A.
- Current Interface: 0/4 - 20 mA (galvanically separated).
- Turbine Power Output: +12 V for Hall sensor (turbine), Max. output current 20 mA.
Mechanical / Hydraulic
- Water Inlet Connection: Standard for plastic hoses 6/4 x 1 (6 mm OD / 4 mm ID, 1 mm wall thickness). Maximum hose length: 5 m.
- Water Outlet Connection: For outlet hose with 12 mm internal diameter.
- Indicator Bottle Size: 500 ml (standard).
Commissioning
- Ensure the instrument is only used with Heyl Testomat® indicators.
- Indicator extraction is automatic via the DOSIClip pump. The intake and transport hoses must be filled with indicator.
- Water inlet should be opened slowly. The measuring chamber should fill in 2 to 6 seconds.
Instrument Settings and Data Input
Display Functions
- Limit Value Status Displays (1 & 2): Illuminate red if limit value is reached/exceeded; green if below the limit value.
- Text Display (2 lines): Shows current analysis result, statuses, and programming data. Line 1 displays the measured value (< or > for out-of-range). Line 2 displays set limit values LV1 and LV2.
- Alarm (red LED): Indicates malfunctioning, error, or warning message.
- Analysis Message (yellow LED): Indicates a current analysis is in progress.
- "BOB operation" Symbol (╘╛): Displayed when "operation without permanent supervision" is active.
Operating Elements and Function Keys
- Cursor Block: For navigating menus, selecting functions, and entering data.
- "ENTER" Key: Confirms selections and data input.
- "Manual" Key: Manually starts an analysis.
- "STANDBY" Key: Sets the instrument to standby mode.
- "Horn" Key: Acknowledges error and warning messages.
- "M" Key: Opens the programming menu.
- "i" Key: Retrieves instrument information and settings.
Operating System Programming
- Functions and data are entered via menus using the cursor block and "ENTER" key.
- An asterisk (*) indicates an active function or selection.
Basic Program Data
- Type of Reagent: Select indicator bottle size (100ml or 500ml) and indicator type (TH2005, TH2025, TH2100, TH2250). Factory setting: 500ml bottle, TH2005 indicator.
- Operating Mode: Time control, quantity control (via water meter), or a combination.
- Time Control: Internal start via timer.
- Analysis Interval (Interval Pause): Settable from 0 to 99 minutes. The total analysis interval includes flushing time and analysis duration.
- Quantity Control: Analysis triggered after a programmed water quantity is measured (1-9999 litres).
- Water Meter Type: Select water meter constant (litre/impulse) or reciprocal value (Imp/l). Factory setting: 100 litres/impulse.
- Quantity Control/Time Priority: Defines which control method has priority.
- Display Unit: Selectable units are °dH, °f, ppm CaCO₃, and mmol/l. Factory setting: °dH.
- Flushing Time: Settable in seconds (00-99). Recommended > 60 seconds for longer downtimes or intervals.
Limit Value Monitoring
- Two limit values (LV1 and LV2) can be programmed.
- Hysteresis: Suppression of the first, second, or third "bad" analysis (setting 1, 2, or 3). Default: 1.
- Switch Functions (LV1/LV2):
- 0: Duration (output switches while limit is exceeded).
- 1: Impulse (output switches for a settable time 't').
- 2: Interval (output switches at intervals 't' while limit is exceeded).
- 3: Two-point (for LV1 only, requires different LV1 and LV2 values).
BOB Operation (Operation Without Permanent Supervision)
- Enables continuous monitoring of the indicator level for up to 72 hours.
- Requires time-controlled operating mode.
- Outputs an alarm message if the indicator level is insufficient for the BOB period.
Description of Signal Inputs/Outputs
- Stop Input (Terminals 18, 19): External analysis stop. Programmable as normally closed or normally open. Prevents analysis start or stops a current analysis. Manual start has priority.
- Water Meter Input (Terminals 20, 21): For quantity-dependent analysis triggering.
- Current Interface (Terminals 22, 23): 0/4 - 20 mA output.
- +12 V Output (Terminal 24): For Hall sensor (turbine).
Relay Outputs
- LV1 and LV2 Limit Value Outputs (Terminals 9-11, 12-14): Two independent change-over contacts for limit value monitoring.
- Alarm/Message Output (Terminals 15-17): Change-over contact for fault message output.
Disclaimer: This extraction may miss figures, footnotes, or revisions. Contractual data must match the OEM PDF revision used on the project.