Datasheet summary
Testomat 2000 Antox
Engineering summary from PDF text extraction for Testomat 2000 Antox. Verify every value with the OEM datasheet.
Summary
Product identification
Testomat 2000® Antox is a special version of Testomat 2000® with an additional pump for dosing a reducing agent. It is designed to control water quality in areas where measurements can be affected by oxidizing agents like chlorine, chlorine dioxide, or ozone.
Application
- Control of water quality in areas where measurement errors can arise due to oxidizing agents.
Measuring range
- Water hardness: 0.05-25 °dH
- Carbonate hardness: 0.5-20 °dH
- p-value: 1-15 mmol/l
- Minus m-value: 0.05-0.5 mmol/l (depending on the choice of the indicator)
Technical Data
Electrical
- Power supply: 230–240 VAC, 115 VAC, 24 VAC, all 50–60 Hz
- Power consumption: max. 30 VA
- Protection class: I
Mechanical / hydraulic
- Degree of protection: IP 65
- Dimensions (W x H x D): 380 x 480 x 280 mm
- Weight: approx. 9.5 kg
- Menu languages: German, English, French, Italian, Polish (others upon request)
- Option: booster pump (see accessories)
Operating limits
- Ambient temperature: 10–45 °C (50 - 104°F)
- Water temperature: 10 to 40 °C (50 to 104°F)
Feed water
- Operating pressure:
- 1 to 8 bar (15 to 116 psi)
- 0.3 to 1 bar (5 to 15 psi), after removing the valve body from the regulator/filter housing
- Water inlet: Opaque hose with 6 mm external diameter / 4 mm internal diameter
- Water outlet: Hose with 6 mm external diameter / 4 mm internal diameter
Disclaimer: This extraction may miss figures, footnotes, or revisions. Contractual data must match the OEM PDF revision used on the project.