Datasheet summary
Testomat EVO TH
Engineering summary from PDF text extraction for Testomat EVO TH. Verify every value with the OEM datasheet.
Summary
Product Identification
- Product Name: Testomat® EVO TH
- Description: Online analysis instrument for residual total hardness (water hardness).
Intended Use
The Testomat® EVO TH is designed for the automatic determination and monitoring of residual total hardness in water. The measurement scope is determined by the selected indicator and user programming.
Safety Information
- Read and understand the operating instructions thoroughly before use.
- Ensure operating instructions are accessible to all users.
- Observe safety precautions for reagents, chemicals, and cleaning agents, and consult Safety Data Sheets.
- RISK: Indicates a significant danger that will definitely result in severe injury or death if not avoided.
- WARNING: Indicates a possible danger that may result in severe injury or death if not avoided.
- CAUTION: Indicates a potentially dangerous situation that could result in minor to moderate injury or property damage if not avoided.
- NOTE: Indicates important information that could adversely affect operational procedures if not heeded.
General Information
Requirements for Installation Site
- Indoor locations only.
- Ambient temperature: 10 to 40°C.
- Altitude: Under 2000 m.
- Maximum relative humidity: 80% at up to 31°C (declining to 50% at 40°C).
- Must be protected against wetness, moisture, splashed water, and condensate.
- Surge category: II.
- Soiling degree: II.
Requirements for Cables and Lines
- Dielectric strength: 30 V to 260 V according to nominal voltage.
- Cable duct clamping range: 4.5 mm – 10 mm external diameter.
- Terminal strips require core cross-sections of 0.08 mm² to 2.5 mm² (single-wire or fine-wire with ferrules).
- For fine-wire cores with plastic collar ferrules, cross-section up to 1.5 mm².
- Single-wire cores: AWG28 – AWG12.
Requirements for Cable Ducts
- Housing recesses are intended for M16 ducts.
- Duct opening should be smooth and rounded.
- Secure bending protection (five times max. cable diameter).
- Duct must include strain relief.
- Material flammability rating: V1 or better.
Assembly
- Disconnect power before assembly or connection/disconnection from power.
- Connect only to the mains voltage specified on the type plate.
- Observe technical data and environmental parameters.
- Lay mains voltage and relay output cables separately.
- Do not operate without partition walls or terminal area covers.
- ESD Safety: Take necessary measures to avoid electrostatic discharge onto components when opening the casing. Ensure you are earthed.
- Power Switch: The device has no power switch; use an external, clearly marked switch.
- Ensure maximum electrical load capacity of switching outputs is not exceeded (total load must not reach 4A).
- Never attempt to repair the device yourself; repairs must be by authorized personnel.
- Cable Ducting: Loosen strain relief, remove blanking plug, insert cable, tighten union nut.
Connecting Water Intake and Discharge
- Water Temperature: Must be between 10 and 40°C. Water exceeding 40°C can cause burns or damage. A cooler is required for temperatures above 40°C.
- Water Pressure: 0.3 to 8 bar. For pressures of 0.3 to 1 bar, the flow governor core may need to be removed. For pressures under 0.3 bar or suction from a tank, a booster pump (e.g., MepuClip) is recommended. For operation exceeding 8 bar, a pressure reducer is required.
- Intake Lines: Recommended under 3m. For longer lines (>3m), configure purge periods > 60s. For cable lengths > 5-10m, external flushing via AUX is recommended.
- Water Intake Connection:
- Connect directly to the sampling pipe, vertically upwards.
- Install a manual stop valve.
- Use opaque plastic pressure hose 6/4 x 1 (max. 5m).
- Purge intake to remove dirt.
- Water Drain Connection:
- Use supplied funnel and drain hose (internal diameter 12 mm/14 mm).
- Lay hose free of back pressure and siphon effect.
Connecting Mains and Devices
- Risk of Injury: Disconnect all power before assembly.
- Use verified cables with sufficient line cross-section.
- Disconnecting Device: Install an appliance switch or circuit breaker (IEC 608947-1, IEC 60947-3) within reach and clearly marked.
- Electromagnetic Fields: Lay connecting cables separately from mains cables and keep them as short as possible. Shield from strong fields; use a mains filter if necessary. Connect the protective conductor.
- Terminal Designations:
- PE (IN): Mains protective conductor (for 230V and 100-240V).
- N (U) (IN): Grid, Neutral conductor (U for 24V).
- L (V) (IN): Grid, Phase (V for 24V).
- n (OUT): Neutral conductor, secured with 4A.
- l (OUT): Phase, secured with 4A.
- System Components: Connect to output terminals of relays 1-4. If requiring mains power, connect external switched mains voltage to the root contact of the respective relay. Connect neutral conductor and protective conductor as applicable.
- External Voltage: If applying external voltages to relay contacts, affix a warning label to the device.
Terminal Outputs
- LV1/LV2 (Limit Value Outputs): Floating relay output, max. 240 VAC, 4 A. Terminals for root c, NO, NC.
- AUX (Auxiliaries): Floating relay output, max. 240 VAC, 4 A. Terminals for root c, NO, NC.
- Alarm (Fault Signal Output): Floating relay output, max. 240 VAC, 4 A. Terminals for root c, NO, NC.
Connecting Inputs and Outputs
- CAUTION: Do not expose connections to external voltage. Ensure cores are securely in place. Incorrect connections will damage the device.
- Current Loop (0/4-20 mA): Galvanically isolated (Out +, Out - terminals).
- Ext. Ack. (External Acknowledge): Input for error/alarm reset. Break/closing contact programmable; use isolated contact.
- WM (Water Meter): Input for water meter signal. Common ground for inputs. Use isolated contact or observe meter specifications.
- Stop: Input for external analysis disruption. Common ground for inputs. Use isolated break/closing contacts.
RS232 Interface Terminal Assignment (Sub-D 9 poles)
- 2: RXD (Data reception)
- 3: TXD (Output of measured values/alarms)
- 5: GND (Ground)
Indicator Bottle Connection
- Flawless operation is guaranteed only with Heyl Testomat 2000® indicators.
- Ensure the screw cap with hole and insert are correctly assembled onto the indicator bottle's suction hose.
- Water Intake Opening: Slowly open the manual stop valve. The measuring chamber should fill within two to six seconds. Restrict intake if water sprays from the measuring chamber hose.
- Automatic Venting: Occurs after switching on; indicator lines are vented and the measuring chamber purged until indicator is detected. This process cannot be interrupted. Acknowledge "voltage failure" error after completion.
Device Settings and Data Input
Functions of Control and Display Elements
- External Power Switch: Used to turn the device on/off. Wait at least five seconds after switching off before switching on again.
- Internal Device Fuse: Protects against overload and short circuit.
Display Functions
- Limit Value Status Indications (Red/Green LEDs): Indicate if limit values are exceeded (red) or not (green).
- Display: Shows current analytical result, states, and programming data. Configured limit values (LV1, LV2) appear on the left. Measurement range exceeded indicated by "<" or ">".
- LEDs:
- Pause (Flashing Green): Analysis interval suspended.
- LV2/LV1 (Red/Green): Limit value status.
- In (Green): Inlet valve open.
- Analysis Message (Yellow): Analysis in progress.
- Out (Green): Outlet valve open.
- Dosing (Yellow): Dosing pump active.
- BOB operation (Green): BOB operation active.
- Service (Yellow): Maintenance period elapsed.
- Alarm (Red): Functional error, error message, or warning message.
Control and Function Keys
- Manual Button (): Starts manual analysis.
- PAUSE Button (): Puts the device into standby mode (no automatic analysis).
- Horn Button (): Acknowledge error and warning messages.
- M Button (): Accesses the programming menu.
- i Button (): Accesses device information and settings.
Programming Keys (Cursor Block)
- Used for menu navigation, function selection, and data input.
- OK Button: Confirms submenu selections, data input, and adopted settings.
- Navigation Keys (Up/Down/Left/Right): Navigate menus, select options, change data positions and values.
Operating System Symbols (Menu Navigation)
- M / I: Indicates current menu (basic program/service or information).
- / : Indicates additional selection options above or below the current option.
- / : Allows viewing settings (e.g., individual errors).
- +: Indicates the selected option includes an additional submenu.
Available Indicators for Testomat® EVO TH Devices
| Parameter/Indicator Type | Unit | Range (Unit) | Resolution (Unit) |
|---|---|---|---|
| Water Hardness (TH 2005) | °dH | 0.05 - 0.50 | 0.01 |
| °f | 0.09 - 0.89 | 0.02 | |
| [ppm] CaCO₃ | 0.89 - 8.93 | 0.2 | |
| mmol/l | 0.01 - 0.09 | 0.01 | |
| Water Hardness (TH 2025) | °dH | 0.25 - 2.50 | 0.05 |
| °f | 0.45 - 4.48 | 0.1 | |
| [ppm] CaCO₃ | 4.5 - 44.8 | 0.9 | |
| mmol/l | 0.04 - 0.45 | 0.01 | |
| Water Hardness (TH 2100) | °dH | 1.0 - 10.0 | 0.2 |
| °f | 1.8 - 17.9 | 0.4 | |
| [ppm] CaCO₃ | 18 - 179 | 3.8 | |
| mmol/l | 0.18 - 1.79 | 0.04 | |
| Water Hardness (TH 2250) | °dH | 2.5 - 25.0 | 0.5 |
| °f | 4.5 - 44.8 | 1.0 | |
| [ppm] CaCO₃ | 45 - 448 | 10 | |
| mmol/l | 0.45 - 4.48 | 0.1 |
- Use only Heyl indicators; external indicators may cause measurement deviations, errors, or damage, voiding the warranty.
Operating Requirements
- Problem-free operation requires Heyl Testomat indicators and operation within a pH range of 4 – 10.5.
- Operate only within specified parameters in "Technical Data".
- Interference from Heavy Metal Ions:
- Iron: Over 0.5 mg/l
- Copper: Over 0.1 mg/l
- Aluminum: Over 0.1 mg/l (may cause brown-red indication)
- Carbonic Acid (CO₂): If test water contains > 20 mg/l CO₂, erroneous evaluations are possible. Use an irrigator in such cases.
- Test water must be clean and free of bubbles.
- Disruptive ingredients can be determined with TESTOVAL® test comparators.
- Potential Measuring Chamber Soiling: Excessive carbonate hardness, disinfecting agents, or silicates can lead to soiling and erroneous evaluations over time.
- Perform regular visual inspections: indicator expiry, leakproof hose connections, air in dosing hoses, water connections, device cleanliness, measuring chamber/drain channel cleanliness.
- Regular maintenance is crucial.
Delivery Includes
- 1 Testomat® EVO TH
- 1 plastic bag with screw cap and insert.
- 1 package with drain funnel, 2 screws for power supply unit, power supply.
- 1 User manual.
Service Description
- Features:
- Simplified menu-driven operation via backlit graphic LCD.
- Measurement of overall hardness (°dH, °f, ppm CaCO₃, mmol/l).
- Precise piston-dosing pump.
- 500 ml indicator storage bottle.
- Analysis trigger options: automatic interval, water meter, external disruption.
- Two independent limit values with adjustable switching functions.
- SD card logging for measurement data and notifications/alarms (100 measurements, 50 notifications buffer).
- Import/Export of settings.
- Error history (20 notifications).
- Firmware update via SD card.
- Built-in self-test.
- Optional wireless data retrieval with WLAN SD card.
- Integration features: Fault signal output, 0/4-20 mA current loop, RS232 serial interface.
Disclaimer
This extraction may miss figures, footnotes, or revisions. Contractual data must match the OEM PDF revision used on the project.