Testomat 2000 CAL

Engineering summary from PDF text extraction for Testomat 2000 CAL. Verify every value with the OEM datasheet.

Summary

Product identification

Testomat 2000® CAL

Application

Control of water quality where calibration of the measuring instrument is important, for example, in the pharmaceutical industry.

Measuring range

  • Water hardness: 0.05–25 °dH
  • Carbonate hardness: 0.5–20 °dH
  • p-value: 1–15 mmol/l
  • minus m-value: 0.05–0.5 mmol/l (depending on indicator choice)

Technical Data

  • Power supply: 230–240 VAC, 115 VAC, 24 VAC (all 50–60 Hz)
  • Power consumption: max. 30 VA
  • Protection class: I
  • Degree of protection: IP 65
  • Ambient temperature: 10–45 °C (50 - 104°F)
  • Option: booster pump (see accessories)
  • Dimensions (W x H x D): 380 x 480 x 280 mm
  • Weight: approx. 9.5 kg
  • Menu languages: German, English, French, Italian (others upon request)

Operating limits

  • Operating pressure:
    • 1 to 8 bar (15 to 116 psi)
    • 0.3 to 1 bar (5 to 15 psi) after removing the valve body from the regulator/filter housing
  • Water temperature: 10 to 40 °C (50 to 104°F)

Hydraulic

  • Water inlet: Opaque hose with 6 mm external diameter / 4 mm internal diameter
  • Water outlet: Hose with 6 mm external diameter / 4 mm internal diameter

Disclaimer: This extraction may miss figures, footnotes, or revisions. Contractual data must match the OEM PDF revision used on the project.